Review

Medical Devices for Tremor Suppression: Current Status and Future Directions

Massachusetts College of Pharmacy and Health Sciences University, School of Pharmacy, Boston, MA 02115, USA
*
Author to whom correspondence should be addressed.
Biosensors 2021, 11(4), 99; https://doi.org/10.3390/bios11040099
Submission received: 22 February 2021 / Revised: 18 March 2021 / Accepted: 24 March 2021 / Published: 30 March 2021

Abstract

:
Tremors are the most prevalent movement disorder that interferes with the patient’s daily living, and physical activities, ultimately leading to a reduced quality of life. Due to the pathophysiology of tremor, developing effective pharmacotherapies, which are only suboptimal in the management of tremor, has many challenges. Thus, a range of therapies are necessary in managing this progressive, aging-associated disorder. Surgical interventions such as deep brain stimulation are able to provide durable tremor control. However, due to high costs, patient and practitioner preference, and perceived high risks, their utilization is minimized. Medical devices are placed in a unique position to bridge this gap between lifestyle interventions, pharmacotherapies, and surgical treatments to provide safe and effective tremor suppression. Herein, we review the mechanisms of action, safety and efficacy profiles, and clinical applications of different medical devices that are currently available or have been previously investigated for tremor suppression. These devices are primarily noninvasive, which can be a beneficial addition to the patient’s existing pharmacotherapy and/or lifestyle intervention.

1. Introduction

Tremors, as defined by the task force of the International Parkinson and Movement Disorder Society (IPMDS), are an involuntary, rhythmic, oscillatory movement of a body part [1]. Essential tremor (ET) is recognized as the most prevalent pathological tremor among adults, affecting about 0.9% of the global population [2]. However, the true prevalence of ET may be higher, as it is believed that these patients may not seek medical attention [3]. Tremors, usually asymmetrically distributed, are frequently seen in patients with Parkinson’s disease (PD), which affects more than six million individuals worldwide [4]. The presence of resting tremor supports the diagnosis of PD [5]. Different clinical subtypes and classifications of tremor disorders have also been identified [1]. The etiologies of tremor include other neurodegenerative diseases such as Wilson’s disease, chromosomal aneuploidy, mitochondrial genetic disorders, infectious and inflammatory diseases, endocrine and metabolic disorders, neuropathies and spinal muscular atrophies, toxin-/drug-induced tremor pathology, and brain neoplasms and injury, as well as several environmental causes [1].
Tremors impact many aspects of the patient’s daily living and interfere with many physical activities at home and in the workplace [6,7,8,9,10]. One clinical-epidemiological study compared the quality of life, including physical and psychosocial aspects, between patients with ET and PD using the Quality of Life in Essential Tremor (QUEST) questionnaire [11]. Patients with ET had a higher QUEST total score and QUEST physical subscore than patients with PD (p < 0.05). This suggests that patients with ET suffers significantly more physical and psychosocial impairment than those with PD [11]. Additionally, among patients suffering from tremor, their psychological strain may be significantly more affected than their physical disabilities [6,12]. The psychological toll of tremor may extend beyond the patients themselves. The Clinical Pathological Study of Cognitive Impairment in Essential Tremor (COGNET), a longitudinal study that evaluates cognitive function in older adults with ET, reported that both patients with ET and those close to them suffer psychological stress [13]. In addition, patients may develop feelings of social isolation [11,14] and depression [6,11,13]. Due to the incredible burden put on individuals diagnosed with ET or PD, a multitude of approaches have been investigated to improve the symptoms and quality of life of those afflicted. These range from lifestyle interventions, pharmacotherapy, and surgical treatments.
Lifestyle interventions focusing on the use of weighted utensils can reduce the amplitude of tremor and alleviate the challenges patients face in their activities of daily living (ADLs) [15,16]. With additional weights, these utensils (e.g., spoon) can assist patients to eat and drink. In 2017, the National Institute for Health and Care Excellence (NICE) produced guidelines for the management of PD in adults [5]. Patients in the early stages of PD may benefit from physio- and occupational therapy if they experience motor symptoms or have difficulties with ADLs [5]. However, lifestyle and the nonpharmacological management of ET were not discussed in the guidelines produced by the American Academy of Neurology (AAN) and the IPMDS [17,18,19]. A systematic review of 19 studies found that physical therapy, limb cooling, vibration therapy, use of limb weights, bright light therapy, and transcranial magnetic stimulation were all examples of investigated treatments of tremor [20]. However, these studies mainly included convenience samples, and the long-term effectiveness of these interventions was not assessed [20].
Pharmacotherapy for the treatment of ET is suboptimal and only treats the symptoms. Many patients do not respond to the existing medications indicated for ET and do not experience a significant improvement in their daily living. Currently, propranolol and primidone are the two first-line therapies [15,16,17,18,19,21]. Across randomized controlled trials (RCTs), propranolol and primidone monotherapy produce a mean reduction in the tremor amplitude of 54.1% and 59.9%, respectively, as measured by accelerometry [22]. Nonetheless, 56.3% of patients eventually discontinued the use of either medications [23]. Topiramate is also recommended as a first-line therapy by the guidelines of the Italian Movement Disorders Association (IMDA) [24] and is considered clinically useful at higher doses by the IPMDS task force [19]. However, it is recommended by the AAN guidelines as a second-line therapy [17,18]. Second-line medications have been reported to be less efficacious in reducing the amplitude of tremors. These include alprazolam, atenolol, gabapentin, and sotalol, as well as the aforementioned topiramate [17,18]. In contrast, there is no consensus in the management of PD tremors. The current NICE guidelines recommend levodopa as the first-line therapy for management of all motor symptoms in patients in the early stages of PD [5].
Deep brain stimulation (DBS), whose efficacy has been demonstrated through closed loop approaches [25,26] and interleaving stimulation [27], is the most common surgical treatment to date, providing durable tremor control, especially for patients with medically refractory ET or advanced PD. The effectiveness of DBS in ET and PD tremor is thought to be due to the direct electrical stimulation to the ventral intermediate nucleus (VIM) possibly disrupting the synchronous firing of thalamic neurons [28,29]. In addition to the VIM, the subthalamic nucleus, internal globus pallidus, and pedunculopontine nucleus are also effective targets for DBS in patients with PD tremors [30]. The use of DBS was approved by the Food and Drug Administration (FDA) for ET in 1997, for advanced PD in 2002, and for mid-stage PD in 2016. As of late, radiofrequency thalamotomy has become less favored. An RCT comparing DBS with thalamotomy in 68 patients with tremor due to ET, PD, or multiple sclerosis found that DBS results in fewer adverse effects (p = 0.024) and a greater increase in the Frenchay Activities Index score, which assess 15 ADLs. This suggests a greater improvement in the functional status when compared to thalamotomy [31]. Although surgical treatments for tremors, including DBS, stereotactic radiosurgery (SRS), and magnetic resonance-guided focused ultrasound (MRgFUS), are more efficacious than pharmacotherapy [32], the utilization of these procedures remains low. Limiting factors may include high surgical costs [33,34], access to care [35,36], and patient preference [35]. Other perceived barriers to DBS include practitioner preference [34,37], high resource and labor intensity [34,38], and perceptions of serious surgical risk [34,38,39].
Thus, a growing unmet need for safe and effective tremor control and suppression sets the stage for a range of therapies to bridge this gap between lifestyle modifications, pharmacotherapy, and surgical treatment. Using a variety of noninvasive suppression mechanisms, medical devices fit within this gap to provide effective tremor suppression at a lower risk than surgery. The increasing interest in this area has led to the birth of a new classification of external upper limb tremor stimulators. In 2018, the de novo classification request of Cala ONE (Cala Health, Burlingame, CA, USA) received FDA approval [40].
Herein, we focus on the mechanisms of action, safety and efficacy profiles, and clinical applications of different categories of medical devices that are available clinically or previously investigated for tremor suppression. Furthermore, we highlight the limitations of these devices. Such information may then be translated biomechanically and clinically for potential future advancements of medical device for tremor suppression.

2. Early Innovations

Over the past several decades, a variety of different orthotic and stimulatory approaches has been proposed to target or reverse the abnormal rhythmic activities in the neural pathways of the cerebellum and the thalamus. Beginning 1987, Rosen and colleagues proposed several devices that employed energy dissipation to suppress tremors. The damped joystick is a hand control device designed to facilitate the control of wheelchairs and other applications [41,42,43,44]. This device consists of a sealed chamber filled with viscous fluid and a spherical ball that acts as a damping element to suppress the involuntary movements of the position-sensing actuator. The controlled energy-dissipation orthosis (CEDO) is a wheelchair-mounted device that provides velocity-dependent loading with magnetic particle brakes to a limb coupling cuff [45,46,47]. Similarly, the modulated energy dissipation (MED) manipulator also provides damping via magnetic particle brakes with real-time digital control [48,49].
The success of these early works in showing that velocity-dependent loading can attenuate tremor and involuntary motions led to the development of wearable orthosis with mechanical loading. Other approaches, including electrical stimulation systems and assistive feeding devices, have also been proposed. Most of these are classified as Class I medical devices, meaning that they are registered with the FDA but not subjected to any premarketing review.

3. Electrical Stimulation Systems

3.1. Median and Radial Nerve Excitation

High frequency transcutaneous electrical nerve stimulation (TENS) has been widely studied and used in the treatment of nociceptive and neuropathic pain [50,51,52,53]. The use of TENS in the treatment of movement disorders, including myoclonic dystonia and ET, was first explored by Toglia and Izzo in 1985 [54]. While the exact mechanism of TENS remains unclear, putative mechanisms focus on its ability to modulate the afferent transmission of sensory information from the periphery to the central nervous system (CNS) [50]. Conventional TENS intends to selectively stimulate the large, myelinated peripheral proprioceptive A-beta (Aβ) sensory fibers [50]. The excitation of the Aβ fibers reduces the transmission of the sensory signals elicited by noxious stimulus, thereby reducing the pain perception [55,56,57,58]. These Aβ fibers carry proprioceptive sensory information into the thalamic circuits that are hypothesized to be involved in tremor generation [59]. Most [54,60,61,62,63], but not all [64], studies suggest that treatment with TENS in patients who have tremors was associated with improved muscle strength and tremor reduction. However, sham-controlled randomized trials are needed to confirm these findings due to potential confounding effects associated with the reason for use.
In 2018, Cala ONE was the first wearable transcutaneous electrical nerve stimulator to be approved by the FDA [40]. The newer version of this device, Cala Trio (Cala Health, USA; previously known as Cala TWO), is currently FDA-registered. The PROspective study for SymPtomatic relief of Essential tremor with Cala Therapy (PROSPECT) pivotal trial for Cala Trio was completed in 2019 [65], but it is still waiting for approval by the FDA. Clinically, Cala Trio is designed to replace Cala ONE for use in the transient, symptomatic relief of hand tremors in adults with ET. This device can be worn for therapy on the left or right wrist.
Cala Trio involves two working electrodes positioned over the median and radial nerves on the anterior surface of the wrist and a counter electrode placed on the posterior surface of the wrist. An accelerometer within this device measures the frequency of the patient’s tremor, allowing individualized calibration of the stimulation intensity. The two working electrodes deliver electrical signals that intermittently excite the median and radial nerves in the upper limbs. Peripheral sensory nerves, including the median and radial nerves, also project to the VIM and the neural circuits that are implicated in ET. Similar to DBS, electrical stimulation of the VIM peripherally via the median [66,67] and radial [68] nerves elicits very fast oscillations, which induce thalamicneuronal oscillations and disrupt the pathological oscillations of tremors (Figure 1). A study involving five patients with tremors due to ET or PD demonstrated that electrical stimulation of the median and radial nerves leads to a 57% tremor suppression (p < 0.01) [69]. Over time, this stimulation with Cala Trio aims to normalize the neural firing in the pathological tremor network in the CNS to reduce tremors.
The pivotal trial for Cala ONE, a sham-controlled randomized trial of a single 40-min TENS session among 77 patients with ET, found no significant improvements in the Archimedes spiral task, as measured using the Tremor Research Group Essential Tremor Rating Assessment Scale (TETRAS) (p = 0.26) [70]. However, the Cala ONE stimulation did show significantly improved upper limb TETRAS tremor scores (p = 0.017) and subject-rated Bain and Findley ADL scores (p = 0.001), corresponding to a 42% (versus 28% with sham) and a 49% (versus 27% with sham) reduction in tremor amplitude, respectively [70]. The PROSPECT pivotal trial for Cala Trio, an open-label study of TENS treatment in adults with ET, compared twice-daily home therapy TENS sessions over a three-month period among 263 patients [71]. The results, based on 205 patients who completed the study, showed that TENS treatment via Cala Trio resulted in significant improvements in both the TETRAS and subject-rated Bain and Findley ADL scores (p < 0.0001) [71]. Among the 193 patients included in the secondary analysis, 54% experienced a ≥50% reduction in tremor amplitude [71]. However, 14 patients did not respond to the therapy, suggesting that not all patients with ET will benefit from Cala Trio [71]. It is important to note that the open label, single-arm design of the PROSPECT trial limits the generality of Cala Trio’s effect; therefore, future studies with more robust designs (e.g., RCTs) would be valuable to assess its efficacy.
Device-related adverse events were mild to moderate in severity. Nonserious adverse events were observed in 18% of patients, including skin irritations (redness, itchiness, and/or swelling); soreness or lesions; and discomfort (stinging and/or sensation of weakness) or burns [71]. These adverse events were all resolved with the use of a topical ointment, decreased stimulation intensity, or discontinued therapy [71]. Contraindications to the use of Cala Trio include having currently implanted electrical medical device (e.g., pacemaker, defibrillator, and deep brain stimulator), suspected or diagnosed epilepsy or other seizure disorders or pregnancy. This device should also not be applied on skin eruptions, open wounds, cancerous lesions, or swollen/infected/inflamed areas.
A 50% reduction of tremor amplitude is comparable to the first-line propranolol and primidone pharmacotherapies [17], which are considered clinically useful for the treatment of ET [19]. Cala Trio can play an important role in patients who are not eligible for surgical intervention or do not respond to pharmacotherapy. It has a similar, favorable safety profile to Cala ONE, whose risk and benefit determination met the FDA’s requirements. This device is noninvasive, with 85% of patients reporting its convenience and ease of use [71]. Currently, it is uncertain whether Cala Trio could reduce or replace the need of medications in the treatment of ET. Thus, physicians need to evaluate how it will fit along with pharmacotherapy and/or lifestyle interventions for each patient with the consideration of tremor severity. Post-approval studies could address this question and provide further insights into the long-term safety and efficacy of Cala Trio.

3.2. Antagonistic Muscles Activation

In contrast to TENS, which stimulates sensory nerves, functional electrical stimulation (FES) provides stimulation to motor nerves to trigger muscle contraction. FES for tremor suppression was pioneered by Prochazka and colleagues in 1989 [72,73] and clinically assessed in 1992 [74,75]. Briefly, FES was associated with a tremor suppression of 73% in ET, 62% in PD tremors, and 38% in cerebellar tremors [75]. The recognized limitations of these early works included the potentially unreliable placement of surface electrodes, which could lead to insufficient tremor suppression [74,75]. Although the implantation of percutaneous intramuscular electrodes could solve this problem, this approach is invasive and reserved for patients with severe tremor [64]. Nonetheless, the results from these pilot works led to the first functional electrical stimulator developed specifically for tremor suppression [76,77]. Comparing the previous approach using an analog filter [72,73,74,75], the use of an optimized digital filter [76,77] in a portable functional electrical stimulator, with the enabled self-tuning and adaptation of more complex algorithms, showed improvements in suppressing tremors. In six participants who were healthy or with PD tremor, the functional electrical stimulator based on a digital filter showed an 84% tremor suppression, compared to a 65% when an analog filter was used [77]. The current approaches in utilizing FES to suppress tremors echo these early works, involving primarily two strategies: out-of-phase and co-contraction stimulations [78].
The MOTIMOVE system (3F-Fit Fabricando Faber, Belgrade, Serbia), based on an out-of-phase stimulation, obtained a CE marking for use in the European Union in 2019 but has not been approved by the FDA (Figure 2). The use of the MOTIMOVE system has been studied in patients with ET, PD tremors [79], and hemiplegia [80]. Two prototypes, the TREMOR neurorobot and the Tremor’s glove, have adapted the co-contraction stimulation. Both devices have been assessed in patients with ET or PD but are currently not approved for clinical use.
The MOTIMOVE system consists of a multichannel stimulator that provides support to activate several electrodes, placed on the forearm and upper arm above the flexor and extensor muscle points, that enable the selective muscle activation via distributed, asynchronous electrical stimulation. The inertial sensors within MOTIMOVE deliver real-time estimation of tremulous movements to a host computer, which provides control over the stimulation of muscles. This system delivers out-of-phase stimulation by sending electrical current pulses to the flexor and extensor muscles, triggering the depolarization of motor neurons that counteracts the tremorgenic activity. A pilot study of MOTIMOVE revealed a 67% tremor suppression in six of seven patients with ET or PD [79]. One patient, however, did not respond, suggesting that out-of-phase stimulation may not work for all patients with tremor [79]. Additional clinical studies evaluating the MOTIMOVE system are claimed to be currently in progress in Serbia, France, and Hungary, which will hopefully demonstrate its efficacy in tremor suppression and feasibility.
The TREMOR neurorobot and the Tremor’s glove adopt a similar design as the MOTIMOVE, consisting of electrodes that provide muscle stimulation, inertial sensors that capture biomechanical characterization signals of tremor, and a controller. Both devices adapt the co-contraction stimulation strategy, which applies mechanical loading via continuous transcutaneous stimulation to a pair of antagonistic muscles, increasing the stiffness of the limb. In turn, this filters out the mechanical manifestation of tremorgenic activity, which are oscillations in the muscle tissue. Like MOTIMOVE, the TREMOR neurorobot stimulates the flexor and extensor muscles of the forearm. This device was found to have a 52% tremor suppression in six patients with ET or PD tremors (p < 0.001) [81]. Conversely, the Tremor’s glove stimulates the abductor pollicis brevis and the first and second dorsal interossei muscles of the hand. In a sham-controlled randomized trial of 30 patients with medically refractory tremor in PD, the use of the Tremor’s glove was associated with a significant reduction in the Unified Parkinson’s Disease Rating Scale (UPDRS) score (p = 0.001), suggesting improved experiences of daily living and motor complications [82].
The A-alpha (Aα) sensory fiber, a primary afferent nerve fiber that innervates antagonistic muscle pairs, appears to have a crucial role in the complex neural pathways that are involved in tremor pathophysiology. The reciprocal inhibition of Aα fibers seems to decrease the excitability of antagonist motor neurons and increase the excitability of agonist motor neurons [83]. While it is not entirely clear whether the reciprocal activation of Aα fibers results in tremor, intermittent stimulation of the Aα fibers innervating the flexor and extensor muscles via FES has been studied in patients with ET or PD, showing a 58% tremor suppression [78]. Another study of 14 patients with PD tremor also observed reduction in tremor amplitude and frequency [84]. This suggests that the excitability of antagonist and agonist motor neurons can be modulated, thereby supporting the mechanism by which FES attenuates tremor.
Muscle fatigue is commonly seen as a nonserious adverse event in patients who are treated with FES, owing to the fact that both out-of-phase and co-contraction stimulations lead to the activation of joints and muscle contraction [85]. The use of the Tremor’s glove can also result in numbness of the hand and burning sensation [82]. Contraindications for FES include a prior implanted electrical device, cancer, osteomyelitis, thrombosis/hemorrhage, epilepsy, or pregnancy [86]. In each case, it is incumbent on physicians to evaluate the risk and benefit of a FES treatment based on the patient’s medical history.
Functional electrical stimulators are minimally invasive and demonstrate sufficient efficacy in the suppression of tremor. However, muscle fatigue during repeated FES-induced contraction limits their long-term use. To address this limitation, several emerging technologies have been proposed to reduce or counter muscle fatigue during FES [85]. These functional electrical stimulators have only been studied in small cohorts of patients. Large scale, sham-controlled randomized trials are necessary to validate the efficacy and safety of these devices.

4. Wearable Orthoses

The first reported mechanical solution for the suppression of hand tremors was focused on clasping the patient’s arm to prevent involuntary spasms, patented by Terry and Hoyt in 1980 [87]. However, this approach was not developed further. In 1998, the Viscous Beam orthosis [88] was developed based on previously established principles of energy dissipation [45,46,48] to suppress tremor along the wrist flexion/extension. This device showed success, demonstrating that energy dissipation could be employed in an orthosis. However, it was limited by the fixed damping rate, leading to inconsistent tremor suppression [88].
Tremor suppression orthoses for the upper limbs, wrist, and elbow joints are classified into active, semi-active, or passive. Active orthoses work by generating an active force that counteracts the involuntary motions while supporting the voluntary motions in patients with tremors. In contrast, semi-active and passive orthoses leverage energy dissipation or absorption to suppress involuntary movements. Unlike passive orthoses, the damping magnitude of semi-active orthoses can be adjusted by an active controller. Tremelo (Five Microns, Fresno, CA, USA), Steadi-One (Steadiwear, Toronto, ON, Canada), and Readi-Steadi (Readi-Steadi, Gonzales, LA, USA) are the three passive orthoses currently available for use in patients with tremor. Both Tremelo and Steadi-One are FDA-registered, while Readi-Steadi is FDA-exempted. Active and semi-active orthoses are currently being researched but are not clinically available.

4.1. Active Suppression

In 2005, the Wearable Orthosis for Tremor Assessment and Suppression (WOTAS) exoskeleton was developed as part of the Dynamically Responsive Interventions for Tremor Suppression (DRIFTS) project of the European Commission [89,90,91,92]. WOTAS consists of sensors that measure rotational motions around the joints, electrical direct current (DC) motors that act as actuators to exert force to suppress tremor by converting electrical energy into mechanical energy, and a controller. This device is placed parallel to the upper limb, suppressing tremor in the wrist flexion/extension and pronation/supination and the elbow flexion/extension. In ten patients with tremors, WOTAS demonstrated a 40% tremor suppression [92]. The major drawback of this device is that it is large and bulky, posing social exclusion concerns [91].
Subsequent active orthoses were developed with similar designs and mechatronics to WOTAS but vary in the types of actuators to reduce the weight and improve the tremor suppression efficacy. The pneumatic actuator, which has a large power-to-weight ratio, was implemented in an orthosis, along with an adaptive tremor estimation algorithm, to suppress tremors in the wrist flexion/extension and adduction/abduction [93,94,95]. The results at the testbench using datasets from ten patients with ET or PD tremor showed a 98.1% tremor suppression [95]. The adaptive disturbance rejection controller, utilizing a permanent magnet linear motor (PMLM), demonstrated a 97.6% tremor suppression when examined with five tremulous signals from patients with PD [96]. Compared to the pneumatic actuator, the PMLM is simpler and faster to control and requires only one sensor [96]. The voluntary-driven elbow orthosis, using an electronically communicated (EC) motor, provided a 99.8% tremor suppression in lab simulation using data from a patient with ET [97]. The wearable tremor suppression glove (WTSG), consisting of an actuation box that includes a multi-channel mechatronic splitter (MMS), aims to provide power support from a single input source to multiple output applications [98]. The MMS incorporates a power EC motor and a steering EC motor to suppress tremors in the wrists and hands [98]. The efficacy of this actuation system, however, has not been evaluated.
Other active orthoses have sought to integrate complex sensor systems to characterize voluntary motions and detect tremors. The myoelectric-controlled upper limb orthosis incorporated an algorithm that recognizes and extracts voluntary movements from surface myoelectric signals [99,100,101,102]. The results from six participants who were healthy or with ET showed a recognition rate of 82% [102]. The newer version of this orthosis adopted a different design that improved flexibility and ease of wear [103,104]. In a healthy participant with FES-induced muscle contraction, the myoelectric-controlled orthosis reduced oscillations in the elbow flexion/extension by about 50–80% [104]. Huen and colleagues implemented context aware body sensor network (BSN) sensors into an upper limb orthosis to enable the detection of six ADLs [105]. In six healthy participants with simulated tremor movements, the BSN-integrated exoskeleton exhibited a 70% accuracy rate in identifying ADLs and a 77% tremor suppression [105].

4.2. Semi-Active Suppression

Several semi-active orthoses utilize magnetorheological (MR) fluids as a strategy to provide tremor suppression. MR fluids consist of magnetizable, microscopic particles dispersed in oil or water. Upon encountering a magnetic field, these particles experience attractive force, and the viscosity of the MR fluids increases, opposing the existing flow. This rheological property has been exploited in tremor suppression orthoses by varying magnetic field intensities to tune the resistance force for tremor suppression [106].
The Double Viscous Beam (DVB) orthosis, positioned on the dorsal surface of the forearm, consists of a chamber of MR fluids and two shear plates to make up a passive actuator, applying mechanical loads for tremor suppression [107]. Compared to the previous approach [88], the DVB orthosis has an improved responsiveness to the viscous resistance as a result of the increased shear strength. This orthosis is coupled to a sensor and a controller to optimize the actuation performance.
Case and colleagues developed a wearable orthosis that incorporated four MR dampers for the wrist flexion/extension and abduction/adduction, the elbow flexion/extension, and the forearm pronation/supination [108,109,110,111]. An estimation algorithm for tremor frequency and a controller were used to measure the amount of resistance force needed to counteract tremulous movements. The resistance force generated by the MR dampers depends on a piston-coil design.
More recently, the MR damper-based soft exoskeleton for the tremor suppression (SETS) system was proposed to suppress tremor in the wrist [112]. Unlike previously designed semi-active orthoses, the SETS system equips a controllable flexible semi-active actuator that dynamically adapts to the motions of the wrist joint, providing tremor suppression in the wrist flexion/extension, abduction/adduction, rotation. This device also integrates passive suppression with two hyper-elastic blades, which suppress tremor in the wrist supination/pronation. The SETS system demonstrates potential clinical utility with its compatibility with the human wrist, real-time tunability based on tremor frequency, and lightweight design.
The carbon fiber-based, lightweight orthosis developed by Herrnstadt and Menon is magnetically activated by an electromagnetic brake (EB) [113]. When the tremor frequency and joint angular displacement are detected by a sensor and potentiometer, respectively, a pulse width modulation signal is sent from the controller to produce a magnetic field, exciting the EB. In turn, the EB actuates the orthosis to generate a resistive force for tremor suppression. In comparison to electric motors such as the DC motor, EBs are capable of producing a higher force while consuming less power. In three healthy participants with simulated tremor motions, the use of the EB-based orthosis demonstrated an 88% tremor suppression [113].
Apart from magnetically driven semi-active orthoses, Kalaiarasi and Kumar designed a pneumatically controlled hand cuff [114]. Similarly, this device is built along with an accelerometer that sends tremor frequency data to a controller. When the threshold is met, an air pump inflates the hand cuff, yielding a resistance force in a reciprocating, linear motion to suppress the tremor. Inflation and deflation of the hand cuff are enabled by two separate valves. The limited efficacy of this approach was observed in one patient with ET who experienced a 30% tremor suppression [114].

4.3. Passive Suppression

Tremelo utilized two tuned vibration absorbers (TVAs) that are positioned over the dorsal and ventral surfaces of the arm (Figure 3). Each TVA contains a mass-spring-damper system in which the vibration energy of involuntary motions of the shaking arm during tremor are transferred from the spring to the added mass. This results in reduced tremulous movements and substantial motions of the added mass within the TVA. This device is purely mechanical, eliminating the need of a power source. Preliminary results showed an 85% tremor suppression in a patient with PD tremors [115]. Recruitment for a pilot clinical study is ongoing, which should provide further data.
Steadi-One is mechanical device that integrates a tuned mass damper (TMD), which obviates the need for a power source [116] (Figure 4). Like TVAs, the TMD embodies a mass-spring-damper system. The difference between TMDs and TVAs is the presence of a dissipating element, which, in Steadi-One, is a non-Newtonian fluid in the interior space of the TMD. When the vibration energy is transferred to the added mass, this non-Newtonian fluid becomes viscous, reducing its amplitude of motions. There are no publicly available data to support its efficacy. However, it is claimed (https://www.steadiwear.com, accessed on 26 March 2021) to have an 85–90% tremor suppression during the lab simulation.
The Readi-Steadi glove embeds a multitude of metal disks that aims to add inertia to the tremulous hand. A preliminary study involving 40 participants who were healthy or with ET observed a 50% tremor suppression [117]. The metal disks function as sensory tricks that influence the aberrant sensorimotor integration to suppress tremor. While there is no study that examines the effectiveness of the sensory trick phenomenon in patients with ET or PD tremor, it has been studied in 30 patients with musician’s dystonia [118]. By wearing a glove, patients with more severe symptoms of dystonia showed better improvements in fine motor control (Pearson’s r = −0.45; p = 0.01) [118].
The Task-Adjustable Passive Orthosis (TAPO) has a textile glove design to enhance wearability and comfort for daily activities. An air-filled structure, inflated on-demand by hand or electrical pump, is fitted within the glove on the dorsal surface of the hand. The inflated TAPO applies pressure to the back of the hand and the forearm, suppressing the involuntary motions in the wrist flexion/extension, ulnar/radial deviation, and pronation/supination. The proof of concept of TAPO has been examined in a patient with PD tremors performing six ADLs. The use of TAPO was associated with a significant tremor suppression in three specific tasks, including 82% while drinking (p = 0.03), 79% while pouring (p = 0.03), and 74% while drawing a spiral (p = 0.03) [119].
More recently, Lu and Huang examined and established a mechanical model for particle damping for passive vibration suppression in tremors [120]. Particle dampers involve the potential of energy absorption and dissipation through momentum exchange between moving particles and vibrating walls. There are several advantages of using particle dampers, including simple construction, low cost, robustness and reliability, wide damping frequency band, and insensitivity to extreme temperature [120]. At a high tremor frequency, the provided damping of the particle damper became nearly independent to the frequency and amplitude of tremor, indicating that it is suitable for tremor suppression [120].
Two other passive orthoses have also been previously investigated for their use in tremor suppression. The Vib-Bracelet, also designed with an incorporated TMD, suppresses tremors in the wrist pronation/supination [121,122]. The result at the testbench using tremor data from one patient with PD showed an 85% tremor suppression [122]. Another approach, proposed by Takanokura and colleagues [123], involved implementing air dashpots into an orthosis to suppress tremors in the wrist flexion/extension and ulnar/radial deviation, as well as the elbow flexion/extension. In a healthy participant with electrical stimulation-induced muscle contraction, this orthosis demonstrated an involuntary movement suppression of 62% in the wrist when two air dashpots were used and 82% in the elbow [123].

4.4. Mechanism Underpinning the Efficacy of Wearable Orthoses in Tremor Suppression

Although some of the underlying causes of tremors remain unknown, several putative interactive factors contributing to the motor expression of tremor have been hypothesized. These include the oscillating tendencies of the joint and muscular mechanical systems, short- and long-loop reflexes of the spinal cord and the brainstem, and the closed-loop feedback systems of higher motor centers such as the cerebellum [124]. Unlike electrical stimulation systems, wearable orthoses target the clinical manifestations of tremors. By generating an opposite force of equal magnitude, these devices attempt to mechanically counteract the involuntary movements.

4.5. Comparing Active, Semi-Active, and Passive Orthoses for Tremor Suppression

Unlike semi-active and passive orthoses, active orthoses often rely on actuators coupled to a signal transmission system, resulting in their heavy and unwieldy nature. A reduction in the overall weights to improve their wearability is an important research priority. For example, Kelley and Kauffman recently proposed substituting the traditional actuators with the soft and compliant dielectric elastomer stack actuators to enable an orthosis conforming to the human joints [125,126]. In lieu of the metallic structure of previous orthoses, the BSN-integrated exoskeleton leveraged plastic materials to reduce the weight [105]. By using an MMS to support multiple output applications with one drive motor, the WTSG has a reduced size and weight [98].
Wearable orthoses are primarily noninvasive in suppressing tremors. However, their safety profile has not been established, because most wearable orthoses were only assessed in small cohorts of patients or at the testbench with data simulation. Furthermore, the small sample size of these studies may undermine the reliability of the data. It is likely that wearable orthoses will become the most widely used medical devices for tremor suppression, given their promising efficacy. Major challenges include developing orthoses that are lightweight and soft in texture, studying the orthotic placement that will result in maximized tremor suppression, and improving the ergonomic design based on the anatomy of the upper limbs. Addressing these questions with further studies should enhance our understanding of the feasibility and practicality of clinically implementing wearable orthoses to suppress tremor.

5. Assistive Feeding Devices

The Neater Eater (Neater Solutions, Derbyshire, UK) was introduced by Michaelis in 1988 [127]. However, it was not available for use in the US until its registration with the FDA in 1993. This is a table-mounted device that involves internal spring-assisted lifting to support a clip-on utensil to enable eating. It relies on viscous damping to absorb tremors and fast movements via the flowing of a viscous fluid that dissipates the kinetic energy. A brief report interviewing 39 participants with various neuromuscular conditions found that the Neater Eater is associated with positive impacts in independence, self-confidence, and quality of life [128]. Follow-up and monitoring may be necessary to prevent fatigue and muscle build-up from using this device on a regular basis.
Liftware Steady (Verily Life Sciences, South San Francisco, CA, USA), registered with the FDA in 2013, is a handheld device designed to help patients with ET or PD tremors eat. It consists of a motion-generating platform, capable of directing two DC motors to move the utensil opposite to the direction of the tremor. The patient’s involuntary movements are detected by an accelerometer, which are then transmitted to a controller, providing control over the motion-generating platform. A pilot study involving 15 patients with ET demonstrated an improvement in tremor with the device, as measured by Fahn-Tolosa-Marin Tremor Rating Scale (TRS), while holding (p = 0.016), eating (p = 0.001), and transferring objects (p = 0.001) [129]. When using the device, patients expressed improved symptoms of ET while eating (p < 0.001) and transferring objects (p = 0.013) but not holding them (p = 0.14), as measured by the subject-rated Clinical Global Impression Scale (CGI-S) [129]. Data demonstrated a 73% tremor suppression across the three tasks [129]. Compared to other adaptive utensils in 22 patients with ET or PD tremor, Liftware Steady was preferred [130].
The Gyenno Spoon (GYENNO Technologies, Shenzhen, China), another handheld assistive feeding device, consists of a longitudinal motor and a transverse motor that are able to generate movements in two different directions opposite to the tremor direction [131]. Both motors are linked to a control module, which receives vibration data from multiple movement sensors within the device. Its ergonomic design and proposed efficacy led to its registration with the FDA in 2016 for use for patients with ET or PD tremors. While it claims (https://www.gyenno.com/spoon-en, accessed on 26 March 2021) to have an 85% tremor suppression, no clinical data has been published.
Contrary to Liftware Steady and the Gyenno Spoon, the Neater Eater requires no power supply and has the ability to automatically bring the spoon forward to the patient’s mouth with an internal spring. However, both Liftware Steady and the Gyenno Spoon enable data collection of the patient’s tremor. This can be beneficial in allowing physicians to monitor their patients’ tremor improvement or progression when Liftware Steady or the Gyenno Spoon is supplemented with pharmacotherapy. Nonetheless, it is important to note that the functionality of these devices is limited to provide support in feeding only.

6. Other Devices

6.1. Gyroscopic Stabilization

The GyroGlove (GyroGear, London, UK) is developed with a plurality of gyroscopes, mounted to a fabric glove on the dorsal surface of the hand, that function to counteract tremors [132]. Each gyroscope includes a rotatable disc that is capable of rotating about an axis to resist involuntary motions. This allows its angular momentum to be conserved when a rotational displacement is encountered, with opposing force inputs from any direction. Unlike previous gyroscopic devices using a single gyroscope [133,134], the use of multiple gyroscopes in the GyroGlove allows it to suppress involuntary movements in multiple planar directions. This device is currently in the advanced stage of development, and data is needed to inform its efficacy.

6.2. Haptic Stimulation Systems

The Emma Watch (Microsoft, Redmond, WA, USA) is a wrist-worn device consisting of several vibration-generating actuators aimed at providing haptic stimulation to the wrist. Mechanosensitive receptors, such as the Pacinian and Meissner corpuscles in the upper limbs, deliver afferent signals to the cuneate nucleus in response to vibratory stimuli [135,136]. Proprioceptive inputs from the cuneate nucleus are projected to the thalamus [136,137,138], implicating its possible role in the neural pathway associated with tremors [139]. However, a study involving 18 patients with ET reported that a mechanical vibration to the hand and forearm via piezoelectric actuators resulted in no homogenous effect in the tremor amplitude [139]. Across different frequencies of vibratory stimuli, 50–72% of patients experienced an increase in tremor amplitude, while 5–22% of patients showed a decrease [139]. The Emma Watch will require clinical validations, since the use of haptic stimulation for tremor suppression has, to date, only led to questionable efficacy.

7. Tremor Suppression Devices: Place in Therapy

The onset of ET can occur early in childhood due to familial factors, but the majority of cases of ET appeared after the age of 40 [140]. One study investigated the correlation between the age of onset and the progression of ET in 115 patients [141]. Patients with an age of onset later than 60 years experienced a more rapid progression when compared to patients with a younger age of onset (p < 0.001) [141]. Since the onset of ET and PD tremors typically occurs in middle to late adulthood, aging-associated diseases such as dementia [142,143] and mild cognitive impairment [144,145,146] intersect with both of these conditions. These neurological disorders may further preclude patients from adhering to pharmacotherapies.
The medical devices described above offer alternative options for the suppression of tremors (Table 1), especially in patients who are not eligible for surgical interventions (i.e., DBS, SRS, and MRgFUS). However, the use of these devices is patient specific. For example, although Cala Trio has an aesthetic design that will likely not pose any social concerns, wearable orthoses may be a better option if the patient has any contraindication to the use of electrical stimulation systems. Depending on the patient’s needs, assistive feeding devices may be a useful addition to the patient’s daily living. Most of the devices that are available for use are subjected to the FDA’s Class I general control for safety and efficacy assurance. In addition to the general control, Cala ONE requires Class II special control for its performance standards and special prescriber labeling.
ET is associated with a staggering cost of direct medical expenses, indirect productivity and income losses, nonmedical expenses, and disability benefits. The unemployment rate increases to about 88% in patients whose ET progresses from mild to severe [147], leading to forced early retirements. Collectively, patients with mild ET have a 1.83-year average loss of employment, corresponding to a $280 billion in income loss [147]. In patients with moderate to severe tremors, the average loss of employment is 6.5 years [147]. ET and PD tremors likely increase the economic burden more than currently estimated due to their progressive natures and the underreported cases. The development of a medical device for tremor suppression is an under-researched area. Most of the investigational devices discussed were abandoned before entering the market. However, it is imperative that the search for safe and effective tremor suppression devices continues, given the overall economic burden of tremors. Given that most of the currently available devices are based on preliminary data, more investigation is needed to understand the safety and efficacy of these devices before their use in clinical practice can be supported. Cost-effectiveness data are necessary and important to convince insurance programs to provide coverage, alleviating the financial constraints on patients and caregivers.

8. Future Perspectives

The devices currently studied have employed distinctive mechanistic approaches. The weight of evidence supporting their efficacy challenges the notion that tremors originate from a single, dominant pathway. Additional pathological insights, such as the loss of Purkinje cells in ET [148,149] and increased central oscillator synchronization in the basal ganglia in PD tremors [150], along with several mechanistic targets of tremor suppression devices, highlight the advances in our understanding of how tremors may be generated. Perhaps the most pertinent pathway implicated in tremors is the cortico-ponto-cerebello-thalamo-cortical loop, which serves as the basis for successful surgical interventions [21]. These findings suggest an integrative multi-pathway model for tremor pathogenesis. The relevance of these pathways necessitates a further clarification of the complexities and inter-related causes of tremors, which is central to spur the future development of safer and more effective devices for tremor suppression.
The lack of consensus on the characterization and electrophysiology of tremor previously represented two major diagnostic pitfalls [151]. However, in 2018, the IPMDS task force reviewed the vast uncertainties to update its consensus classification criteria for tremor disorders [1]. Besides ET and PD tremors, it is important to recognize that a wide range of other tremor conditions also affect the upper limbs with varying clinical features and etiologies [1]. Future studies could investigate whether the efficacy of these devices is generalizable to other tremor conditions. As seen in the pivotal Cala ONE trial [70], tremor suppression can, in part, be attributed to the surgical placebo effect. Since the studies of most of these devices were descriptive in design, sham-controlled randomized trials are warranted to confirm their efficacy. Lastly, evaluating the concurrent use of one or more devices, along with pharmacotherapy/lifestyle interventions, may derive insightful data to explain the benefits and overall impact of a multimodal strategy in the management of tremors.

9. Conclusions

Although tremors are not a life-threatening movement disorder, they can be disabling and negatively impact the patient’s quality of life. Our limited knowledge of the pathophysiology of tremors has given rise to the challenge of developing effective or curative pharmacotherapies. In the past several decades, many medical devices, with a broad range of mechanisms, have been developed to suppress tremors in different aspects of daily living. Based on the current evidence, some of these devices appear to have promise as potentially safe and effective options in the medical armamentarium for tremor suppression. Most of these devices are noninvasive and placed externally around the wrists or the upper limbs. It is important to note that externally wearing a device could pose a cosmetic and social concern, so understanding the acceptability of tremor medical devices among patients with tremors is warranted. Nonetheless, it is likely that the future of tremor management will benefit from the addition of medical devices into the patient’s existing pharmacotherapy and/or lifestyle intervention. However, given the high variability in the quality of the current studies, future research is needed to better understand the long-term efficacy, safety, and cost-effectiveness of the tremor suppression devices to fulfill this promise.

Author Contributions

Conceptualization: R.P.; Writing—original draft: J.M.; Writing—reviewing and editing—All authors. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data sharing not applicable.

Acknowledgments

The authors wish to thank the School of Pharmacy at the Massachusetts College of Pharmacy and Health Sciences University for their financial support of this project.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Bhatia, K.P.; Bain, P.; Bajaj, N.; Elble, R.J.; Hallett, M.; Louis, E.D.; Raethjen, J.; Stamelou, M.; Testa, C.M.; Deuschl, G.; et al. Consensus Statement on the classification of tremors. From the task force on tremor of the International Parkinson and Movement Disorder Society. Mov. Disord. 2018, 33, 75–87. [Google Scholar] [CrossRef]
  2. Louis, E.D.; Ferreira, J.J. How common is the most common adult movement disorder? Update on the worldwide prevalence of essential tremor. Mov. Disord. 2010, 25, 534–541. [Google Scholar] [CrossRef]
  3. Louis, E.D.; Ottman, R.; Hauser, W.A. How common is the most common adult movement disorder? Estimates of the prevalence of essential tremor throughout the world. Mov. Disord. 1998, 13, 5–10. [Google Scholar] [CrossRef] [PubMed]
  4. Dorsey, E.R.; Elbaz, A.; Nichols, E.; Abd-Allah, F.; Abdelalim, A.; Adsuar, J.C.; Ansha, M.G.; Brayne, C.; Choi, J.-Y.J.; Collado-Mateo, D.; et al. Global, regional, and national burden of Parkinson’s disease, 1990–2016: A systematic analysis for the Global Burden of Disease Study 2016. Lancet Neurol. 2018, 17, 939–953. [Google Scholar] [CrossRef] [Green Version]
  5. National Institute for Health and Care Excellence (UK). Parkinson’s Disease in Adults: Diagnosis and Management; National Institute for Health and Care Excellence (UK): London, UK, 2017. [Google Scholar]
  6. Louis, E.D.; Barnes, L.; Albert, S.M.; Cote, L.; Schneier, F.R.; Pullman, S.L.; Yu, Q. Correlates of functional disability in essential tremor. Mov. Disord. 2001, 16, 914–920. [Google Scholar] [CrossRef]
  7. Elble, R.J.; Brilliant, M.; Leffler, K.; Higgins, C. Quantification of essential tremor in writing and drawing. Mov. Disord. 1996, 11, 70–78. [Google Scholar] [CrossRef]
  8. Héroux, M.E.; Parisi, S.L.; Larocerie-Salgado, J.; Norman, K.E. Upper-Extremity Disability in Essential Tremor. Arch. Phys. Med. Rehabil. 2006, 87, 661–670. [Google Scholar] [CrossRef] [PubMed]
  9. Norman, K.E.; D’Amboise, S.N.; Pari, G.; Héroux, M.E. Tremor during movement correlates well with disability in people with essential tremor. Mov. Disord. 2011, 26, 2088–2094. [Google Scholar] [CrossRef] [PubMed]
  10. Rajput, A.H.; Robinson, C.A. Essential tremor course and disability: A clinicopathologic study of 20 cases. Neurology 2004, 62, 932–936. [Google Scholar] [CrossRef]
  11. Louis, E.D.; Machado, D.G. Tremor-related quality of life: A comparison of essential tremor vs. Parkinson’s disease patients. Park. Relat. Disord. 2015, 21, 729–735. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  12. Lorenz, D.; Schwieger, D.; Moises, H.; Deuschl, G. Quality of life and personality in essential tremor patients. Mov. Disord. 2006, 21, 1114–1118. [Google Scholar] [CrossRef]
  13. Monin, J.K.; Gutierrez, J.; Kellner, S.; Morgan, S.; Collins, K.; Rohl, B.; Migliore, F.; Cosentino, S.; Huey, E.; Louis, E.D. Psychological Suffering in Essential Tremor: A Study of Patients and Those Who Are Close to Them. Tremor Other Hyperkinetic Mov. 2017, 7, 526. [Google Scholar] [CrossRef]
  14. Schneier, F.R.; Barnes, L.F.; Albert, S.M.; Louis, E.D. Characteristics of Social Phobia among Persons with Essential Tremor. J. Clin. Psychiatry 2001, 62, 367–372. [Google Scholar] [CrossRef]
  15. Damen, J.A.A.G. Author’s reply to Woodward. BMJ 2016, 354, i4485. [Google Scholar] [CrossRef]
  16. Elias, W.J.; Shah, B.B. Tremor. JAMA 2014, 311, 948–954. [Google Scholar] [CrossRef] [PubMed]
  17. Zesiewicz, T.A.; Elble, R.; Louis, E.D.; Hauser, R.A.; Sullivan, K.L.; Dewey, R.B.; Ondo, W.G.; Gronseth, G.S.; Weiner, W.J. Practice Parameter: Therapies for essential tremor: Report of the Quality Standards Subcommittee of the American Academy of Neurology. Neurology 2005, 64, 2008–2020. [Google Scholar] [CrossRef] [Green Version]
  18. Zesiewicz, T.A.; Elble, R.J.; Louis, E.D.; Gronseth, G.S.; Ondo, W.G.; Dewey, R.B.; Okun, M.S.; Sullivan, K.L.; Weiner, W.J. Evidence-based guideline update: Treatment of essential tremor: Report of the Quality Standards Subcommittee of the American Academy of Neurology. Neurology 2011, 77, 1752–1755. [Google Scholar] [CrossRef] [PubMed]
  19. Ferreira, J.J.; Mestre, T.A.; Lyons, K.E.; Benito-León, J.; Tan, E.; Abbruzzese, G.; Hallett, M.; Haubenberger, D.; Elble, R.; Deuschl, G.; et al. MDS evidence-based review of treatments for essential tremor. Mov. Disord. 2019, 34, 950–958. [Google Scholar] [CrossRef]
  20. O’Connor, R.J.; Kini, M.U. Non-pharmacological and non-surgical interventions for tremor: A systematic review. Park. Relat. Disord. 2011, 17, 509–515. [Google Scholar] [CrossRef]
  21. Haubenberger, D.; Hallett, M. Essential Tremor. N. Engl. J. Med. 2018, 378, 1802–1810. [Google Scholar] [CrossRef]
  22. Deuschl, G.; Raethjen, J.; Hellriegel, H.; Elble, R. Treatment of patients with essential tremor. Lancet Neurol. 2011, 10, 148–161. [Google Scholar] [CrossRef]
  23. Diaz, N.L.; Louis, E.D. Survey of medication usage patterns among essential tremor patients: Movement disorder specialists vs. general neurologists. Park. Relat. Disord. 2010, 16, 604–607. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  24. Zappia, M.; Italian Movement Disorders Association (DISMOV-SIN) Essential Tremor Committee; Albanese, A.; Bruno, E.; Colosimo, C.; Filippini, G.; Martinelli, P.; Nicoletti, A.; Quattrocchi, G. Treatment of essential tremor: A systematic review of evidence and recommendations from the Italian Movement Disorders Association. J. Neurol. 2012, 260, 714–740. [Google Scholar] [CrossRef] [PubMed]
  25. Velisar, A.; Syrkin-Nikolau, J.; Blumenfeld, Z.; Trager, M.; Afzal, M.; Prabhakar, V.; Bronte-Stewart, H. Dual threshold neural closed loop deep brain stimulation in Parkinson disease patients. Brain Stimul. 2019, 12, 868–876. [Google Scholar] [CrossRef]
  26. Weerasinghe, G.; Duchet, B.; Cagnan, H.; Brown, P.; Bick, C.; Bogacz, R. Predicting the effects of deep brain stimulation using a reduced coupled oscillator model. PLoS Comput. Biol. 2019, 15, e1006575. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  27. Kern, D.S.; Picillo, M.; Thompson, J.A.; Sammartino, F.; Di Biase, L.; Munhoz, R.P.; Fasano, A. Interleaving Stimulation in Parkinson’s Disease, Tremor, and Dystonia. Ster. Funct. Neurosurg. 2018, 96, 379–391. [Google Scholar] [CrossRef] [PubMed]
  28. Chen, K.S.; Chen, R. Invasive and Noninvasive Brain Stimulation in Parkinson’s Disease: Clinical Effects and Future Perspectives. Clin. Pharmacol. Ther. 2019, 106, 763–775. [Google Scholar] [CrossRef]
  29. Farokhniaee, A.; McIntyre, C.C. Theoretical principles of deep brain stimulation induced synaptic suppression. Brain Stimul. 2019, 12, 1402–1409. [Google Scholar] [CrossRef]
  30. Mao, Z.; Ling, Z.; Pan, L.; Xu, X.; Cui, Z.; Liang, S.; Yu, X. Comparison of Efficacy of Deep Brain Stimulation of Different Targets in Parkinson’s Disease: A Network Meta-Analysis. Front. Aging Neurosci. 2019, 11, 23. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  31. Schuurman, P.R.; Bosch, D.A.; Bossuyt, P.M.; Bonsel, G.J.; Van Someren, E.J.; De Bie, R.M.; Merkus, M.P.; Speelman, J.D. A Comparison of Continuous Thalamic Stimulation and Thalamotomy for Suppression of Severe Tremor. N. Engl. J. Med. 2000, 342, 461–468. [Google Scholar] [CrossRef] [Green Version]
  32. Elble, R.J.; Shih, L.; Cozzens, J.W. Surgical treatments for essential tremor. Expert Rev. Neurother. 2018, 18, 303–321. [Google Scholar] [CrossRef]
  33. Ravikumar, V.K.; Parker, J.J.; Hornbeck, T.S.; Santini, V.E.; Pauly, K.B.; Wintermark, M.; Ghanouni, P.; Stein, S.C.; Halpern, C.H. Cost-effectiveness of focused ultrasound, radiosurgery, and DBS for essential tremor. Mov. Disord. 2017, 32, 1165–1173. [Google Scholar] [CrossRef]
  34. Lozano, A.M.; Lipsman, N.; Bergman, H.; Brown, P.; Chabardes, S.; Chang, J.W.; Matthews, K.; McIntyre, C.C.; Schlaepfer, T.E.; Schulder, M.; et al. Deep brain stimulation: Current challenges and future directions. Nat. Rev. Neurol. 2019, 15, 148–160. [Google Scholar] [CrossRef] [PubMed]
  35. Walters, H.; Shah, B.B. Focused Ultrasound and Other Lesioning Therapies in Movement Disorders. Curr. Neurol. Neurosci. Rep. 2019, 19, 66. [Google Scholar] [CrossRef] [PubMed]
  36. Chan, A.K.; McGovern, R.A.; Brown, L.T.; Sheehy, J.P.; Zacharia, B.E.; Mikell, C.B.; Bruce, S.S.; Ford, B.; McKhann, G.M. Disparities in Access to Deep Brain Stimulation Surgery for Parkinson Disease. JAMA Neurol. 2014, 71, 291–299. [Google Scholar] [CrossRef] [Green Version]
  37. Lange, M.; Mauerer, J.; Schlaier, J.; Janzen, A.; Zeman, F.; Bogdahn, U.; Brawanski, A.; Hochreiter, A. Underutilization of deep brain stimulation for Parkinson’s disease? A survey on possible clinical reasons. Acta Neurochir. 2017, 159, 771–778. [Google Scholar] [CrossRef] [PubMed]
  38. Shukla, A.W.; Deeb, W.; Patel, B.; Ramirez-Zamora, A. Is deep brain stimulation therapy underutilized for movement disorders? Expert Rev. Neurother. 2018, 18, 899–901. [Google Scholar] [CrossRef]
  39. Kim, M.-R.; Yun, J.Y.; Jeon, B.; Lim, Y.H.; Kim, K.R.; Yang, H.-J.; Paek, S.H. Patients’ reluctance to undergo deep brain stimulation for Parkinson’s disease. Park. Relat. Disord. 2016, 23, 91–94. [Google Scholar] [CrossRef]
  40. Food and Drug Administration. Medical Devices; Neurological Devices; Classification of the External Upper Limb Tremor Stimulator. Final order. Fed. Regist. 2018, 83, 52315–52316. [Google Scholar]
  41. Beringhause, S.; Rosen, M.; Huang, S. Evaluation of a damped joystick for people disabled by intention tremor. In Proceedings of the 12th Annual Conference on Rehabilitation Technology, New Orleans, LA, USA, 25–30 June 1989; pp. 41–42. [Google Scholar]
  42. Hendriks, J.; Rosen, M.; Berube, N.; Aisen, M. A second-generation joystick for people disabled by tremor. In Proceedings of the 14th Annual RESNA Conference, Kansas City, MO, USA, 21–26 June 1991; pp. 248–251. [Google Scholar]
  43. Rosen, M.J. Tremor Suppressing Hand Controls. U.S. Patent 4,689,449, 25 August 1987. [Google Scholar]
  44. Rosen, M.J. Multiple Degree of Freedom Damped Hand Controls. U.S. Patent 5,107,080, 21 April 1992. [Google Scholar]
  45. Arnold, A.S.; Rosen, M.J.; Aisen, M.L. Evaluation of a controlled-energy-dissipation orthosis for tremor suppression. J. Electromyogr. Kinesiol. 1993, 3, 131–148. [Google Scholar] [CrossRef]
  46. Rosen, M.J.; Arnold, A.S.; Baiges, I.J.; Aisen, M.L.; Eglowstein, S.R. Design of a controlled-energy-dissipation orthosis (CEDO) for functional suppression of intention tremors. J. Rehabil. Res. Dev. 1995, 32, 1–16. [Google Scholar] [PubMed]
  47. Rosen, M.J.; Baiges, I.J. Whole-Arm Orthosis for Steadying Limb Motion. U.S. Patent 5,231,998, 3 August 1993. [Google Scholar]
  48. Maxwell, S.M. A Modulated-Energy-Dissipation Manipulator and Application to Suppressing Human Arm Tremor. Ph.D. Thesis, Massachusetts Institute of Technology, Cambridge, MA, USA, 1990. [Google Scholar]
  49. Maxwell, S.M. A System for Resisting Limb Movement. EP Patent 0,569,489 B1, 10 May 1995. [Google Scholar]
  50. Johnson, M. Transcutaneous Electrical Nerve Stimulation: Mechanisms, Clinical Application and Evidence. Rev. Pain 2007, 1, 7–11. [Google Scholar] [CrossRef] [PubMed]
  51. Gibson, W.; Wand, B.M.; Meads, C.; Catley, M.J.; E O’Connell, N. Transcutaneous electrical nerve stimulation (TENS) for chronic pain—An overview of Cochrane Reviews. Cochrane Database Syst. Rev. 2019, 4, CD011890. [Google Scholar] [CrossRef] [PubMed]
  52. Johnson, M.I.; Paley, C.A.; Howe, T.E.; Sluka, K.A. Transcutaneous electrical nerve stimulation for acute pain. Cochrane Database Syst. Rev. 2015, 6, CD006142. [Google Scholar] [CrossRef] [Green Version]
  53. Gibson, W.; Wand, B.M.; E O’Connell, N. Transcutaneous electrical nerve stimulation (TENS) for neuropathic pain in adults. Cochrane Database Syst. Rev. 2017, 2017, 011976. [Google Scholar] [CrossRef] [Green Version]
  54. Toglia, J.U.; Izzo, K. Treatment of myoclonic dystonia with transcutaneous electrical nerve stimulation. Neurol. Sci. 1985, 6, 75–78. [Google Scholar] [CrossRef]
  55. Campbell, J.N.; Raja, S.N.; Meyer, R.A.; MacKinnon, S.E. Myelinated afferents signal the hyperalgesia associated with nerve injury. Pain 1988, 32, 89–94. [Google Scholar] [CrossRef]
  56. Truini, A.; Padua, L.; Biasiotta, A.; Caliandro, P.; Pazzaglia, C.; Galeotti, F.; Inghilleri, M.; Cruccu, G. Differential involvement of A-delta and A-beta fibres in neuropathic pain related to carpal tunnel syndrome. Pain 2009, 145, 105–109. [Google Scholar] [CrossRef]
  57. Xu, Z.-Z.; Kim, Y.H.; Bang, S.; Zhang, Y.; Berta, T.; Wang, F.; Oh, S.B.; Ji, R.-R. Inhibition of mechanical allodynia in neuropathic pain by TLR5-mediated A-fiber blockade. Nat. Med. 2015, 21, 1326–1331. [Google Scholar] [CrossRef]
  58. Nagi, S.S.; Marshall, A.G.; Makdani, A.; Jarocka, E.; Liljencrantz, J.; Ridderström, M.; Shaikh, S.; O’Neill, F.; Saade, D.; Donkervoort, S.; et al. An ultrafast system for signaling mechanical pain in human skin. Sci. Adv. 2019, 5, eaaw1297. [Google Scholar] [CrossRef] [Green Version]
  59. Garcia, K.; Wray, J.K.; Kumar, S. Spinal Cord Stimulation, StatPearls; StatPearls Publishing: Treasure Island, FL, USA, 2020. [Google Scholar]
  60. Ferrara, J.; Stamey, W.; Strutt, A.M.; Adam, O.R.; Jankovic, J. Transcutaneous Electrical Stimulation (TENS) for Psychogenic Movement Disorders. J. Neuropsychiatry Clin. Neurosci. 2011, 23, 141–148. [Google Scholar] [CrossRef]
  61. Serrano-Muñoz, D.; Avendaño-Coy, J.; Simón-Martínez, C.; Taylor, J.; Gómez-Soriano, J. Effect of high-frequency alternating current transcutaneous stimulation over muscle strength: A controlled pilot study. J. Neuroeng. Rehabil. 2018, 15, 1–4. [Google Scholar] [CrossRef] [Green Version]
  62. Hao, M.-Z.; Xu, S.-Q.; Hu, Z.-X.; Xu, F.-L.; Niu, C.-X.M.; Xiao, Q.; Lan, N. Inhibition of Parkinsonian tremor with cutaneous afferent evoked by transcutaneous electrical nerve stimulation. J. Neuroeng. Rehabil. 2017, 14, 75. [Google Scholar] [CrossRef] [PubMed]
  63. Hao, M.-Z.; He, X.; Kipke, D.R.; Lan, N. Effects of electrical stimulation of cutaneous afferents on corticospinal transmission of tremor signals in patients with Parkinson’s disease. In Proceedings of the 2013 6th International IEEE/EMBS Conference on Neural Engineering (NER), San Diego, CA, USA, 6–8 November 2013. [Google Scholar]
  64. Munhoz, R.P.; Hanajima, R.; Ashby, P.; Lang, A.E. Acute effect of transcutaneous electrical nerve stimulation on tremor. Mov. Disord. 2002, 18, 191–194. [Google Scholar] [CrossRef] [PubMed]
  65. U.S. National Library of Medicine. Prospective Study for Symptomatic Relief of ET with Cala Therapy (PROSPECT). 2018. Available online: https://clinicaltrials.gov/ct2/show/NCT03597100 (accessed on 20 July 2020).
  66. Hanajima, R. Somatosensory evoked potentials (SEPs) recorded from deep brain stimulation (DBS) electrodes in the thalamus and subthalamic nucleus (STN). Clin. Neurophysiol. 2004, 115, 424–434. [Google Scholar] [CrossRef] [PubMed]
  67. Hanajima, R.; Chen, R.; Ashby, P.; Lozano, A.M.; Hutchison, W.D.; Davis, K.D.; Dostrovsky, J.O. Very Fast Oscillations Evoked by Median Nerve Stimulation in the Human Thalamus and Subthalamic Nucleus. J. Neurophysiol. 2004, 92, 3171–3182. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  68. Bathien, N.; Rondot, P.; Toma, S. Inhibition and synchronisation of tremor induced by a muscle twitch. J. Neurol. Neurosurg. Psychiatry 1980, 43, 713–718. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  69. Dosen, S.; Muceli, S.; Dideriksen, J.L.; Romero, J.P.; Rocon, E.; Pons, J.; Farina, D. Online Tremor Suppression Using Electromyography and Low-Level Electrical Stimulation. IEEE Trans. Neural Syst. Rehabil. Eng. 2014, 23, 385–395. [Google Scholar] [CrossRef] [PubMed]
  70. Pahwa, R.; Dhall, R.; Ostrem, J.; Gwinn, R.; Lyons, K.; Ro, S.; Dietiker, C.; Luthra, N.; Ms, P.C.; Hamner, S.; et al. An Acute Randomized Controlled Trial of Noninvasive Peripheral Nerve Stimulation in Essential Tremor. Neuromodulation 2018, 22, 537–545. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  71. Isaacson, S.H.; Peckham, E.; Tse, W.; Waln, O.; Way, C.; Petrossian, M.T.; Dahodwala, N.; Soileau, M.J.; Lew, M.; Dietiker, C.; et al. Prospective Home-use Study on Non-invasive Neuromodulation Therapy for Essential Tremor. Tremor Other Hyperkinetic Mov. 2020, 10, 29. [Google Scholar] [CrossRef]
  72. Elek, J.; Prochazka, A. Attenuation of wrist tremor with closed-loop electrical stimulation of muscles. J. Physiol. 1989, 414, 17P. [Google Scholar]
  73. Javidan, M.; Elek, J.; Prochazka, A. Tremor reduction by functional electrical stimulation. Neurology 1990, 40, 369. [Google Scholar]
  74. Prochazka, A.; Elek, J.; Javidan, M. Attenuation of pathological tremors by functional electrical stimulation I: Method. Ann. Biomed. Eng. 1992, 20, 205–224. [Google Scholar] [CrossRef] [Green Version]
  75. Javidan, M.; Elek, J.; Prochazka, A. Attenuation of pathological tremors by functional electrical stimulation II: Clinical evaluation. Ann. Biomed. Eng. 1992, 20, 225–236. [Google Scholar] [CrossRef] [PubMed]
  76. Law, J.J. Reduction of Pathological Tremor by Functional Electrical Stimulation Using Digital Feedback Control. Master’s Thesis, University of Alberta, Edmonton, AB, Canada, 1995. [Google Scholar]
  77. Gillard, D.; Cameron, T.; Prochazka, A.; Gauthier, M. Tremor suppression using functional electrical stimulation: A comparison between digital and analog controllers. IEEE Trans. Rehabil. Eng. 1999, 7, 385–388. [Google Scholar] [CrossRef]
  78. Dideriksen, J.L.; Laine, C.M.; Dosen, S.; Muceli, S.; Rocon, E.; Pons, J.L.; Benito-Leon, J.; Farina, D. Electrical Stimulation of Afferent Pathways for the Suppression of Pathological Tremor. Front. Neurosci. 2017, 11, 178. [Google Scholar] [CrossRef]
  79. Maneski, L.P.; Jorgovanović, N.; Ilić, V.; Došen, S.; Keller, T.; Popović, M.B.; Popović, D.B. Electrical stimulation for the suppression of pathological tremor. Med. Biol. Eng. Comput. 2011, 49, 1187–1193. [Google Scholar] [CrossRef]
  80. Popović, D.; Popović-Maneski, L. The Instrumented Shoe Insole for Rule-Based Control of Gait in Persons with Hemiplegia. EasyChair Prepr. 2019, p. 1345. Available online: https://easychair.org/publications/preprint/ZF3Z (accessed on 26 March 2021).
  81. Gallego, J.Á.; Rocon, E.; Belda-Lois, J.M.; Pons, J.L. A neuroprosthesis for tremor management through the control of muscle co-contraction. J. Neuroeng. Rehabil. 2013, 10, 36. [Google Scholar] [CrossRef] [Green Version]
  82. Jitkritsadakul, O.; Thanawattano, C.; Anan, C.; Bhidayasiri, R. Tremor’s glove-an innovative electrical muscle stimulation therapy for intractable tremor in Parkinson’s disease: A randomized sham-controlled trial. J. Neurol. Sci. 2017, 381, 331–340. [Google Scholar] [CrossRef] [PubMed]
  83. Meunier, S.; Pol, S.; Houeto, J.L.; Vidailhet, M. Abnormal reciprocal inhibition between antagonist muscles in Parkinson’s disease. Brain 2000, 123, 1017–1026. [Google Scholar] [CrossRef] [Green Version]
  84. Heo, J.-H.; Jeon, H.-M.; Choi, E.-B.; Kwon, D.-Y.; Eom, G.-M. Effect of Sensory Electrical Stimulation on Resting Tremors in Patients with Parkinson’s Disease and SWEDDS. J. Mech. Med. Biol. 2019, 19, 1940033. [Google Scholar] [CrossRef] [Green Version]
  85. Bickel, C.S.; Gregory, C.M.; Dean, J.C. Motor unit recruitment during neuromuscular electrical stimulation: A critical appraisal. Graefe’s Arch. Clin. Exp. Ophthalmol. 2011, 111, 2399–2407. [Google Scholar] [CrossRef] [PubMed]
  86. Martin, R.; Sadowsky, C.; Obst, K.; Meyer, B.; McDonald, J. Functional Electrical Stimulation in Spinal Cord Injury: From Theory to Practice. Top. Spinal Cord Inj. Rehabil. 2012, 18, 28–33. [Google Scholar] [CrossRef] [Green Version]
  87. Terry, T.E.; Hoyt, L.J., Sr. Cerebral Palsy Arm and Hand Brace. U.S. Patent 4,237,873, 9 December 1980. [Google Scholar]
  88. Kotovsky, J.; Rosen, M.J. A wearable tremor-suppression orthosis. J. Rehabil. Res. Dev. 1998, 35, 373–387. [Google Scholar] [PubMed]
  89. Rocon, E.; Ruiz, A.; Pons, J.L.; Belda-Lois, J.; Sánchez-Lacuesta, J. Rehabilitation Robotics: A Wearable Exo-Skeleton for Tremor Assessment and Suppression. In Proceedings of the Proceedings of the 2005 IEEE International Conference on Robotics and Automation, Barcelona, Spain, 18–22 April 2005. [Google Scholar]
  90. Rocon, E.; Ruiz, A.; Brunetti, F.; Pons, J.L.; Belda-Lois, J.; Sánchez-Lacuesta, J. On the use of an active wearable exoskeleton for tremor suppression via biomechanical loading. In Proceedings of the 2006 IEEE International Conference on Robotics and Automation, Orlando, FL, USA, 15–19 May 2006. [Google Scholar]
  91. Manto, M.; Rocon, E.; Pons, J.L.; Belda, J.M.; Camut, S. Evaluation of a wearable orthosis and an associated algorithm for tremor suppression. Physiol. Meas. 2007, 28, 415–425. [Google Scholar] [CrossRef]
  92. Rocón, E.; Belda-Lois, J.M.; Ruiz, A.; Manto, M.; Moreno, J.C.; Pons, J.L. Design and Validation of a Rehabilitation Robotic Exoskeleton for Tremor Assessment and Suppression. IEEE Trans. Neural Syst. Rehabil. Eng. 2007, 15, 367–378. [Google Scholar] [CrossRef] [Green Version]
  93. Taheri, B.; Case, D.; Richer, E. Active Tremor Estimation and Suppression in Human Elbow Joint. In Proceedings of the ASME 2011 Dynamic Systems and Control Conference, Arlington, VI, USA, 31 October–2 November 2011. [Google Scholar]
  94. Taheri, B.; Case, D.; Richer, E. Robust Controller for Tremor Suppression at Musculoskeletal Level in Human Wrist. IEEE Trans. Neural Syst. Rehabil. Eng. 2013, 22, 379–388. [Google Scholar] [CrossRef]
  95. Taheri, B.; Case, D.; Richer, E. Adaptive Suppression of Severe Pathological Tremor by Torque Estimation Method. IEEE/ASME Trans. Mechatron. 2014, 20, 717–727. [Google Scholar] [CrossRef]
  96. Zamanian, A.H.; Richer, E. Adaptive disturbance rejection controller for pathological tremor suppression with permanent magnet linear motor. In Proceedings of the ASME 2017 Dynamic Systems and Control Conference, Tysons, VI, USA, 11–13 October 2017. [Google Scholar]
  97. Herrnstadt, G.; Menon, C. Voluntary-Driven Elbow Orthosis with Speed-Controlled Tremor Suppression. Front. Bioeng. Biotechnol. 2016, 4, 29. [Google Scholar] [CrossRef] [Green Version]
  98. Zhou, Y.; Naish, M.D.; Jenkins, M.E.; Trejos, A.L. Design and validation of a novel mechatronic transmission system for a wearable tremor suppression device. Robot. Auton. Syst. 2017, 91, 38–48. [Google Scholar] [CrossRef]
  99. Ando, T.; Watanabe, M.; Fujie, M.G. Extraction of voluntary movement for an EMG controlled exoskeltal robot of tremor patients. In Proceedings of the 2009 4th International IEEE/EMBS Conference on Neural Engineering, Antalya, Turkey, 29 April–2 May 2009. [Google Scholar]
  100. Seki, M.; Matsumoto, Y.; Ando, T.; Kobayashi, Y.; Fujie, M.G.; Iijima, H.; Nagaoka, M. Development of robotic upper limb orthosis with tremor suppressiblity and elbow joint movability. In Proceedings of the 2011 IEEE International Conference on Systems, Man, and Cybernetics, Anchorage, AK, USA, 9–12 October 2011. [Google Scholar]
  101. Seki, M.; Matsumoto, Y.; Ando, T.; Kobayashi, Y.; Iijima, H.; Nagaoka, M.; Fujie, M.G. The weight load inconsistency effect on voluntary movement recognition of essential tremor patient. In Proceedings of the 2011 IEEE International Conference on Robotics and Biomimetics, Karon Beach, Thailand, 7–11 December 2011. [Google Scholar]
  102. Ando, T.; Watanabe, M.; Nishimoto, K.; Matsumoto, Y.; Seki, M.; Fujie, M.G. Myoelectric-Controlled Exoskeletal Elbow Robot to Suppress Essential Tremor: Extraction of Elbow Flexion Movement Using STFTs and TDNN. J. Robot. Mechatron. 2012, 24, 141–149. [Google Scholar] [CrossRef]
  103. Matsumoto, Y.; Amemiya, M.; Kaneishi, D.; Nakashima, Y.; Seki, M.; Ando, T.; Kobayashi, Y.; Iijima, H.; Nagaoka, M.; Fujie, M.G. Development of an elbow-forearm interlock joint mechanism toward an exoskeleton for patients with essential tremor. In Proceedings of the 2014 IEEE/RSJ International Conference on Intelligent Robots and Systems, Chicago, IL, USA, 14–18 September 2014. [Google Scholar]
  104. Matsumoto, Y.; Seki, M.; Ando, T.; Kobayashi, Y.; Nakashima, Y.; Iijima, H.; Nagaoka, M.; Fujie, M.G. Development of an Exoskeleton to Support Eating Movements in Patients with Essential Tremor. J. Robot. Mechatron. 2013, 25, 949–958. [Google Scholar] [CrossRef]
  105. Huen, D.; Liu, J.; Lo, B. An integrated wearable robot for tremor suppression with context aware sensing. In Proceedings of the 2016 IEEE 13th International Conference on Wearable and Implantable Body Sensor Networks (BSN), San Francisco, CA, USA, 14–17 June 2016. [Google Scholar]
  106. Fromme, N.P.; Camenzind, M.; Riener, R.; Rossi, R.M. Need for mechanically and ergonomically enhanced tremor-suppression orthoses for the upper limb: A systematic review. J. Neuroeng. Rehabil. 2019, 16, 1–15. [Google Scholar] [CrossRef] [Green Version]
  107. Loureiro, R.; Belda-Lois, J.M.; Lima, E.; Pons, J.; Sanchez-Lacuesta, J.; Harwin, W.S. Upper Limb Tremor Suppression in ADL Via an Orthosis Incorporating a Controllable Double Viscous Beam Actuator. In Proceedings of the 9th International Conference on Rehabilitation Robotics, Chicago, IL, USA, 28 June–1 July 2005. [Google Scholar]
  108. Case, D.; Taheri, B.; Richer, E. Dynamic Magnetorheological Damper for Othotic Tremor Suppression. In Proceedings of the 2011 Hawaii University International Conference on Mathematics and Engineering, Honolulu, HI, USA, 13–15 June 2011. [Google Scholar]
  109. Case, D.; Taheri, B.; Richer, E. Multiphysics modeling of magnetorheological dampers. Int. J. Multiphysics 2013, 7, 61–76. [Google Scholar] [CrossRef]
  110. Case, D.; Taheri, B.; Richer, E. A Lumped-Parameter Model for Adaptive Dynamic MR Damper Control. IEEE ASME Trans. Mechatron. 2014, 20, 1–8. [Google Scholar] [CrossRef]
  111. Case, D.; Taheri, B.; Richer, E. Active control of MR wearable robotic orthosis for pathological tremor suppression. In Proceedings of the ASME 2015 Dynamic Systems and Control Conference, Columbus, OH, USA, 28–30 October 2015. [Google Scholar]
  112. Zahedi, A.; Zhang, B.; Yi, A.; Zhang, D. A Soft Exoskeleton for Tremor Suppression Equipped with Flexible Semiactive Actuator. Soft Robot. 2020. [Google Scholar] [CrossRef] [PubMed]
  113. Herrnstadt, G.; Menon, C. On-Off Tremor Suppression Orthosis with Electromagnetic Brake. Int. J. Mech. Eng. Mechatron. 2013, 1, 7–14. [Google Scholar] [CrossRef] [Green Version]
  114. Kalaiarasi, A.; Kumar, L.A. Sensor Based Portable Tremor Suppression Device for Stroke Patients. Electrother. Res. 2018, 43, 29–37. [Google Scholar] [CrossRef]
  115. Rudraraju, S.; Nguyen, T. Wearable Tremor Reduction Device (TRD) for Human Hands and Arms. In Proceedings of the 2018 Design of Medical Devices Conference, Minneapolis, MI, USA, 9–12 April 2018. [Google Scholar]
  116. Elias, M.; Patel, S.; Maamary, E.; Araneta, L.; Obaid, N. Apparatus for Damping Involuntary Hand Motions. U.S. Patent 0,216,628 A1, 18 July 2019. [Google Scholar]
  117. Hunter, R.; Pivach, L.; Madere, K.; Van Gemmert, A.W.A. Potential benefits of the Readi-Steadi on essential tremor. In Proceedings of the 5th Annual LSU Discover Day, Baton Rouge, LA, USA, 10 April 2018. [Google Scholar]
  118. Paulig, J.; Jabusch, H.-C.; Groãÿbach, M.; Boullet, L.; Altenmüller, E. Sensory trick phenomenon improves motor control in pianists with dystonia: Prognostic value of glove-effect. Front. Psychol. 2014, 5, 1012. [Google Scholar] [CrossRef] [Green Version]
  119. Fromme, N.P.; Camenzind, M.; Riener, R.; Rossi, R.M. Design of a lightweight passive orthosis for tremor suppression. J. Neuroeng. Rehabil. 2020, 17, 1–15. [Google Scholar] [CrossRef] [Green Version]
  120. Lu, Z.; Huang, Z. Analytical and experimental studies on particle damper used for tremor suppression. J. Vib. Control 2020, 1–11. [Google Scholar] [CrossRef]
  121. Katz, R.; Buki, E.; Zacksenhouse, M. Attenuating Tremor Using Passive Devices. Stud. Health Technol. Inform. 2017, 242, 741–747. [Google Scholar]
  122. Buki, E.; Katz, R.; Zacksenhouse, M.; Schlesinger, I. Vib-bracelet: A passive absorber for attenuating forearm tremor. Med Biol. Eng. Comput. 2017, 56, 923–930. [Google Scholar] [CrossRef] [PubMed]
  123. Takanokura, M.; Sugahara, R.; Miyake, N.; Ishiguro, K.; Muto, T.; Sakamoto, K. Upper-limb orthoses implemented with air dashpots for suppression of pathological tremor in daily activities. In Proceedings of the 23rd Congress of International Society of Biomechanics, Brussels, Belgium, 2–3 July 2011. [Google Scholar]
  124. Lusardi, M.M. Tremor, chorea and other involuntary movement. In Geriatric Rehabilitation Manual, 2nd ed.; Kauffman, T., Barr, J., Moran, M., Eds.; Elsevier: New York, NY, USA, 2007; pp. 215–225. [Google Scholar]
  125. Kelley, C.R.; Kauffman, J.L. Tremor-Active Controller for Dielectric Elastomer-Based Pathological Tremor Suppression. IEEE/ASME Trans. Mechatron. 2020, 25, 1143–1148. [Google Scholar] [CrossRef]
  126. Kelley, C.R.; Kauffman, J.L. Scaled Tremor Suppression with Folded Dielectric Elastomer Stack Actuators. In Electroactive Polymer Actuators and Devices (EAPAD) XXII; International Society for Optics and Photonics: Bellingham, WA, USA, 2020. [Google Scholar] [CrossRef]
  127. Michaelis, J. Introducing the neater eater. Action Res. 1988, 6, 2–3. [Google Scholar]
  128. Mandy, A.; Sims, T.; Stew, G.; Onions, D. Manual Feeding Device Experiences of People with a Neurodisability. Am. J. Occup. Ther. 2018, 72, 7203345010p1–7203345010p5. [Google Scholar] [CrossRef]
  129. Pathak, A.; Redmond, J.A.; Allen, M.; Chou, K.L. A noninvasive handheld assistive device to accommodate essential tremor: A pilot study. Mov. Disord. 2013, 29, 838–842. [Google Scholar] [CrossRef] [Green Version]
  130. Sabari, J.; Stefanov, D.G.; Chan, J.; Goed, L.; Starr, J. Adapted Feeding Utensils for People With Parkinson’s-Related or Essential Tremor. Am. J. Occup. Ther. 2019, 73, 7302205120p1–7302205120p9. [Google Scholar] [CrossRef] [PubMed]
  131. Zhu, Y.; Ren, K. Arm Vibration Damping Device. U.S. Patent 0,327,023 A1, 16 November 2017. [Google Scholar]
  132. De Panisse, P.; Ibrahim, Y.; Medeisis, J.; Tiarvando, L.; Vaklev, N.L.; Ong, J.F.; Gan, B.; Koh, B.; Soler, X.L.; Choong Ngan Lou, W. Tremor Stabilization Apparatus and Methods. U.S. Patent 0266820 A1, 20 September 2018. [Google Scholar]
  133. Hall, W.D. Hand-Held Gyroscopic Device. U.S. Patent 5,058,571 A, 22 October 1991. [Google Scholar]
  134. Kalvert, M.A. Adjustable and Tunable Hand Tremor Stabilizer. U.S. Patent 6,730,049 B2, 4 May 2004. [Google Scholar]
  135. Ferrington, D.G.; Nail, B.S.; Rowe, M. Human tactile detection thresholds: Modification by inputs from specific tactile receptor classes. J. Physiol. 1977, 272, 415–433. [Google Scholar] [CrossRef]
  136. Douglas, P.R.; Ferrington, D.G.; Rowe, M. Coding of information about tactile stimuli by neurones of the cuneate nucleus. J. Physiol. 1978, 285, 493–513. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  137. Uemura, Y.; Haque, T.; Sato, F.; Tsutsumi, Y.; Ohara, H.; Oka, A.; Furuta, T.; Bae, Y.C.; Yamashiro, T.; Tachibana, Y.; et al. Proprioceptive thalamus receiving forelimb and neck muscle spindle inputs via the external cuneate nucleus in the rat. Brain Struct. Funct. 2020, 225, 2177–2192. [Google Scholar] [CrossRef]
  138. Tracey, D.J. The projection of joint receptors to the cuneate nucleus in the cat. J. Physiol. 1980, 305, 433–449. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  139. Lora-Millán, J.S.; López-Blanco, R.; Gallego, J.Á.; Méndez-Guerrero, A.; de La Aleja, J.G.; Rocon, E. Mechanical vibration does not systematically reduce the tremor in essential tremor patients. Sci. Rep. 2019, 9, 1–11. [Google Scholar] [CrossRef] [Green Version]
  140. Louis, E.D. The Roles of Age and Aging in Essential Tremor: An Epidemiological Perspective. Neuroepidemiology 2019, 52, 111–118. [Google Scholar] [CrossRef]
  141. Louis, E.D.; Ford, B.; Barnes, L.F. Clinical subtypes of essential tremor. Arch. Neurol. 2000, 57, 1194–1198. [Google Scholar] [CrossRef] [Green Version]
  142. Benito-León, J.; Louis, E.D.; Bermejo-Pareja, F. Elderly-onset essential tremor is associated with dementia. Neurology 2006, 66, 1500–1505. [Google Scholar] [CrossRef] [PubMed]
  143. Hanagasi, H.A.; Tufekcioglu, Z.; Emre, M. Dementia in Parkinson’s disease. J. Neurol. Sci. 2017, 374, 26–31. [Google Scholar] [CrossRef]
  144. Park, I.-S.; Oh, Y.-S.; Lee, K.-S.; Yang, D.-W.; Song, I.-U.; Park, J.-W.; Kim, J.-S. Subtype of Mild Cognitive Impairment in Elderly Patients with Essential Tremor. Alzheimer Dis. Assoc. Disord. 2015, 29, 141–145. [Google Scholar] [CrossRef] [PubMed]
  145. Benito-León, J.; Louis, E.D.; Mitchell, A.J.; Bermejo-Pareja, F. Elderly-Onset Essential Tremor and Mild Cognitive Impairment: A Population-Based Study (NEDICES). J. Alzheimer’s Dis. 2011, 23, 727–735. [Google Scholar] [CrossRef]
  146. Monastero, R.; Cicero, C.E.; Baschi, R.; Davì, M.; Luca, A.; Restivo, V.; Zangara, C.; Fierro, B.; Zappia, M.; Nicoletti, A. Mild cognitive impairment in Parkinson’s disease: The Parkinson’s disease cognitive study (PACOS). J. Neurol. 2018, 265, 1050–1058. [Google Scholar] [CrossRef] [PubMed]
  147. Frost & Sullivan. Assessing the Full Impact of Essential Tremor on Patient Quality of Life and Finances in the United States. 2018. Available online: https://www.insightec.com/media/1550/fs_wp_insightec-et_010819.pdf (accessed on 20 September 2020).
  148. Axelrad, J.E.; Louis, E.D.; Honig, L.S.; Flores, I.; Ross, G.W.; Pahwa, R.; Lyons, K.E.; Faust, P.L.; Vonsattel, J.P.G. Reduced Purkinje Cell Number in Essential Tremor. Arch. Neurol. 2008, 65, 101–107. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  149. Louis, E.D.; Lee, M.; Babij, R.; Ma, K.; Cortés, E.; Vonsattel, J.-P.G.; Faust, P.L. Reduced Purkinje cell dendritic arborization and loss of dendritic spines in essential tremor. Brain 2014, 137, 3142–3148. [Google Scholar] [CrossRef] [Green Version]
  150. Helmich, R.C.; Janssen, M.J.R.; Oyen, W.J.G.; Bloem, B.R.; Toni, I. Pallidal dysfunction drives a cerebellothalamic circuit into Parkinson tremor. Ann. Neurol. 2011, 69, 269–281. [Google Scholar] [CrossRef] [PubMed]
  151. Espay, A.J.; Lang, A.E.; Erro, R.; Merola, A.; Fasano, A.; Berardelli, A.; Bhatia, K.P. Essential pitfalls in “essential” tremor. Mov. Disord. 2017, 32, 325–331. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Cala Trio transcutaneous electrical nerve stimulation. The median and radial nerves, which project to the ventral intermediate nucleus of the thalamus, are stimulated by Cala Trio (Cala Health, Burlingame, CA, USA) through two working electrodes placed on the anterior surface of the wrist.
Figure 1. Cala Trio transcutaneous electrical nerve stimulation. The median and radial nerves, which project to the ventral intermediate nucleus of the thalamus, are stimulated by Cala Trio (Cala Health, Burlingame, CA, USA) through two working electrodes placed on the anterior surface of the wrist.
Biosensors 11 00099 g001
Figure 2. MOTIMOVE functional electrical stimulation system. This device (3F-Fit Fabricando Faber, Serbia) comprises a multichannel stimulator that attaches to several electrodes placed on the flexor and extensor muscles of the forearm, enabling muscle activation.
Figure 2. MOTIMOVE functional electrical stimulation system. This device (3F-Fit Fabricando Faber, Serbia) comprises a multichannel stimulator that attaches to several electrodes placed on the flexor and extensor muscles of the forearm, enabling muscle activation.
Biosensors 11 00099 g002
Figure 3. Tremelo passive orthosis. The vibration energy of tremor is being transferred to a mass-spring-damper system within the two tuned vibration absorbers of Tremelo (Five Microns, Fresno, CA, USA), positioned over the dorsal and ventral surfaces of the arm to counteract the involuntary motions.
Figure 3. Tremelo passive orthosis. The vibration energy of tremor is being transferred to a mass-spring-damper system within the two tuned vibration absorbers of Tremelo (Five Microns, Fresno, CA, USA), positioned over the dorsal and ventral surfaces of the arm to counteract the involuntary motions.
Biosensors 11 00099 g003
Figure 4. Steadi-One passive orthosis. The non-Newtonian fluid within the tuned mass damper of Steadi-One (Steadiwear, Toronto, ON, Canada) becomes viscous when the vibration energy of tremor is transferred to a mass-spring-damper system, acting as a dissipating element that reduces the amplitude of motions.
Figure 4. Steadi-One passive orthosis. The non-Newtonian fluid within the tuned mass damper of Steadi-One (Steadiwear, Toronto, ON, Canada) becomes viscous when the vibration energy of tremor is transferred to a mass-spring-damper system, acting as a dissipating element that reduces the amplitude of motions.
Biosensors 11 00099 g004
Table 1. Summary of the tremor suppression devices and study results.
Table 1. Summary of the tremor suppression devices and study results.
Type of DeviceStudy Participants (n)EfficacyRisksRefs
Electrical Stimulation Systems: Transcutaneous Electrical Nerve Stimulators
Cala ONE ET (77)
  • Improved upper limb TETRAS tremor scores (p = 0.017)
  • Improved subject-rated BF-ADL scores (p = 0.001)
  • Skin irritations (redness, itchiness, and swelling)
  • Soreness or lesions
  • Discomfort (stinging and sensation of weakness) or burns
[70]
Cala Trio *ET (205)
  • Improved upper limb TETRAS tremor scores (p < 0.0001)
  • Improved subject-rated BF-ADL scores (p < 0.0001)
[71]
Electrical Stimulation Systems: Functional Electrical Stimulators
MOTIMOVEET (3); PD tremor (4)67% tremor suppressionMuscle fatigue[79]
TREMOR neurorobotET (4); PD tremor (2)52% tremor suppression[81]
Tremor’s glovePD tremor (30)Reduced UPDRS score (p = 0.001)[82]
Wearable Orthoses: Active Orthoses
WOTAS exoskeletonET (7); MS tremor (1); Posttraumatic tremor (1); Mixed tremor (1)40% tremor suppression [92]Not reported[89,90,91,92]
Pneumatic actuator-based orthosisET (5) §; PD tremor (5) §98.1% tremor suppression [95][93,94,95]
PMLM-based orthosisPD tremor (5) §97.6% tremor suppression[96]
Voluntary-driven elbow orthosisET (1) §99.8% tremor suppression[97]
MMS-based WTSGNot reportedNot reported[98]
Myoelectric-controlled orthosisET (2); Healthy (4)Not reported[99,100,101,102]
Myoelectric-controlled orthosis (ver. 2)Healthy (1)50–80% tremor suppression [104][103,104]
BSN-based orthosisHealthy (6) §77% tremor suppression[105]
Wearable Orthoses: Semi-Active Orthoses
Double viscous beam orthosisNot reportedNot reportedNot reported[107]
MR damper-based orthosisNot reportedNot reported[108,109,110,111]
SETS systemNot reportedNot reported[112]
Electromagnetic brake-based orthosisHealthy (3) §88% tremor suppression[113]
Pneumatic hand cuffET (1)30% tremor suppression[114]
Wearable Orthoses: Passive Orthoses
Tremelo *PD tremor (1)85% tremor suppressionNot reported[115]
Steadi-One *Lab simulation85–90% tremor suppression[116]
Readi-Steadi *ET (20); Healthy (40)50% tremor suppression[117]
Task-Adjustable Passive OrthosisPD tremor (1)
  • 82% tremor suppression while drinking (p = 0.03)
  • 79% tremor suppression while pouring (p = 0.03)
  • 74% tremor suppression while drawing a spiral (p = 0.03)
[119]
Particle DamperNot reported Not reported[120]
Vib-BraceletPD tremor (1) §85% tremor suppression[121,122]
Air-dashpot-based orthosisHealthy (1)
  • 20–62% tremor suppression in the wrist
  • 82% tremor suppression in the elbow
[123]
Assistive Feeding Devices
Neater Eater *Not reportedNot reportedNot reported[127]
Liftware Steady *ET (15)
  • Improved FTM-TRS while holding, eating, and transferring objects (p = 0.001)
  • 73% tremor suppression
[129]
Gyenno Spoon *Not reported85% tremor suppression (claimed)[131]
Gyroscopic Stabilizers
GyroGlove *Not reportedNot reportedNot reported[132]
Haptic Stimulation Systems
Emme WatchNot reportedNot reportedNot reported
BF-ADL, Bain and Findley Activities of Daily Living; BSN, body senor network; ET, essential tremor; FTM-TRS, Fahn-Tolosa-Marin Tremor Rating Scale; MMS, multi-channel mechatronic splitter; MS, multiple sclerosis; PMLM, permanent magnet linear motor; SETS, soft exoskeleton for tremor suppression; TETRAS, Tremor Research Group Essential Tremor Rating Assessment Scale; PD, Parkinson’s disease; UPDRS, Unified Parkinson’s Disease Rating Scale; WOTAS, Wearable Orthosis for Tremor Assessment and Suppression; and WTSG, wearable tremor suppression gloves. * FDA-registered; Class I medical device. FDA-approved; Class II medical device. § Test bench simulation. Induced muscle contraction.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Mo, J.; Priefer, R. Medical Devices for Tremor Suppression: Current Status and Future Directions. Biosensors 2021, 11, 99. https://doi.org/10.3390/bios11040099

AMA Style

Mo J, Priefer R. Medical Devices for Tremor Suppression: Current Status and Future Directions. Biosensors. 2021; 11(4):99. https://doi.org/10.3390/bios11040099

Chicago/Turabian Style

Mo, Jiancheng, and Ronny Priefer. 2021. "Medical Devices for Tremor Suppression: Current Status and Future Directions" Biosensors 11, no. 4: 99. https://doi.org/10.3390/bios11040099

APA Style

Mo, J., & Priefer, R. (2021). Medical Devices for Tremor Suppression: Current Status and Future Directions. Biosensors, 11(4), 99. https://doi.org/10.3390/bios11040099

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.