1. Introduction
Quartz-enhanced photoacoustic spectroscopy (QEPAS) is a type of photoacoustic spectroscopy which employs a piezoelectric quartz tuning fork instead of a microphone as a transducer. Since the introduction of the QEPAS-technique by Kosterev et al. in 2002 [
1] it has emerged to be a very useful tool for photoacoustic detection of a manifold of gaseous substances. Further progress in QEPAS has been made in terms of using custom tuning forks with larger gaps and lower resonance frequencies to use light sources with larger focal diameters such as LEDs or THz-QCLs or by incorporating the photoacoustic cell and transducer, also called spectrophone, in a multipass cell to increase the power density and thus the sensitivity. A summary of achievements for different species and with respect to the used techniques can be found e.g., in the review articles by Patimisco et al. [
2,
3] and Ma [
4]. For a simple QEPAS setup only a bare quartz tuning fork is required; however, in most spectrophones the tuning fork is additionally equipped with acoustic resonators [
5]. Their use can improve the S/N-ratio of a QEPAS spectrophone by a factor of up to 30 [
6]. For a QEPAS system using the first overtone mode, an enhancement factor of even 50 as compared to the bare tuning fork was reported [
7]. On the other hand a strong coupling between the tuning fork and the acoustic resonator exists and the highest signals are achieved when the resonance frequencies of the tuning fork and the one of the acoustic resonator coincide. The resonance frequency of the acoustic resonator is essentially dependent on its length in relation to the velocity of sound. The most common approach is to use two cylindrical pipe elements sandwiching the tuning fork in a way that the radiation is aimed through both segments and between the prongs of the tuning fork. In this configuration best sensitivity is achieved when the length of a segment is between half and a quarter of the sound wavelength. However, since the tuning fork and the acoustic resonator react differently on changes in the environmental conditions the balanced condition is easily lost when the system is operated in real world scenarios.
Provided that the excitation laser addresses multiple species, the large dynamic range of photoacoustic spectroscopy enables the measurement of both, main components as well as trace gases of a gas mixture. This is one demand for biogas analysis. Therefore, we demonstrated that the system is able to measure methane in a carbon dioxide atmosphere (main components) as well as ammonia in small concentrations as present within the biogas.
Besides the compact size and the small measurement volume of only a few cubic centimeters the high Q-factor of the tuning fork is usually highlighted as an advantage of the QEPAS-technique in contrast to conventional photoacoustic spectroscopy which gives rise to the high sensitivity of a QEPAS-spectrophone. However, at the same time, the high Q-factor results in a sharp resonance curve and thus a careful setting of the laser’s modulation frequency is required. Since the resonance frequency of the spectrophone is affected by environmental conditions such as pressure, temperature, and gas composition related density and viscosity, changes of these parameters can easily cause measurement errors. In this paper a new detection scheme for using both electrical and photoacoustic driving of the tuning fork and phase modulation of these two driving forces is discussed.
2. Materials and Methods
In a typical QEPAS setup a laser beam is focused between the prongs of the tuning fork (and if applicable through an acoustic resonator). The emission wavelength of the laser source is tuned to an absorption line of the target molecule and is additionally modulated with the resonance frequency or half the resonance frequency of the tuning fork depending on the modulation approach, 1f-amplitude modulation or 2f-wavelength modulation, respectively. This is usually done using a separate frequency generator with a precision of at least 0.1 Hz. The tuning fork is connected to a trans-impedance amplifier and the signal is then further amplified by a lock-in amplifier which is referenced by the frequency generator. To determine the correct modulation frequency, the frequency can be scanned in proximity of the expected resonance frequency while recording the signal amplitude of the spectrophone and thus obtaining a resonance curve with a maximum at the spectrophone’s resonance frequency. The modulation frequency of the laser has to be realigned frequently if changes in environmental conditions such as pressure, temperature, and gas composition are expected. Even a complete recalibration might be necessary because of changes of the Q-factor if the environmental conditions change perceivably.
Many QEPAS systems provide the ability to record resonance curves non-photo-acoustically by directly exciting the tuning fork with the frequency generator as reported by Kosterev et al. [
8]. This feature is typically used for presetting the modulation frequency before optical alignment of the spectrophone and for verifying the modulation frequency after changes in the environmental conditions, but not for the actual measurement.
Figure 1 shows the signals for photoacoustic excitation (open red squares), for purely electrical excitation (solid green squares), and for simultaneous electrical and photo-acoustic excitation while the phase between them is altered (filled squares). For this feasibility study pure methane was excited by a cw quantum cascade laser operating at ν = 1276.85 cm
−1 using 2f-modulation. The amplitudes were set to 60 mV for electrical excitation and to 40 mV for photoacoustic excitation by adjusting the amplitude and the modulation width respectively.
Figure 1 shows the resulting signal for simultaneous excitation when the phase of the laser modulation is varied from φ = −210° to +210° while the phase of the electrical excitation is fixed. The sinusoidal sum signal varies between R
LI = 20 and 100 mV. The phase difference between two maxima is Δφ = 180°. Since the phase was set with respect to the modulation frequency which is half the resonance frequency of the tuning fork because of 2f-modulation the actual phase difference is 360°, as expected.
In [
9,
10] we already described the ability to incorporate a quartz micro-tuning fork with removed container in an oscillator circuit and to simultaneously focus a laser beam between the tuning fork’s prongs to generate acoustic waves upon absorption of the molecule. Since the laser modulation signal is derived from the oscillation circuit that automatically adjusts to the current resonance frequency of the tuning fork, the laser is always modulated with the correct frequency. The measurement signal and the reference which are fed to the lock-in amplifier are also derived from the oscillator circuit. The measurement signal is thus the sum of the electrically generated amplitude and the photo-acoustically induced amplitude, which furthermore depends on the phase between both signals. Until now, the phase was manually adjusted as a temporal delay in the modulation line to obtain a maximum signal amplitude. A scheme of this method is shown in
Figure 2a. The phase of a photoacoustic wave is governed by the gas mixture’s relaxation dynamics which in turn depends on the collision partner behavior and thus on the gas mixture composition, its pressure and temperature. Therefore a phase setting optimized for a fixed concentration is valid for a limited range of the mentioned parameters only. This partially limits the advantage of the modulation frequency auto tracking: instead of the realignment of the frequency a realignment of the phase is needed. In this paper however, we report on the ability of a continuous phase modulation instead of a fixed setting which yields a steady measurement and additionally separates the photo-acoustically induced amplitude from the offset generated by the oscillator circuit.
Figure 2b depicts how the auto-triggered QEPAS setup can be extended to avoid the critical phase alignment. Instead of setting a fixed delay the phase is continuously modulated between 0 and 360° with a frequency f
phs = 0.37 Hz which is considerably smaller than the resonance frequency of the tuning fork. The output (A) of the first lock-in amplifier is fed to the input of a second lock-in amplifier which is referenced with the phase modulation frequency f
phs. Without target molecules present in the spectrophone, signal (A) represents the amplitude induced by the oscillator circuit only. It is not modulated and thus output (B) of the second lock-in-amplifier is zero. Upon absorption, signal (A) starts oscillating with the frequency f
phs and the photo-acoustically induced amplitude. The output of the second lock-in amplifier (B) shows the oscillation amplitude of (A), i.e., the pure photo-acoustic amplitude without electrically induced offset.
The functionality of the two lock-in stages was incorporated into one device developed by KNESTEL Technologie & Elektronik GmbH. The device consists of a combination of a micro controller and a FPGA where the functionality of both lock-in amplifiers and phase modulation is digitally programmed.
For the measurements detailed in
Section 3, a pulsed external cavity QCL provided by the Fraunhofer IAF was used because it covers both, methane and ammonia which are important for biogas measurements. The laser emits pulses at a repetition rate f
rep = 400 kHz and is over-modulated with a TTL signal with a duty cycle of roughly 30% (either generated by the oscillator circuit or manually set) thus releasing four laser pulses during each high-level. The laser can be tuned in the frequency range 980 cm
−1–1290 cm
−1.
A spectrophone without acoustic resonators was used for the measurements, which were conducted at ambient pressure and room temperature.