Next Article in Journal
Free-Standing Carbon Nanofiber Films with Supported Cobalt Phosphide Nanoparticles as Cathodes for Hydrogen Evolution Reaction in a Microbial Electrolysis Cell
Previous Article in Journal
Nanomaterial-Enhanced Hybrid Disinfection: A Solution to Combat Multidrug-Resistant Bacteria and Antibiotic Resistance Genes in Wastewater
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Portable Homemade Magnetic Hyperthermia Apparatus: Preliminary Results

1
LEPABE—Laboratory for Process Engineering, Environment, Biotechnology and Energy, Faculty of Engineering, University of Porto, Rua Dr. Roberto Frias, 4200-465 Porto, Portugal
2
AliCE—Associate Laboratory in Chemical Engineering, Faculty of Engineering, University of Porto, Rua Dr. Roberto Frias, 4200-465 Porto, Portugal
3
Institute of Molecular and Cellular Biology of Cancer, CSIC/University of Salamanca (GIR Citómica), 37007 Salamanca, Spain
4
CEADIR—Center for Environmental Studies and Rural Revitalization, Avenida Filiberto Villalobos, 119, 37007 Salamanca, Spain
5
Faculty of Natural Sciences, Engineering and Technologies, Lusófona University of Porto, R. de Augusto Rosa 24, 4000-098 Porto, Portugal
6
CEFT—Center of Study of Phenomena’s of Transport, Faculty of Engineering, University of Porto, Rua Dr. Roberto Frias, 4200-465 Porto, Portugal
*
Author to whom correspondence should be addressed.
Nanomaterials 2024, 14(22), 1848; https://doi.org/10.3390/nano14221848
Submission received: 10 August 2024 / Revised: 28 October 2024 / Accepted: 3 November 2024 / Published: 19 November 2024
(This article belongs to the Special Issue New Insights into the Therapeutic Efficacy of Nanomaterials)

Abstract

:
This study aims to describe and evaluate the performance of a new device for magnetic hyperthermia that can produce an alternating magnetic field with adjustable frequency without the need to change capacitors from the resonant bank, as required by other commercial devices. This innovation, among others, is based on using a capacitator bank that dynamically adjusts the frequency. To validate the novel system, a series of experiments were conducted using commercial magnetic nanoparticles (MNPs) demonstrating the device’s effectiveness and allowing us to identify new challenges associated with the design of more powerful devices. A computational model was also used to validate the device and to allow us to determine the best system configuration. The results obtained are consistent with those from other studies using the same MNPs but with magnetic hyperthermia commercial equipment, confirming the good performance of the developed device (e.g., consistent SAR values between 1.37 and 10.80 W/gMNP were obtained, and experiments reaching temperatures above 43 °C were also obtained). This equipment offers additional advantages, including being economical, user-friendly, and portable.

1. Introduction

Over the last twenty years, magnetic hyperthermia (MH) has been studied as a potential method for treating cancer. It is undergoing clinical trials in Germany [1] and is also starting to be tested clinically in other countries [2]. It is an interdisciplinary field that demands knowledge integration from diverse areas such as quantum physics, materials research, electrical engineering, chemical engineering, biology, biochemistry, and medicine. As a result, scientists face substantial challenges in fully comprehending this technique and its associated therapies, for which further exploration is thus needed.
Hyperthermia treatment, involving heat generation, holds immense potential. For cancer therapy, it entails raising the local temperature of the tumor, thus modifying the physiology of the diseased cells and ultimately causing tissue necrosis. This treatment can effectively complement existing cancer therapies, including chemotherapy, radiotherapy, surgery, and immunotherapy.
Hyperthermia treatments can be categorized into different types based on the extent of the temperature increase:
(1)
High hyperthermia reaches temperatures above 46 °C (up to 56 °C), causing direct tissue necrosis, coagulation, or carbonization of cells.
(2)
Moderate hyperthermia (41 °C < T < 46 °C) has various effects at the cellular and tissue levels. In this temperature range, cells undergo heat stress, resulting in the activation and/or initiation of many extracellular degradation mechanisms, such as protein denaturation, protein folding, DNA aggregation, and cross-linking.
(3)
Low hyperthermia uses temperatures below 41 °C and is applied to treat rheumatic diseases in physiotherapy. It is important to notice that the effectiveness of any hyperthermia treatment depends significantly on the temperatures reached in the targeted sites, the duration of exposure, and the specific characteristics of the cancer cells [3]. The National Cancer Institute (NCI) recognizes three different types of hyperthermia, which are classified based on the area where it is applied and the extent of the area to be treated: local, regional, and whole-body hyperthermia. In local hyperthermia, the primary objective is to heat only tumor cells without damaging healthy tissues. Local hyperthermia is currently a major focus due to its ability to target heat within a specified region [4].
Traditionally, hyperthermia treatment was administered through external devices that transfer energy to the tissues via irradiation with light or electromagnetic waves. Today, there are various techniques available for inducing hyperthermia, such as ultrasound [5], radio frequencies [6], microwave range waves [7], infrared radiation [8], nanoparticle-assisted photothermal therapy [9,10,11], and the use of magnetically excitable nanoparticles. However, each of these methods has its own limitations. Oncologists usually combine hyperthermia treatment with radiotherapy, chemotherapy, or both [12]. Some of the challenges in traditional hyperthermia treatment include:
(1)
The inevitable heating of healthy tissue, resulting in burns, blisters, and discomfort;
(2)
Limited penetration of heat into body tissues by microwave, laser, and ultrasound energy;
(3)
Thermal underdosing in the target area, particularly in areas protected by pelvic or nape bones, often leads to recurrent tumor growth, remaining largely unresolved.
The potential use of magnetic materials in cancer treatment through hyperthermia was initially suggested in 1957, aiming to convert magnetic energy into thermal energy [13]. The rapid progress in magnetic nanoparticles (MNPs) synthesis has represented significant advances in this type of hyperthermia. Hyperthermia treatment using MNPs presents multiple advantages over traditional hyperthermia treatment, as outlined in Table 1.
The exploration of new magnetic nanoparticles is currently at the forefront of biomedical research, particularly in magnetic hyperthermia for cancer treatment. Extensive research is being conducted on superparamagnetic nanoparticles, with an average diameter (σ) of just a few tens of nanometers, to assess their potential in this area. In magnetic hyperthermia, these nanoparticles are exposed to an alternating magnetic field (AMF) with high amplitude ( H 0 ) and high-frequency ( f 0 ).
The “tunable” magnetic properties of MNPs are essential for their biomedical applications. These properties, such as magnetic susceptibility ( χ ), blocking temperature ( T B ), relaxation time (τ), and saturation magnetization ( M s ), can be tailored through different synthesis processes to create specific MNPs for magnetic hyperthermia treatments. The specific absorption rate (SAR), proportional to the nanoparticles’ concentration, is a key indicator of energy dissipation as heat per unit of mass and can be calculated using the initial slope method.
Some researchers prefer to calculate the specific loss power ( S L P ) instead of the SAR, as its definition is less ambiguous. Many studies have measured the S L P for different MNPs. The heating generated by MNPs is attributed to various factors such as eddy currents, the Brownian effect (Brownian relaxation), the crossing of anisotropic barriers (Néel relaxation), and magnetic energy losses due to the magnetic hysteresis of the MNPs.
Utilizing the linear response theory (LRT), valid for low-applied fields and/or highly anisotropic magnetic nanoparticles, this heating power can be mathematically expressed as detailed by [18].
S L P = π × μ 0 × χ 0 × H 0 2 × 2 × f 0 × τ ( 1 + ( 2 × π × f 0 × τ ) 2 ) ( W g )                
where S L P , χ 0 ,   H 0 , μ 0 and f 0 represent specific loss power or specific power absorption, static susceptibility, magnetic field amplitude, vacuum permeability, and frequency, respectively, and τ is the relaxation time, as detailed in [19,20]. Equation (1) demonstrates that both the field amplitude and field frequency influence the dissipated power of MNPs.
Before conducting in vivo experiments, the MNPs must undergo thorough preliminary studies in chemical laboratories to characterize their physical and chemical properties. In addition, in vitro experiments in cellular biology laboratories are necessary to evaluate heating effects and the extent of necrosis or apoptosis in cell cultures. A compact experimental setup, such as the device developed in our lab, provides an excellent alternative for conducting rapid trials and measurements in all these laboratories, making it a powerful tool.
The main goal for synthesized MNPs is to produce a high SAR with a low concentration of particles when exposed to an alternating magnetic field. There are various systems proposed for generating alternating magnetic fields, including single- and double-layer solenoids, Helmholtz coils, inductors with a ferromagnetic core (in horseshoe shape), and coils made with Litz wire [21,22]. The single-layer solenoid refrigerated system is the most widely used, even in commercial equipment, and it is the primary focus of this work.
The development of MH equipment remains vastly unexplored due to its intricate technology demands, necessitating both high power and high frequency at the same time. Commercially available devices are, therefore, complex equipment that also does not present the desired compactness and portability. Consequently, and due to its complexity and, hence, associated costs only a handful of companies manufacture this equipment, leading to significant economic barriers that, among others, impede more in-depth research and widespread application of this promising technique.
It is already known that in order to create devices that are able to achieve the frequencies and magnetic fields required for magnetic hyperthermia, the best option is to use a resonant circuit. Many different systems with different configurations have been developed so far to reach this goal. The current options that exist commercially are complicated (thus hard to adapt and costly) and/or do not allow for a dynamic change of frequencies, making the overall device rather complex, expensive, hard to handle, and non-portable. In the last years, research has been performed trying to innovate and achieve devices where a change of frequencies does not require the physical change of capacitors, among other concerns; however, all these solutions present limited success only. For example, Mazon et al. [23] present a device where frequency tuning of a magnetic hyperthermia device is achieved by switching control capacitors. Nonetheless, these and similar cases present the problem that the tactic used for frequency tunning implies the use of microprocessors with a drive circuit and an H bridge and an inverter, demanding also the use of a PLL circuit for better enhancing the results of the output frequencies, making the all-device complex, costly, and non-portable. In the work that is detailed in this article, we present an innovative system that was very challenging to design and develop, as it joined the capacity of frequency tuning by switching control capacitors with the simplified version of a resonant circuit, without the need to use complex microprocessors, inverters, etc. The outcome is a device capable of performing magnetic hyperthermia with dynamical frequency tunning based on a simplified circuit, which makes it low-cost, portable, and flexible, and, thus, allows for simpler and cheaper magnetic hyperthermia research studies and also for possible worldwide commercial use.

2. Materials and Methods

2.1. Magnetic Field Simulations

The magnetic field simulations were conducted using the “Magnetic fields” module from the commercial software COMSOL Multiphysics®, version 6.0. For computational purposes, an HP workstation Z4 (HP Headquarters, Lisbon, Portugal) with an Intel(R) Xeon(R) W-2295 CPU @ 3.00 GHz was used.
A typical solenoid shape was chosen because it preserves the direction of the magnetic field within its interior [24]. The coil itself is composed of hollow copper tubes with an inner radius of 1.5 mm and thickness of 1 mm, into which a water pumping system is connected to control temperature changes due to the Joule effect. To complete the geometry step, the coil was placed inside a parallelepiped domain composed of air. The example of S5 is depicted in Figure 1. The simulation of the magnetic field was performed with a finite element method (FEM), where the domain is discretized into small triangular elements. In each element, the magnetic field is computed through a frequency domain solver. In our simulation, we employed a dense mesh to accurately capture the system’s complexities. For the copper tube and the water domain, we implemented a specialized boundary condition designed to effectively simulate the skin depth effect associated with the electric current. This approach allowed us to better represent the penetration of the current into the conductive material, ensuring a more realistic analysis of the interaction between the copper tube and the surrounding water. In total, the mesh consists of 365,281 elements with a minimum relative quality of 0.178 and an average quality of 0.6485, being totally composed of 3 main domains (water, coil, air), 18 faces, 36 edges, and, finally, by 24 points.
Magnetostatic numerical models:
× H = J
B = μ r μ 0 H
B = × A
where J is the current density vector, given by the product of the material’s electrical conductivity (σ) and the electric field (E), J = σ   E . This field is computed based on the type of excitation in the coil. In this study, the coils are voltage-driven, so the electric field is derived from the divergence of the electric potential, E = · V . To account for the Joule effect, the electrical losses are calculated using the following equation:
Q = 1 T t t + T J E   d t
where Q represents the electromagnetic losses. Finally, magnetic insulation boundary conditions, n ^ × A were applied to the faces of the air parallelepiped domain. The materials were selected from the “built-in” libraries from the COMSOL software. The properties are listed in Table 2.

2.2. Hyperthermia Studies

The initial examination of the developed magnetic hyperthermia (MH) system as presented in this article involved the use of commercially available MNPs. These MNPs were carefully chosen based on a comprehensive review of the literature, highlighting their positive results [25,26,27,28,29]. Their characteristics and designations are detailed in Table 3. Hence, seven different iron-oxide-based magnetic liquid samples from Chemicell GmbH (Berlin, Germany) were evaluated (Table 4), following studies by Kallumadil et al. [26]); Yaremenko, A.V et al. [30]; and Wang, Huang et al. [31]. These samples varied in size and core-shell configurations, all featuring a magnetite core. The MNPs had previously undergone comprehensive testing and characterization, particularly concerning their hydrodynamic diameters.
The heating efficacy is represented by the specific absorption rate (SAR) value, expressed in W/g(Magnetic NanoParticles). This value corresponds to the amount of energy converted into heat (J) per unit time (s) and mass (mmag) of the magnetic material. The SAR was calculated using the initial slope method (Figure 2), which considers the first few minutes (Equation (6)) within the linear variation range of the heating curve (adiabatic regime). This period is typically set at 100 s based on theoretical assumptions (low frequency and magnetic field).
S . A . R ( Specific   Absortion   Rate ) = C msol × m s o l m m a g d T d t = ( m H 2 O × C p H 2 O +   m M N P   × C p M N P )   m M N P   × d T d t
where C msol is the specific heat of the solution, m s o l   is the mass of the solution, m m a g is the mass of the MNPs, and d T d t is the maximum value of the initial linear slope.
The concentration range for measuring the specific absorption rate (SAR) of the of the MNPs was 5 to 25 mgMNP/mL. This measurement was conducted using the developed apparatus by determining the heating curve (temperature variation over time). For statistical accuracy, three measurements were taken for each sample.
In Appendix A, details are shown about how magnetic fields were measured and accounted for.

3. Results and Discussion

3.1. Configuration, Specifications, and Fundamental Operational Principles of an Alternating Magnetic Field Generator

In the past two decades, there has been remarkable progress in the development of devices for magnetic hyperthermia applications. These devices utilize alternating magnetic fields at high frequencies and power levels [33]. However, most of these devices are primarily designed for whole-body applications [1]. Due to their large size, these devices are not suitable for use in research laboratories. Recently, various switched-mode resonant inverters have been developed for medical use in electromagnetic thermotherapy. These inverters use voltage-fed high-frequency applicators with power metal-oxide-semiconductor field-effect transistors (MOSFETs) operating based on resonant circuits as further explained below [34]. To achieve the necessary alternating magnetic field for hyperthermia, a high alternating electric current at a specific frequency must be passed through a coil, typically at high frequencies (>100 kHz). To accomplish this, two main technologies are commonly used: resonant circuits and H-bridge-based circuits, also known as inverters. For more information on how these circuits work and their components, see Cabrera et al., 2019 [35].
In this groundbreaking work, a resonant circuit was employed, harnessing a technique that has undergone extensive study over the past three decades. This approach has captured substantial attention from both academic and industrial research communities due to its remarkable characteristics: smooth waveforms, exceptional efficiency, and high-power density. Resonant circuits are based on transforming a constant direct current power source into a sinusoidal wave. This is made possible through a resonant circuit based on a parallel LC (inductor/capacitor) configuration, as depicted in Figure 3.
Resonant converters encompass a diverse and extensive family, which can pose a straightforward description challenge. However, one common feature shared by most, if not all, is their reliance on a “resonant inverter” to convert DC voltage into sinusoidal voltage and deliver AC power. These converters utilize major topologies such as full-bridge zero voltage switching (ZVS), half-bridge ZVS, full-bridge zero current switching (ZCS), and half-bridge ZCS series converters [36]. For this project, we opted for the half-bridge ZVS series converter, a simple yet widely used choice (see Figure 1). The ZVS system is typically employed in high-frequency switching applications, offering advantages including reduced power losses, enhanced reliability, and improved overall performance of power electronic systems [37].
To attain maximum performance, the converter needs to operate at the resonant frequency. Equations (7) and (8) precisely define the resonance frequency fr and the peak voltage VM across the coil for the basic ZVS series converter at the operating frequency.
f r = 1 2 × π × L × C
V M = X L × I m a x = 2 × π × f r × L × I m a x
In Equations (7) and (8), L denotes the coil’s inductance, while C represents the system’s capacitance. XL stands for inductive reactance, which varies with frequency.
In our research, we attempted to use the simplest and most efficient system that would allow us to construct an economical and portable device for use by the scientific community.
The experimental program initiated with the establishment of a resonant circuit to generate a sinusoidal wave. The wave’s frequency is contingent on the coil configuration and the capacitors positioned in parallel to form the resonant oscillator circuit. Typically, this circuit maintains a fixed frequency, obstructing the analysis of particle behavior at different frequencies for identifying the optimal operating value. To surmount this limitation, a bank of capacitors, managed by a set of switches (six in this scenario), was introduced into the circuit. These switches enable the parallel placement of more or fewer capacitors with the test coil, thereby modifying the resonant frequency of the LC circuit and consequently adjusting the test frequency (see Figure 4). This apparatus has demonstrated remarkable effectiveness in operating at high powers and high frequencies (frequency range 69–303 khz and generated magnetic field range 4 mT–16 mT).
Upon testing our initial resonant system (see Figure 5a) with a fixed bank of capacitors, we achieved notable positive results. These results were particularly evident when we modified the coil’s configuration to increase field concentration by adjusting the current flow due to lower impedance (2.5 to 5 µH inductance) and by changing the frequency.
In Figure 5, the progression from our initial system prototype to the current development stage is depicted, even into a new, more powerful prototype currently under testing. The magnetic hyperthermia device assembly consists of five key components:
(1)
Power source (VELLEMAN, Model: LABPS6030SM);
(2)
Electric circuit (designed by us and manufactured by JMP Electronics (Jinhu, China));
(3)
Cooling system (designed by us);
(4)
Oscilloscope coupled with a probe for measuring the magnetic field (see Appendix A) and verifying the waveform passing through the coil (Promax Electronics, Model: OD-624);
(5)
Temperature recorder and probe for measuring the sample temperature (Chaviun1822 with probe K presenting an accuracy ±2%). In the magnitude range of the magnetic field, the inference is minimal, as demonstrated in the next section [32,33,38].
The power source was carefully chosen to ensure the required voltage (ΔV) for generating an electric current capable of creating a high-intensity magnetic field. The specific MOSFET selected, IRFP260, can handle a maximum current of 60 A and support a voltage of 200 V.
The resonant system and coil were meticulously designed to produce a uniform magnetic field inside the sample. Several coils were designed and tested, the default system being composed of a coil, made of copper and comprising a 2 cm diameter, 11 cm length, and 17 turns, designed to be cooled by flowing water to maintain room temperature. It is important to notice that under ideal conditions, Equations (7) and (8) are well fitted to the real maximum current and frequency of the resonant circuit, respectively.
The maximum intensity of the magnetic field in the coils is given by Equation (9),
B M A X = μ 0 × I c u r r . m a x × N l e
N represents the number of turns of the coil, μ0 stands for the magnetic permeability in a vacuum (4π × 10−7 H/m), Icurr.max denotes the maximum current intensity, and le indicates the length of the solenoid. The high current running through the system causes the solenoid to heat up due to the Joule effect. To maintain a constant temperature, it is essential to cool it by circulating water within the solenoid. Priority is given to improve the insulation system, and currently, we are exploring materials such as mineral wool, AMORIM-expanded cork, and Styrofoam for the sample holder and insulation. Figure 6 and Figure 7 illustrate a diagram of the setup and a schematic representation of the system with the cooled coil containing a sample inside.
To improve the performance of the magnetic hyperthermia system, various experiments were conducted with different coil configurations (see Table 4) that are depicted in Figure 8. The results of these experiments will be presented throughout the text.

3.2. Results of Magnetic Field Simulations

This section presents the numerical results obtained for the different coils used in the hyperthermia studies described in this work.
Parametric frequency studies of solenoid coils
The geometric parameters of each solenoid coil can be found in Table 4. In these numerical studies, only coils S5, S6, and S8 were simulated (see example of S5 in Figure 9), for reasons detailed in the next sub-section. For coil S8, two simulations were conducted with different numbers of turns, 12 and 17. To obtain the electric properties of the selected coils, a parametric frequency run was performed using frequencies of 72, 96, 134, and 302 kHz. The applied voltage to excite the coils was set to a constant value of 60 V.
Comparison of coils at various frequencies
To compare each coil, two criteria were established: the shape of the magnetic field density at the center of the coil and the electric properties of each coil. The first criterion involved evaluating the magnetic field density relative to the coil’s length. This method allowed us to determine whether the magnetic field density inside the coil followed a horseshoe profile or remained constant along its length. The second criterion focused on the electrical properties, such as impedance, current, and induction, when different frequencies were applied to the solenoid.
The initial solenoids (S1–S4) were tested but did not yield significant results; therefore, simulations were only performed for S5–S8. Solenoid S1 did not produce a satisfactory magnetic field value and, like S2, exhibited a less concentrated, non-uniform field with high impedance. S3 presented challenges in implementing an insulation system due to its narrow design, while S4 had a less concentrated magnetic field compared to S2. Additionally, coil S7 was discarded because it heated up as the electric current passed through it.
Figure 10 displays the results of the first criterion evaluation. The blue solid lines represent coils operating at a frequency of 72 kHz, while the red dotted lines, black dashed lines, and green dashed–dot lines correspond to frequencies of 96 kHz, 134 kHz, and 302 kHz, respectively. The symbols (*), (◊), (□), and (☼) denote solenoid coils S5, S6, S8 with 12 turns, and S8 with 17 turns, respectively.
Analysis of the image reveals that the magnetic field strength decreases with increasing frequency, as demonstrated by [40]. Overall, coil S6 exhibited the highest magnetic field density, reaching peaks above 16 mT at 72 kHz and approximately 3.5 mT at 302 kHz. In contrast, coil S5 had the lowest value, peaking at around 3.5 mT at 72 kHz and below 1 mT at 302 kHz. However, coil S8, both with 12 turns and 17 turns, demonstrated a consistent magnetic field strength across all frequencies, showing a flat profile.
The second criterion analysis is based on Table 5, which presents the electrical properties of each solenoid excited by a sinusoidal applied voltage of 60 V. As shown, S6 exhibited the highest current values and consequently the lowest impedance inductance, ranging from 1.07 μH to 1.08 μH. S8 achieved impedances ranging from 2.13 μH to 2.15 μH for 12 turns and from 3.03 μH to 3.06 μH for 17 turns. Similar to the trend observed in the first criterion evaluation, S5 demonstrated the poorest performance, with inductances ranging between 3.91 μH and 3.93 μH. This is consistent with Equation (10), as the impedance varies linearly with the inductance of the coil and the frequency. Additionally, Equation (11) describes the inductance, indicating a quadratic relationship with the number of turns and an inverse linear relationship with the length of the coil. Thus, the simulation results align with the expected outcomes.

3.3. Hyperthermia Essays

In order to assess the effectiveness of the insulation system, a critical component for adiabatic heating, we conducted tests using distilled water (DW) and observed its heating patterns. As depicted in Figure 11, over a 10 min heating period at a frequency of 69 kHz, the temperature rose by a mere 2.5 °C across three separate days, affirming the consistency of our measurements. Initially, we meticulously monitored the sample’s temperature on a minute-by-minute basis to ascertain any signs of heating. Once the device’s reliability was confirmed, we obtained continuous data and meticulously documented it for the purpose of calculating temperature variation per second. This approach enabled us to utilize more precise methodologies, such as the Lucas box method and the increment-corrected method, to derive the SAR value [39,41].
The initial experiments were conducted using solenoid S5, and the first particles tested were FluidmagARA and FluidmagUCA. FluidmagARA consists of coated particles, while FluidmagUCA features uncoated particles with anionic charges. The nanoparticle concentration used was 25 mg/mL. FluidmagARA particles are composed of a multi-domain magnetite core coated with a layer of glucuronic acid [29,42]. At a frequency of 98 kHz, the temperature increased by   9.2   ° C   in the FluidmagUCA sample and 6.8   ° C   in the FluidmagARA sample. These results align with the parameters listed in Table 5, as FluidmagARA particles are smaller than UCA particles, resulting in fewer less magnetic domains and a lower iron percentage. Additionally, it is known that the coating on nanoparticles can reduce the magnetic saturation value (Ms), which is consistent with the observed results. The heating curves and SAR values are presented in Figure 12, showing SAR values of 1.76 W/gMNP-UCA and 1.37 W/gMNP-ARA.
To verify that the heating was specifically due to the presence of the magnetic particles, we conducted a control experiment using a blank sample of distilled water (DW) with the same volume as the analyzed samples. The heating curve for this blank sample showed a temperature increase of only 2 °C. Additionally, we studied a sample of FluidMagD100nn at a frequency of 69 kHz and repeated the experiment at least three times to assess its reproducibility. The resulting specific absorption rate (SAR) was 4.34 W/gMNP (69 kHz), and the system and procedure exhibited excellent reproducibility, consistently reaching nearly the same final temperature within the last 10 min, even when starting at a lower temperature. The higher SAR values observed align with those reported by Kallumadil and colleagues [30]. In Figure 13, it can be seen that the temperature did not reach the expected plateau associated with magnetic hyperthermia, where a stable equilibrium between heat lost and heat generated is achieved. The main aim of this study was to confirm that the heating was indeed caused by the magnetic nanoparticles and to validate that the device operates correctly with acceptable reproducibility. This was confirmed by the third repetition: although the initial temperature was lower, the final temperature was almost the same in the last 10 min. Given more time, the final temperature would likely have reached a stable state similar to previously observed results.
Figure 14 illustrates the heating curve of FluidmagD50 at 69 kHz using solenoid S5. The SAR values obtained were 2.47 W/gMNP for FluidmagD50 and 3.5 W/gMNP for FluidmagD100. Even though distilled water samples were used in all experiments, their results are not displayed in the graphs to avoid redundancy.
The study investigated how frequency impacts the heating curves of the samples to find the best frequency. Figure 15 displays the linear fitting that was utilized to find the initial slope for calculating the SAR value, as outlined in Equation (9).
The slight variation in frequency, ranging from 63 to 138 kHz, leads to a corresponding change in the final temperature. This change is due to the differences in the current magnitude at different frequencies. At 63 kHz, the current is higher compared to 138 kHz due to impedance. To address this issue, it is necessary to either increase the voltage of the power source or design a new coil. The frequency of 78 kHz produces the best results, as it is higher than 63 kHz but still maintains a similar current magnitude.
The SAR values obtained are as follows: 2.43 W/gMNP at 63 kHz, 2.22 W/gMNP at 81 kHz, 3.54 W/gMNP at 78 kHz, and 2.03 W/gMNP at 138 kHz. Since the magnetic field is directly proportional to the current, as described in Equation (4), increasing the current significantly impacts this parameter. It is possible that if we had continued the experiment at 138 kHz, the temperature would reach a value like what was observed at 13 min at 78 kHz. In the case of the FluidMagDx100nn sample, a final temperature of 36.7 °C was reached at 78 kHz using the system configuration with the refrigerated solenoid S5. This produced a SAR value of 3.54 W/gMNP (refer to Figure 15). These results are promising, as they were achieved using a lower frequency of 78 kHz and a lower magnetic field compared to the study by Kallumadil and colleagues. Furthermore, experiments were carried out with various types of particles to compare heating curves at 138 kHz and to determine the optimal system for heating studies. The SAR values obtained were as follows: FluidMagCMX = 1.56 W/gMNP, FluidMagLip200nm = 1.92 W/gMNP, FluidMagDx100nn = 2.03 W/gMNP, and FluidMagDx50nm = 2.70 W/gMNP (refer to Figure 16).
Figure 16 summarizes the results obtained for the four samples at 138 kHz. Interestingly, the FluidMag200 nm sample seems to exhibit lower heating efficiency, whereas the FluidMagCMX sample, with a broader size distribution as reported by the manufacturer, demonstrates a less intense heating effect. This difference may be attributed to magnetic properties, such as higher magnetization values associated with larger MNPs.
Furthermore, a new set of experiments employing a new coil, designated as S6, with reduced turns, and narrower and shorter dimensions, was performed. Given that the inductive reactance   X L relies on inductance (L) and the length of the solenoid l e , as outlined in Equations (10) and (11), it is anticipated that the alterations in the coil design will impact the heating performance.
X L = 2 × π × f × L [ ohms ]
L = ( N 2 × A × μ ) l e [ Henry ]
In Equation (10), the symbol (f) represents the frequency, L denotes the inductance of the solenoid, N is the number of turns, µ is the modulus of the magnetic permeability of the material, A is the cross-sectional area in m2, and l e is the length of the solenoid. The objective was to reduce the length of the solenoid, thereby decreasing l e , and to concentrate the magnetic field by using a narrower inductor and a reduced cross-sectional area with the new coil, S6. The FluidMagDx100nm sample, with a concentration of 12.5 mgMNP/mL, was tested to assess improvements in heating performance. The results showed final temperatures ranging from 28.8 °C to 29.7 °C on the frequencies 110–170 kHz. The corresponding SAR values were 2.62 W/gMNP at 141 kHz, 3.60 W/gMNP at 110 kHz, and 3.49 W/gMNP at 170 kHz (see Figure 17).
A new coil configuration, designated as S8, was designed with more turns and a smaller radius, resulting in a more concentrated magnetic field. This configuration improved system performance, achieving an alternating magnetic field (AMF) H = 4   ± 2   mT . The heating of the FluidmagUCA sample, with a concentration of 12.5 mg/mL, improved, reaching 35.5 °C in 9 min at 72 kHz, with a SAR of 4.83 W/gMNP UCA (see Figure 18). For the FluidMagD100nm sample, the SAR was 3.1 W/gMNP, which is consistent with the literature values [43].
These results can be explained by the S8 coil’s 17 turns and smaller diameter, which concentrate the magnetic field lines. Additionally, the current passing through the solenoid is nearly doubled (see Table 5), which is directly proportional to the intensity of the alternating magnetic field.
Regarding the MagFluidD100nm, the previous results were significantly affected, reinforcing the notion that the maximum temperature threshold was likely reached within the frequency range previously observed with coil S6 (see Figure 19).
FluidMagDx50nm samples were used at different frequencies with a concentration of 25 mgMNP/mL in this new device configuration. As anticipated, the results improved, and the graph also shows the reference for the DW sample and the repeatability in this new configuration (see Figure 20). The SAR values were as follows: 3.55 W/gMNP at 302 kHz, 5.40 W/gMNP at 99 kHz, and 7.74 W/gMNP at 132 kHz. The optimal frequency was around 132 kHz. The Chemicell samples exhibit rapid heating under the 132 kHz applied magnetic field, consistent with the findings of Eggeman and co-workers [44], with the temperature notably increase at 132 kHz.
The heating curves for the two samples of FluidMagDX50nm are shown in Figure 21, as a function of average concentration (5–25 mg/mL). The Chemicell samples were heated using an applied magnetic field at 132 kHz, with a field amplitude of approximately 8 mT. The initial slope decreases when the sample is diluted from 25 mg/mL to 5 mg/mL, which aligns with the findings of [44]. Therefore, it appears that dilution does not improve the heating efficiency in this type of MNPs. As state previously, macroscopic dilution does not necessarily change the inter-particle distance within a cluster; it may only increase the average cluster spacing, as the MNPs are composed of clusters as described before.
At a frequency of 132 kHz, FluidMagDX50nm exhibits a specific absorption rate (SAR) of 1.95 W/g for a sample concentration of 5 mg/mL and 9.13 W/g for a sample concentration of 25 mg/mL. The final temperature for the more concentrated sample reached 47.2 °C.
To summarize, we tested the FluidMagDX50nm sample at 101.5 kHz using a new configuration where we changed the capacitors to a specific value determined by us through Equation (7) to achieve a higher frequency. We selected these particles because they have previously produced good results [30]. Therefore, we wanted to determine if our device could achieve comparable or even better results than those obtained by other researchers. We conducted a comprehensive experiment for magnetic hyperthermia, which included heating and cooling curves. The results were very promising, as depicted in Figure 22, yielding a SAR of 10.8 W/gMNP.
A resume of the final temperature and results obtained from all experiences are presented in Table 6 with the respective average and deviation; almost all of the experiences were performed in triplicate as a minimum.

4. Conclusions

In this study, we developed, built, and tested a new magnetic hyperthermia device. We evaluated its performance using commercial magnetic nanoparticles and found that it produced results similar to those of commercial devices. We optimized the system to achieve higher magnetic fields and frequencies, and the simulation results confirmed our practical findings. This device is more cost-effective and portable compared to existing options, and we aim to make this technology more accessible.
According to our findings and previous research [25,26,45], the Chemicell iron oxide sample FluidMagDx appears to be a suitable material for magnetic hyperthermia (MHT) applications. The cluster size is large enough to prevent endocytosis at a cell membrane, yet small enough to avoid activating bacterial defense mechanisms. The coating material, dextran, makes it biocompatible. However, the stability of the clusters inside biological organisms is still unknown [46]. As a result, “FluidMAG” from Chemicell GmbH showed a significant heating response that can be adjusted by the field and frequency of the AC driving field and the solution concentration.
Our results confirm that significant optimization can be achieved in a magnetic hyperthermia device, such as increasing the frequency, the magnitude of the magnetic field, or even designing a programmable system.
According to our results, factors such as the size, composition, and shape of particles may not be enough to accurately predict the magnetic response to different frequencies and, as a result, the rate of heat generation in an AC magnetic field. We can only make initial assumptions based on previous research, so our next step is to create various types of magnetic nanoparticles and see how they affect the heating rate.
The concentration of iron in the sample influences the heating, but local clustering is also important.
In our upcoming research, we will concentrate on creating MNPs and adjusting their properties to improve their performance. We are in the process of designing and building a new apparatus that is similar to the resonant system but capable of generating higher magnetic fields and frequencies, as outlined in Castelo-Grande et al. [39].
We have encountered an issue with controlling test frequencies and have addressed it by implementing a circuit based on the H-bridge using MOSFET transistors (IRFP260 or IRFP4668) as switches. These switches are controlled by a computational system (Arduino MEGA platform), enabling us to generate square waves at the desired frequency. However, we faced challenges in controlling the switches on the upper part of the H-bridge (H-bridge high side or low side).
Based on our analysis, it is evident that there is still significant room for improvement in the field of magnetic nanoparticle hyperthermia. This could involve developing more user-friendly devices with a wider range of designs and configurations, thereby paving the way for further exploration. In terms of MNPs, there is still much to learn in order to accurately predict the optimal frequency and field for MH, which are closely linked to intrinsic parameters [47].
For possible future developments, including for example, ex vivo and in vivo biological sample applications, and comparison with simulated data (on the hyperthermia treatment), collaborations with health institutes of Porto and Salamanca have been initiated.

5. Patents

A patent (reference PPP 20242006604014) was filed concerning the developed innovative magnetic hyperthermia device described in this paper.

Author Contributions

Conceptualization, T.C.-G., P.A.A. and L.G.; methodology, T.C.-G., L.G. and E.C.; software, L.G. and E.C.; validation, T.C.-G., L.G. and E.C.; formal analysis, T.C.-G., P.A.A. and D.B.; investigation, T.C.-G., P.A.A. and E.C.; resources, T.C.-G., P.A.A., L.G., E.C. and D.B. data curation, T.C.-G., P.A.A. and E.C.; writing—original draft preparation, T.C.-G.; writing—review and editing, T.C.-G., P.A.A. and D.B., visualization, T.C.-G. and P.A.A.; supervision, T.C.-G., P.A.A. and D.B.; project administration, T.C.-G. and P.A.A.; funding acquisition, T.C.-G., P.A.A. and D.B. All authors have read and agreed to the published version of the manuscript.

Funding

This work was financially supported by (a) Contract 2021 02188 CEECIND (Teresa Castelo-Grande); (b) Project MAGLIDUO (POCI-01-0145-FEDER-028020), funded by FEDER funds through COMPETE2020—Programa Operacional Competitividade e Internacionalização (POCI) and by national funds (PIDDAC) through FCT/MCTES; (c) LEPABE, UIDB/00511/2020 (DOI: 10.54499/UIDB/00511/2020) and UIDP/00511/2020 (DOI: 10.54499/UIDP/00511/2020) and ALiCE, LA/P/0045/2020 (DOI: 10.54499/LA/P/0045/2020); (d) Project MAT2016-76824-C3-2-R, “Diseño de Nanohilos Magneticos para su uso en Tecnologias Limpias”, MINECO, Spain; (e) Project 2019/00057/001—“Application of Magnetic Methods and Magnetic Substances for Environmental Treatment and Biomedicine”.

Data Availability Statement

Data are contained within the article.

Acknowledgments

Helena Sant’Ovaia (Instituto de Ciências da Terra—Pólo da Faculdade de Ciências U.P.) is gratefully acknowledged for her assistance with measuring the magnetic particle properties.

Conflicts of Interest

The authors declare no conflicts of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

Appendix A

Magnetic field measurement
The LF-R 400 H field is a passive near-field probe that has a large diameter (25 mm), making it highly sensitive and suitable for measurements in ranges of up to 10 cm around assemblies and fixtures, as it is proven capable of detecting a larger area of the magnetic field specially comparing with the LF-R 50 (10 mm) or LF-R 3 (3 mm) near-field probes). It has a current attenuation sheath and is, therefore, electrically shielded. It can be connected to a spectrum analyzer or an oscilloscope with a 50 Ω input. An oscilloscope with a 1:1 tip was used, which recorded the Vvalue (peak to peak) and verified the sinusoidal waveform, as well as the frequency obtained in the coil.
Equations used to calculate the magnetic field
Probe RF-R 400-1- Langer EMV-Technik:
d B = = μ [ d B μ ] A m = U p r o b e [ d B μ V ] + [ d B μ A μ V v m ]
A signal attenuator was also used to be able to read the obtained signal. For this we have to convert the value read by the probe with the attenuator. Since this is 5 dB, the following equation must be solved:
5 ( d B ) = 20 × l o g 10 ( x )               where   we   want   to   obtain   x
Solving, we obtain the value of x = 0.56, which means it attenuates 56% of the value.
o r   5 ( d B ) = 20 × l o g 10 ( V o u t V i n )   w h e r e   w e   w a n t   t o   o b t a i n   v i n
For example, if the value read was [∆V]pp = 95 ± 5 V <=>95/0.56 = 169.64, which gives approximation—[∆V]p = 85, then it is necessary to resolve the following equation in order to obtain AMF.
H ( k A m ) = r e f v a l u e ( k H z ) × 10 ( V p 20 ) / 1000
where r e f v a l u e ( k H z ) is a value obtained from the data sheet of the equipment and it is the function of frequency used, V p value of the high of pic of the wavelength shown in the oscilloscope. In the particular case of this work, the range of magnetic field is from 3.8–6.5 kA/m or 4.6–8 mT depending on the configuration system (different solenoids).

References

  1. Mahmoudi, K.; Bouras, A.; Bozec, D.; Ivkov, R.; Hadjipanayis, C. Magnetic hyperthermia therapy for the treatment of glioblastoma: A review of the therapy’s history, efficacy and application in humans. Int. J. Hyperth. 2018, 34, 1316–1328. [Google Scholar] [CrossRef] [PubMed]
  2. Rivera, D.; Schupper, A.J.; Bouras, A.; Anastasiadou, M.; Kleinberg, L.; Kraitchman, D.L.; Attaluri, A.; Ivkov, R.; Hadjipanayis, C.G. Neurosurgical Applications of Magnetic Hyperthermia Therapy. Neurosurg. Clin. N. Am. 2023, 34, 269–283. [Google Scholar] [CrossRef] [PubMed]
  3. Raaphorst, G.P.; Freeman, M.L.; Dewey, W.C. Radiosensitivity and recovery from radiation damage in cultured CHO cells exposed to hyperthermia at 42.5 or 45.0 °C. Radiat. Res. 1979, 79, 390–402. [Google Scholar] [CrossRef] [PubMed]
  4. Habash, R.W.Y.; Bansal, R.; Krewski, D.; Alhafid, H. Thermal therapy. Part 2: Hyperthermia techniques. Crit. Rev. Biomed. Eng. 2006, 34, 491–542. [Google Scholar] [CrossRef]
  5. Zhu, L.; Altman, M.B.; Laszlo, A.; Straube, W.; Zoberi, I.; Hallahan, D.E.; Chen, H. Ultrasound Hyperthermia Technology for Radiosensitization. Ultrasound Med. Biol. 2019, 45, 1025–1043. [Google Scholar] [CrossRef]
  6. Gongalsky, M.; Gvindzhiliia, G.; Tamarov, K.; Shalygina, O.; Pavlikov ASolovyev, V.; Kudryavtsev, A.; Sivakov, A.; Osminkina, L.A. Radiofrequency Hyperthermia of Cancer Cells Enhanced by Silicic Acid Ions Released During the Biodegradation of Porous Silicon Nanowires. ACS Omega 2019, 4, 10662–10669. [Google Scholar] [CrossRef]
  7. Altintas, G.; Akduman, I.; Janjic, A.; Yilmaz, T. A Novel Approach on Microwave Hyperthermia. Diagnostics 2021, 11, 493. [Google Scholar] [CrossRef]
  8. Wan Mohd Zawawi WF, A.; Hibma, M.H.; Salim, M.I.; Jemon, K. Hyperthermia by near infrared radiation induced immune cells activation and infiltration in breast tumor. Sci. Rep. 2021, 11, 10278. [Google Scholar] [CrossRef]
  9. Gamal, H.; Tawfik, W.; El-Sayyad, H.I.H.; Emam, A.N.; Mohamed Fahmy, H.; El-Ghaweet, H.A. A New Vision of Photothermal Therapy Assisted with Gold Nanorods for the Treatment of Mammary Cancers in Adult Female Rats. Nanoscale Adv. 2024, 6, 170–187. [Google Scholar] [CrossRef]
  10. Bianchi, L.; Baroni, S.; Paroni, G.; Violatto, M.B.; Moscatiello, G.Y.; Panini, N.; Russo, L.; Fiordaliso, F.; Colombo, L.; Diomede, L.; et al. Thermal Effects and Biological Response of Breast and Pancreatic Cancer Cells Undergoing Gold Nanorod-Assisted Photothermal Therapy. J. Photochem. Photobiol. B Biol. 2024, 259, 112993. [Google Scholar] [CrossRef]
  11. Nag, S.; Mitra, O.; Tripathi, G.; Adur, I.; Mohanto, S.; Nama, M.; Samanta, S.; Gowda, B.H.J.; Subramaniyan, V.; Sundararajan, V.; et al. Nanomaterials-Assisted Photothermal Therapy for Breast Cancer: State-of-the-Art Advances and Future Perspectives. Photodiagnosis Photodyn. Ther. 2024, 45, 103959. [Google Scholar] [CrossRef] [PubMed]
  12. Pacheco, A.R.F.; Cardoso, B.D.; Pires, A.; Pereira, A.M.; Araújo, J.P.; Carvalho, V.M.; Rodrigues, R.O.; Coutinho, P.J.G.; Castelo-Grande, T.; Augusto, P.A.; et al. Development of pH-Sensitive Magnetoliposomes Containing Shape Anisotropic Nanoparticles for Potential Application in Combined Cancer Therapy. Nanomaterials 2023, 13, 1051. [Google Scholar] [CrossRef] [PubMed]
  13. Gilchrist, R.K.; Medal, R.; Shorey, W.D.; Hanselman, R.C.; Parrott, J.C.; Taylor, C.B. Selective inductive heating of lymph nodes. Ann. Surg. 1957, 146, 596–606. [Google Scholar] [CrossRef] [PubMed]
  14. Farzin, A.; Etesami, S.A.; Quint, J.; Memic, A.; Tamayol, A. Magnetic Nanoparticles in Cancer Therapy and Diagnosis. Adv. Healthc. Mater. 2020, 9, e1901058. [Google Scholar] [CrossRef]
  15. Latorre, M.; Rinaldi, C. Applications of magnetic nanoparticles in medicine: Magnetic fluid hyperthermia. Puerto Rico Health Sci. J. 2009, 28, 227–238. [Google Scholar]
  16. Hilger, I. The role of magnetic hyperthermia in heating-based oncological therapies. In Magnetic Materials and Technologies for Medical Applications; Woodhead Publishing: London, UK, 2022; pp. 265–285. [Google Scholar] [CrossRef]
  17. Dutz, S.; Hergt, R. Magnetic nanoparticle heating and heat transfer on a microscale: Basic principles, realities and physical limitations of hyperthermia for tumour therapy. Int. J. Hyperth. 2013, 29, 790–800. [Google Scholar] [CrossRef]
  18. Carrião, M.S.; Aquino, V.R.; Landi, G.T.; Verde, E.L.; Sousa, M.H.; Bakuzis, A.F. Giant-spin nonlinear response theory of magnetic nanoparticle hyperthermia: A field dependence study. J. Appl. Phys. 2017, 121, 173901. [Google Scholar] [CrossRef]
  19. Phong, P.T.; Nguyen, L.H.; Lee, I.-J.; Phuc, N.X. Computer simulations of contributions of Néel and Brown relaxation to specific loss power of magnetic fluids in hyperthermia. J. Electron. Mater. 2017, 46, 2393–2405. [Google Scholar] [CrossRef]
  20. Deatsch, A.E.; Evans, B.A. Heating efficiency in magnetic nanoparticle hyperthermia. J. Magn. Magn. Mater. 2014, 354, 163–172. [Google Scholar] [CrossRef]
  21. Skumiel, A.; Leszczyński, B.; Molcan, M.; Timko, M. The comparison of magnetic circuits used in magnetic hyperthermia. J. Magn. Magn. Mater. 2016, 420, 177–184. [Google Scholar] [CrossRef]
  22. Papadopoulos, C.; Efthimiadou, E.K.; Pissas, M.; Fuentes, D.; Boukos, N.; Psycharis, V.; Kordas, G.; Loukopoulos, V.C.; Kagadis, G.C. Magnetic fluid hyperthermia simulations in evaluation of SAR calculation methods. Phys. Medica 2020, 71, 39–52. [Google Scholar] [CrossRef] [PubMed]
  23. Mazon, E.E.; Sámano, A.H.; Calleja, H.; Quintero, L.H.; Paz, J.A.; Cano, M.E. A frequency tuner for resonant inverters suitable for magnetic hyperthermia applications. Meas. Sci. Technol. 2017, 28, 095901. [Google Scholar] [CrossRef]
  24. Griffiths, D.J. Introduction to Electrodynamics: Pearson New International, 4th ed.; Pearson Education, Inc.: London, UK, 2013. [Google Scholar]
  25. Makarova, M.V.; Suzuki, K.; Kon, H.; Dubey, P.; Sonobe, H.; Matsumura, T.; Saito, H. High-Frequency Magnetic Field Energy Imaging of Magnetic Recording Head by Alternating Magnetic Force Microscopy (A-MFM) with Superparamagnetic Tip. Appl. Sci. 2023, 13, 4843. [Google Scholar] [CrossRef]
  26. Kallumadil, M.; Tadac, M.; Nakagawac, T.; Abec, M.; Southern, P.; Pankhurst, Q.A. Suitability of commercial colloids for magnetic hyperthermia. J. Magn. Magn. Mater. 2009, 321, 1509–1513. [Google Scholar] [CrossRef]
  27. Angelakeris, M.; Li, Z.A.; Sakellari, D.; Simeonidis, K.; Spasova, M.; Farle, M. Can commercial ferrofluids be exploited in AC magnetic hyperthermia treatment to address diverse biomedical aspects? EPJ Web Conf. 2014, 75, 08002. [Google Scholar] [CrossRef]
  28. Lanier, O.L.; Korotych, O.I.; Monsalve, A.G.; Wable, D.; Savliwala, S.; Grooms, N.W.; Nacea, C.; Tuitt, O.R.; Dobson, J. Evaluation of magnetic nanoparticles for magnetic fluid hyperthermia. Int. J. Hyperth. 2019, 36, 686. [Google Scholar] [CrossRef]
  29. Sakellari, D.; Mathioudaki, S.; Kalpaxidou, Z.; Simeonidis, K.; Angelakeris, M. Exploring multifunctional potential of commercial ferrofluids by magnetic particle hyperthermia. J. Magn. Magn. Mater. 2015, 380, 360. [Google Scholar] [CrossRef]
  30. Yaremenko, A.V.; Zelepukin, I.V.; Ivanov, I.N.; Melikov, R.O.; Pechnikova, N.A.; Dzhalilova, D.S.; Mirkasymov, A.B.; Bragina, V.A.; Nikitin, M.P.; Deyev, S.M.; et al. Influence of magnetic nanoparticle biotransformation on contrasting efficiency and iron metabolism. J. Nanobiotechnol. 2022, 20, 535. [Google Scholar] [CrossRef]
  31. Wang, S.-Y.; Huang, S.; Borca-Tasciuc, D.-A. Potential Sources of Errors in Measuring and Evaluating the Specific Loss Power of Magnetic Nanoparticles in an Alternating Magnetic Field. IEEE Trans. Magn. 2013, 49, 255–262. [Google Scholar] [CrossRef]
  32. Dabaghi, M.; Quaas, R.; Hilger, I. The Treatment of Heterotopic Human Colon Xenograft Tumors in Mice with 5-Fluorouracil Attached to Magnetic Nanoparticles in Combination with Magnetic Hyperthermia Is More Efficient than Either Therapy Alone. Cancers 2020, 12, 2562. [Google Scholar] [CrossRef]
  33. Jordan, A.; Scholz, R.; Maier-Hauff, K.; Johannsen, M.; Wust, P.; Nadobny, J.; Schirra, H.; Schmidt, H.; Deger, S.; Loening, S.; et al. Presentation of a New Magnetic Field Therapy System for the Treatment of Human Solid Tumors with Magnetic Fluid Hyperthermia. J. Magn. Magn. Mater. 2001, 225, 118–126. [Google Scholar] [CrossRef]
  34. Tai, C.-C.; Chen, M.-K. A compact half-bridge induction heating system for magnetic nanoparticle thermotherapy applications, Biomedical Engineering: Applications. Basis Commun. 2007, 19, 27–35. [Google Scholar] [CrossRef]
  35. Cabrera, D.; Rubia-Rodríguez, I.; Garaio, E.; Plazaola, F.; Dupré, L.; Farrow, N.; Terán, F.J.; Ortega, D. Chapter 5—Instrumentation for Magnetic Hyperthermia. In Micro and Nano Technologies, Nanomaterials for Magnetic and Optical Hyperthermia Applications; Fratila, R.M., De La Fuente, J.M., Eds.; Elsevier: Amsterdam, The Netherlands, 2019; pp. 111–138. ISBN 9780128139288. [Google Scholar] [CrossRef]
  36. Rashid, M.H.; Hui, S.Y.; Shu-Hung Chung, H. Power Electronics Handbook, 4th ed.; Rashid, M.H., Ed.; Butterworth-Heinemann: Oxford, UK, 2018. [Google Scholar] [CrossRef]
  37. Truong, C.-T.; Choi, S.-J. Improved ZVS Criterion for Series Resonant Converters. IEEE Access 2024, 12, 5333–5344. [Google Scholar] [CrossRef]
  38. Skumiel, A.; Hornowski, T.; Józefczak, A.; Koralewski, M.; Leszczyński, B. Uses and limitation of different thermometers for measuring heating efficiency of magnetic fluids. Appl. Therm. Eng. 2016, 100, 1308–1318. [Google Scholar] [CrossRef]
  39. Castelo-Grande, T.; Augusto, P.A.; Gomes, L.; Lopes AR, C.; Araújo, J.P.; Barbosa, D. Economic and Accessible Portable Homemade Magnetic Hyperthermia System: Influence of the Shape, Characteristics and Type of Nanoparticles in Its Effectiveness. Materials 2024, 17, 2279. [Google Scholar] [CrossRef]
  40. Pinakidou, F.; Simeonidis, K.; Myrovali, E.; Brzhezinskaya, M.; Paloura, E.C.; Angelakeris, M.; Katsikini, M. Addressing the effect of magnetic particle hyperthermia application on the composition and spatial distribution of iron oxide nanoparticles using X-ray spectroscopic techniques. J. Phys. Chem. C 2022, 126, 10101–10109. [Google Scholar] [CrossRef]
  41. Ring, H.L.; Sharma, A.; Ivkov, R.; Bischof, J.C. The impact of data selection and fitting on SAR estimation for magnetic nanoparticle heating. Int. J. Hyperth. 2022, 37, 100–107. [Google Scholar] [CrossRef]
  42. Espinosa DCarlsson, L.B.; Figueiredo Neto, A.M. Influence of nanoparticle size on the nonlinear optical properties of magnetite ferrofluids. Phys. Rev. 2013, 88, 032302. [Google Scholar] [CrossRef]
  43. Shah, R.R.; Davis, T.P.; Glover, A.L.; Nikles, D.E.; Brazel, C.S. Impact of magnetic field parameters and iron oxide nanoparticle properties on heat generation for use in magnetic hyperthermia. J. Magn. Magn. Mater. 2015, 387, 96–106. [Google Scholar] [CrossRef]
  44. Eggeman, A.S.; Majetich, S.A.; Farrell, D.; Pankhurst, Q.A. Size and Concentration Effects on High Frequency Hysteresis of Iron Oxide Nanoparticles. IEEE Trans. Magn. 2007, 43, 2451–2453. [Google Scholar] [CrossRef]
  45. Sakellari, D.; Brintakis, K.; Kostopoulou, A.; Myrovali, E.; Simeonidis, K.; Lappas, A.; Angelakeris, M. Ferrimagnetic nanocrystal assemblies as versatile magnetic particle hyperthermia mediators. Mater. Sci. Eng. C 2016, 58, 187–193. [Google Scholar] [CrossRef] [PubMed]
  46. Rubia-Rodríguez, I.; Santana-Otero, A.; Spassov, S.; Tombácz, E.; Johansson, C.; De La Presa, P.; Teran, F.J.; Morales, M.d.P.; Veintemillas-Verdaguer, S.; Thanh, N.T.K.; et al. Whither Magnetic Hyperthermia? A Tentative Roadmap. Materials 2021, 14, 706. [Google Scholar] [CrossRef] [PubMed]
  47. Lemine, O.M.; Algessair, S.; Madkhali, N.; Al-Najar, B.; El-Boubbou, K. Assessing the Heat Generation and Self-Heating Mechanism of Superparamagnetic Fe3O4 Nanoparticles for Magnetic Hyperthermia Application: The Effects of Concentration, Frequency, and Magnetic Field. Nanomaterials 2023, 13, 45. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Three-dimensional view of solenoid S5. (a) The copper coil contains interior water for cooling inside, while the surrounding domain contains air (b) computational mesh.
Figure 1. Three-dimensional view of solenoid S5. (a) The copper coil contains interior water for cooling inside, while the surrounding domain contains air (b) computational mesh.
Nanomaterials 14 01848 g001
Figure 2. Initial slope method (ISM) for calculating the specific absorption rate (SAR).
Figure 2. Initial slope method (ISM) for calculating the specific absorption rate (SAR).
Nanomaterials 14 01848 g002
Figure 3. Parallel LC resonant circuit.
Figure 3. Parallel LC resonant circuit.
Nanomaterials 14 01848 g003
Figure 4. Image of the switch set that controls the capacitor bank.
Figure 4. Image of the switch set that controls the capacitor bank.
Nanomaterials 14 01848 g004
Figure 5. (a) Initial prototype of our system, (b) the system being currently applied, (c) a new and more powerful prototype system (resonant) that is being tested.
Figure 5. (a) Initial prototype of our system, (b) the system being currently applied, (c) a new and more powerful prototype system (resonant) that is being tested.
Nanomaterials 14 01848 g005
Figure 6. Overview of the system (a) and photographs of the apparatus and of the system (b): A—power source; B—system; C—solenoid; D—cooling system; E—equipment for measuring temperature; F—computer for acquiring and registering the temperature; G and H—oscilloscope coupled with a probe for measuring the magnetic field and verifying the waveform passing through the coil [39].
Figure 6. Overview of the system (a) and photographs of the apparatus and of the system (b): A—power source; B—system; C—solenoid; D—cooling system; E—equipment for measuring temperature; F—computer for acquiring and registering the temperature; G and H—oscilloscope coupled with a probe for measuring the magnetic field and verifying the waveform passing through the coil [39].
Nanomaterials 14 01848 g006aNanomaterials 14 01848 g006b
Figure 7. Insulating system.
Figure 7. Insulating system.
Nanomaterials 14 01848 g007
Figure 8. Images of solenoids described in Table 4.
Figure 8. Images of solenoids described in Table 4.
Nanomaterials 14 01848 g008
Figure 9. Profile of the magnetic field density (mT) of solenoid coil S5 for a driving voltage of 60 V and an operating frequency of 72 kHz. (a) Three-dimensional view, (b) two-dimensional view for the plane x = 0.
Figure 9. Profile of the magnetic field density (mT) of solenoid coil S5 for a driving voltage of 60 V and an operating frequency of 72 kHz. (a) Three-dimensional view, (b) two-dimensional view for the plane x = 0.
Nanomaterials 14 01848 g009
Figure 10. Magnetic field density (mT) along the center of the simulated coils at various operating frequencies.
Figure 10. Magnetic field density (mT) along the center of the simulated coils at various operating frequencies.
Nanomaterials 14 01848 g010
Figure 11. The temperature variation of dstillated water samples was consistent across three experiments conducted on different days at a frequency of 69 kHz with solenoid S5.
Figure 11. The temperature variation of dstillated water samples was consistent across three experiments conducted on different days at a frequency of 69 kHz with solenoid S5.
Nanomaterials 14 01848 g011
Figure 12. Heating curves for the FluidmagARA and FluidmagUCA samples at a frequency 98 kHz using solenoid S5.
Figure 12. Heating curves for the FluidmagARA and FluidmagUCA samples at a frequency 98 kHz using solenoid S5.
Nanomaterials 14 01848 g012
Figure 13. Heating curves of FluidmagD100nm (25 mg/mL) at 69 kHz using solenoid S5.
Figure 13. Heating curves of FluidmagD100nm (25 mg/mL) at 69 kHz using solenoid S5.
Nanomaterials 14 01848 g013
Figure 14. Heating curves of FluidmagD50nm (25 mg/mL) at 69 kHz using solenoid S5.
Figure 14. Heating curves of FluidmagD50nm (25 mg/mL) at 69 kHz using solenoid S5.
Nanomaterials 14 01848 g014
Figure 15. Heating curves for the FluidMagD100nm sample at various frequencies using solenoid S5.
Figure 15. Heating curves for the FluidMagD100nm sample at various frequencies using solenoid S5.
Nanomaterials 14 01848 g015
Figure 16. Heating curves of various samples at a frequency of 138 kHz.
Figure 16. Heating curves of various samples at a frequency of 138 kHz.
Nanomaterials 14 01848 g016
Figure 17. Heating curves of FluidMagDx100nn at different frequencies with a concentration of 12.5 mgMNP/mL using solenoid S6.
Figure 17. Heating curves of FluidMagDx100nn at different frequencies with a concentration of 12.5 mgMNP/mL using solenoid S6.
Nanomaterials 14 01848 g017
Figure 18. Heating curves of FluidMagUCA at 72 kHz with a concentration of 12.5 mgMNP/mL using solenoid S8.
Figure 18. Heating curves of FluidMagUCA at 72 kHz with a concentration of 12.5 mgMNP/mL using solenoid S8.
Nanomaterials 14 01848 g018
Figure 19. Heating curves of FluidMagD100nm at different frequencies with a concentration of 12.5 mgMNP/mL using solenoid S8.
Figure 19. Heating curves of FluidMagD100nm at different frequencies with a concentration of 12.5 mgMNP/mL using solenoid S8.
Nanomaterials 14 01848 g019
Figure 20. Heating curves of FluidMagDx50nm at different frequencies with a concentration of 25 mgMNP/mL using solenoid S8.
Figure 20. Heating curves of FluidMagDx50nm at different frequencies with a concentration of 25 mgMNP/mL using solenoid S8.
Nanomaterials 14 01848 g020
Figure 21. Heating curves of FluidmagDX50nm samples with different concentrations (5 and 25 mg/mlMNP) at 132 kHz using solenoid S8.
Figure 21. Heating curves of FluidmagDX50nm samples with different concentrations (5 and 25 mg/mlMNP) at 132 kHz using solenoid S8.
Nanomaterials 14 01848 g021
Figure 22. Heating and cooling curves of FluidMagDx50nm with the final configuration (S8).
Figure 22. Heating and cooling curves of FluidMagDx50nm with the final configuration (S8).
Nanomaterials 14 01848 g022
Table 1. Advantages and disadvantages of magnetic hyperthermia [14,15,16,17,18].
Table 1. Advantages and disadvantages of magnetic hyperthermia [14,15,16,17,18].
AdvantagesDisadvantages
Magnetic nanoparticles are absorbed by cancer cells, allowing for localized therapeutic heat supply, which increases hyperthermia’s effectivenessMagnetic hyperthermia encounters challenges in enhancing nanoparticle heating power, regulating tumor temperature
Tagging MNPs with tumor-specific binding agents ensures targeted and efficient treatment, maximizing its impact.It is difficult to minimize adverse effects on nearby healthy tissues.
Harmless passage of alternating magnetic field frequencies through the body exclusively generates heat in MNP-containing tissues, ensuring safe and precise treatment.It is necessary to monitor temperature changes at the cellular level using precise and non-invasive techniques.
MNPs’ ability to traverse the blood–brain barrier makes them valuable for treating brain tumors.It is crucial to understand the impact of temperature on the biological processes of cells.
MNPs can be used to create stable colloids, allowing for a variety of drug delivery routes.It is important to comprehend the factors that impact heat transport from MNPs to cells.
Table 2. Properties for magnetostatic simulations.
Table 2. Properties for magnetostatic simulations.
MaterialRelative Permeability
μ r
Relative Permeability
ε r
Electric Conductivity
σ (S/m)
Air110
Copper116 × 107
Water110
Table 3. Properties of the magnetic nanoparticles used in the experiments.
Table 3. Properties of the magnetic nanoparticles used in the experiments.
ProductMatrix/CoverSize (Hydrodynamic
Diameter) (nm)
Density
(g/cm3)
FunctionalizationParticles NumberCore/vh * (nm)Zeta Potential
(mv)
Fluidmag-araPolysaccharide150~1.25Glucuronic acid~1.8 × 1015/g −2.2 × 1014/g23.6 ± 5.6/
121 ± 40
[30]
−22.65 mV
[30]
Fluidmag-ucaNo200~5.2Anionic charge~2.2 × 1014/g
Fluidmag-dxDextran50~1.25Hydroxyl groups~1.3 × 1016/g
Fluidmag-cmxCarboxymethyl–dextrann.a.~1.25Sodium carboxylate 110.6 ± 3.5
[32]
–32.3 ± 0.1
Fluidmag-lipidPhosphatidylcholine200~1.25Phosphatidylcholine~2.2 × 1014/g
Fluidmag-d50Starch50~1.25Hydroxyl groups~1.3 × 1016/g
Fluidmag-d100Starch100~1.25Hydroxyl groups~1.8 × 1015/g
SupermagneticARA, coated with polysaccharides, distilled water solvent (diam. = 150 nnm, conc = 25 mg/mL and 1.8 × 1015 particles/g; d = 1.25 g/cm3). SupermagneticUCA (diam. = 200 nm, conc = 25 mg/mL and 2.2 × 1014 particles/g; d = 5.2 g/cm3), magnetite as core and uncoated, and are charged to be anionic, solvent distilled water. * Other average values published CMX-coatedMNPs 0.98 ± 0.11 content of iron mgFe/mL.
Table 4. Parameters of the applied solenoids.
Table 4. Parameters of the applied solenoids.
SolenoidH (Height)
(mm)
Tube External Diameter
(mm)
Tube
Internal
Diameter (mm)
Solenoid
Internal Diameter
(mm)
Number
of Turns
Space
Between
Turns (mm)
Observations
S133.04.1----54.65 2.7
S2100.76.3173.713 2.0Inductance is 5 μH
S368.06.3150.09 2.0
S466.06.3174.58 Variable
S573.05.0170.09 2–2.5 variable
S657.06.3141.37 VariableInductance is 2.5 μH
S7100.0n.a.---70.011 Variable Foldable (multiwire)
S8119.65.0142.317 VariableYield the best results so far
Table 5. Electrical properties of solenoid at various frequencies.
Table 5. Electrical properties of solenoid at various frequencies.
f(kHz)Coil Current (A) Coil Resistance (Ω)Coil Impedance (Ω)Coil Inductance (μH)
S57233.684∠−89.305°2.160 × 10−21.781∠89.305°3.9372
9625.311∠−89.403°2.470 × 10−22.37∠89.403°3.9296
13418.167∠−89.512°2.811 × 10−23.303∠89.512°3.9226
3028.081∠−89.739°3.377 × 10−27.425∠89.739°3.9127
S672122.183∠−88.89°9.512 × 10−30.491∠88.89°1.0853
9691.917∠−89.046°1.086 × 10−20.653∠89.046°1.0820
13466.04∠−89.22°1.237 × 10−20.909∠89.22°1.0790
30229.423∠−89.58°1.495 × 10−22.039∠89.58°1.0747
S8—12 turns7261.764∠−88.943°1.791 × 10−20.971∠88.943°2.1470
9646.459∠−89.101°2.026 × 10−21.291∠89.101°2.1408
13433.372∠−89.277°2.268 × 10−21.798∠89.277°2.1353
30214.858∠−89.626°2.637 × 10−24.038∠89.626°2.1280
S8—17 turns7243.316∠−88.959°2.516 × 10−21.385∠88.959°3.0614
9632.58∠−89.116°2.841 × 10−21.842∠89.116°3.0528
13423.402∠−89.29°3.178 × 10−22.564∠89.29°3.0450
30210.418∠−89.634°3.676 × 10−25.759∠89.634°3.0349
Table 6. Results of the final temperature respective deviations and conditions obtained in the experiences.
Table 6. Results of the final temperature respective deviations and conditions obtained in the experiences.
Type of ParticleFrequency (kHz)SolenoidTFinal (C°)Figure
FluidmagARA98S527.5 ± 0.412
FluidmagUCA98S523.5 ± 0.512
FluidMagD100nm69S531.0 ± 0.513
FluidMagD50nm69S526.2 ± 0.514
FluidMagD100nm69S531.5 ± 0.414
FluidMagD100nm63S531.0 ± 0.315
FluidMagD100nm78S533.8 ± 0.315
FluidMagD100nm81S525.5 ± 0.4 +15
FluidMagD100nm138S528.5 ± 0.4 ++15
FluidMagCMX138S527.7 ± 0.5 *16
FluidMagLip200nm138S524.4 ± 0.4 **16
FluidMagDx100nn138S531.80.5 +*16
FluidMagDX50nm138S526.9 ± 0.5 +**16
FluidMagDx100nn a110S627.0 ± 0.517
FluidMagDx100nn a141S627.0 ± 0.417
FluidMagDx100nn a170S627.5 ± 0.417
FluidMagDx100nn a235S629.7 ± 0.317
FluidMagUCA a72 S835.5 ± 0.418
FluidMagDx100nm71S828.7 ± 0.419
FluidMagDx100nm96S827.1 ± 0.519
FluidMagDx100nm302S827.3 ± 0.519
FluidMagDx50nm99S834.3 ± 0.420
FluidMagDx50nm132S836.8 ± 0.420
FluidMagDx50nm302S831.6 ± 0.4 ++**20
FluidMagDx50nm b132S823 ± 0.321
FluidMagDx50nm132S840.9 ± 0.21
FluidMagD50nm101.5 kHzS852.2 ± 0.422
+ At 10 min, ++ at 13 min.; * 10 min; ** 12 min; +* 13 min; +** 9 min; a concentration of 12.5 mgMNP/mL b concentration of 5 mg/mL; ++** The study was stopped for 302 kHz 31.6 in 7 min since at the same time it obtained 36.8 (132 kHz) and 34.3 (99 kHz); b 5 mg/mL.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Castelo-Grande, T.; Augusto, P.A.; Gomes, L.; Calvo, E.; Barbosa, D. Portable Homemade Magnetic Hyperthermia Apparatus: Preliminary Results. Nanomaterials 2024, 14, 1848. https://doi.org/10.3390/nano14221848

AMA Style

Castelo-Grande T, Augusto PA, Gomes L, Calvo E, Barbosa D. Portable Homemade Magnetic Hyperthermia Apparatus: Preliminary Results. Nanomaterials. 2024; 14(22):1848. https://doi.org/10.3390/nano14221848

Chicago/Turabian Style

Castelo-Grande, Teresa, Paulo A. Augusto, Lobinho Gomes, Eduardo Calvo, and Domingos Barbosa. 2024. "Portable Homemade Magnetic Hyperthermia Apparatus: Preliminary Results" Nanomaterials 14, no. 22: 1848. https://doi.org/10.3390/nano14221848

APA Style

Castelo-Grande, T., Augusto, P. A., Gomes, L., Calvo, E., & Barbosa, D. (2024). Portable Homemade Magnetic Hyperthermia Apparatus: Preliminary Results. Nanomaterials, 14(22), 1848. https://doi.org/10.3390/nano14221848

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop