Article

Applicability of Selected 3D Printing Materials in Electrochemistry

1
J. Heyrovský Institute of Physical Chemistry of the Czech Academy of Sciences, Dolejškova 3, 182 23 Prague, Czech Republic
2
Department of Analytical Chemistry, Faculty of Science, Charles University, Albertov 6, 128 00 Prague, Czech Republic
3
Department of Chemistry Education, Faculty of Science, Charles University, Albertov 6, 128 00 Prague, Czech Republic
*
Author to whom correspondence should be addressed.
Biosensors 2022, 12(5), 308; https://doi.org/10.3390/bios12050308
Submission received: 11 April 2022 / Revised: 28 April 2022 / Accepted: 30 April 2022 / Published: 7 May 2022

Abstract

:
This manuscript investigates the chemical and structural stability of 3D printing materials (3DPMs) frequently used in electrochemistry. Four 3D printing materials were studied: Clear photopolymer, Elastic photopolymer, PET filament, and PLA filament. Their stability, solubility, structural changes, flexibility, hardness, and color changes were investigated after exposure to selected organic solvents and supporting electrolytes. Furthermore, the available potential windows and behavior of redox probes in selected supporting electrolytes were investigated before and after the exposure of the 3D-printed objects to the electrolytes at various working electrodes. Possible electrochemically active interferences with an origin from the 3DPMs were also monitored to provide a comprehensive outline for the use of 3DPMs in electrochemical platform manufacturing.

1. Introduction

The application of 3D printing is steadily spreading into electrochemistry due to the efficiency it provides to the manufacturing of cells, electrode parts, and even the electrodes themselves, particularly in low-scale production [1]. The printing process is precise, easy, fast, often inexpensive, safe, and allows for versatile production and product modifications [2,3,4,5,6,7].
Various 3D printing techniques have been developed over time. The most common is fused deposition modeling (FDM), based on the protruding a fiber of the thermoplastic material through a heated nozzle [4,5,8]; the melted extrusion is deposited on the growing object in a specified pattern [4,8,9,10]. The most popular thermoplastic materials are polylactic acid (PLA), acrylonitrile butadiene styrene (ABS), acrylonitrile styrene acrylate (ASA), nylon and other polyamides, polypropylene (PP), polycarbonate, or polyetherimide [2,9,11]. Polymers can also be mixed with another material before forming into the filament. This process might help to reinforce the material, but the possibility to add an electrically conductive component (e.g., graphite or graphene) to obtain a conductive printing material [1,7,12,13] is particularly interesting for electrochemistry. Another common 3D printing method is stereolithography (SLA), in which a liquid polymer precursor is photopolymerized using UV or visible radiation. The light beam moves according to the project instructions and polymerizes the material. Alternatively, a diode array as a light source is used for digital light processing (DLP) with the irradiation carried out by particular switchable pixels in the diode field [14,15,16]. SLA uses printing materials with radical polymerization, such as acrylates, vinyl ethers, or epoxy resins. Often a mixture of the various monomers is used to reach the desired properties of the final polymeric material. In comparison with FDM, SLA provides structures with a better resolution, but it cannot combine several printing materials during the printing process [14,17,18]. The disadvantage of SLA and DLP is that printed objects require rinsing by organic solvents, which makes the manufacturing process costly and waste-producing.
Regardless of the used printing technique, its application for electrochemistry requires sufficient chemical stability of the employed materials with regard to the printing conditions and minimum amount of soluble, electrochemically active interferences. Choosing suitable 3D printing materials (3DPMs) to print electrochemical cells and systems is essential to avoid their undesirable changes because the properties of the materials may be affected by long-term exposure to various common solvents, chemical compounds, or extreme pH values [19,20]. The damage or disintegration of the printed objects is the usual result of a careless application; however, some of these processes can be utilized to pretreat or regenerate electrodes by the selective etching of the polymeric material. On the other hand, the printing material can affect the electrochemical properties of the electrolyte/electrode system, e.g., the influence on the oxygen or hydrogen evolution reactions. In addition, the chemical compatibility may differ in the cases of original virgin filament, the printed object, and may even be different from supplementary information provided by the manufacturers and vendors, as shown by Heikkinen et al. [21]. In any case, interactions between the 3DPMs and chemicals used during the experiments should be understood and characterized before printing an object or starting an experiment.
So far, the properties of composite electrodes with glassy carbon fibers [19], PLA [22], and ABS [23] were mainly explored regarding their dependence on the used printing parameters [24,25], fiber modification [26,27], and post-printing treatment [22,28,29,30]. While properties of 3D printed electrodes have been investigated extensively, significantly less attention was paid to insulating materials to be used for other cell parts.
This paper focuses on the comparison of the stability of four 3DPMs and the change of potential windows of electrolytes after 24 h of contact with these materials. The four materials were chosen as the most popular ones for fabrication of particular parts of an electrochemical cell or experimental platform: PLA as a cheap, biodegradable, and the most common material for fabrication of conductive parts (mixed with carbon or metallic additives), PET as a cheap and durable material for electrically non-conductive parts, and two resins: Elastic for the fabrication of mechanically stressed parts and Clear for the transparent parts. The investigation includes the physical changes of these filaments and photopolymerized resins after exposure to alkaline, acidic, and organic solutions [31].
Furthermore, changes in potential windows in such electrolytes after exposure to the 3DPMs are compared using hanging mercury, glassy carbon, and 3D-printed PLA/carbon black electrodes. Finally, the influence on peak potentials and peak currents of model analytes or redox probes after exposure to these 3DPMs will be investigated at the selected electrodes. This will provide a better understanding of the various effects of 3DPMs in electrochemical experiments, allowing a knowledge-based way to select a proper fiber or resin for the desired application.

2. Experimental

2.1. 3D Printed Tested Objects

Four non-conductive 3DPMs were used for the exposure tests. SLA resins were: Elastic (Formlabs, Somerville, MA, USA)—the resin of acrylate monomer, and Clear (Formlabs, Somerville, MA, USA)—a resin made of a mixture of dimethacrylate, methacrylate monomer, and diphenyl(2,4,6-trimethylbenzoyl)phosphine oxide; FDM filaments were PLA Extrafill TrafficWhite (Fillamentum, Hulín, Czech Republic) and EPR InnoPET Natural (Innofil3D, Emmen, Netherlands).
The tested objects for the mechanical stability testing were prepared in the shape of a square plate (dimensions 10 × 10 × 1 mm) and for the voltammetric measurement in the shape of a “7-wall accordion” on a circular base (diameter 15 mm, height 30 mm, wall width 1.0 mm), as shown in Figure 1. The accordion shape was chosen to maximize the surface to study possible electrochemically active interferents that can be dissolved from selected material.
The FDM printer was an original Prusa i3 MK3S+MMU2S (Prusa research, CZ) with Spring Steel Sheet with Smooth Double-sided PEI and E3D V6 0.4 nozzle (0.4 mm diameter, all Prusa research, CZ). Printing parameters for PET and PLA were 200 μm layer thickness, bed temperature 60 °C, and extruder temperature 215 °C and 210 °C for the first and the following layers, respectively. No brims or additional treatment after printing were used for the FMD-printed objects.
Stereolithographic resins Elastic and Clear were used in the SLA printer Formlabs Form 3 (Formlabs, USA), with manufacturer recommended settings: 405 nm LED light, intensity 1.25 mW cm−2, and 100 μm layer thickness. Afterward, they were cleaned two times with isopropanol, left to dry under the ambient temperature, and hardened in Formlabs Cure (Formlabs, USA) at 60 °C for 30 and 20 min, respectively.

2.2. Chemicals

The solvents used for solubility and stability tests were: acetone, acetonitrile (ACN), chloroform, cyclohexanone, dichloromethane, diethylether, dimethyl sulfoxide, ethanol, ethyl acetate, formaldehyde, hexane, isopropanol, methanol (MeOH), 1-penthanol, tetrahydrofuran, toluene, (all Penta, Prague, Czech Republic), and N,N-dimethyl formamide (Sigma Aldrich, MO, USA).
Aqueous 1 mol dm−3 solutions of inorganic electrolytes were prepared from H2SO4, NaOH, and KCl (all Lachema, Brno, Czech Republic). A mixed aqueous–organic electrolyte was prepared from 1 mol dm−3 KCl and MeOH (1:9, v/v).
K3[Fe(CN)6], Cd(NO3)2 (both Lachema, Brno, Czech Republic), and [Ru(NH3)6]Cl3 (Sigma Aldrich, Prague, Czech Republic) (stock solution concentration 1 × 10−3 mol dm−3) were used as model redoxactive probes. All chemicals were of p.a. purity grade, and aqueous solutions were prepared using Milli-Q-Gradient water (Millipore, Prague, Czech Republic, conductivity < 0.05 μS cm−1).

2.3. Apparatus

The voltammetric signals were recorded by PC ETP (Polaro-Sensors, Prague, Czech Republic) controlled by MultiElChem 3.3 software (J. Heyrovský Institute of Physical Chemistry, Prague, Czech Republic). A three-electrode system was used in all cases with Ag|AgCl|KCl(sat.) as the reference electrode (Monokrystaly, Turnov, Czech Republic) and platinum plate (1 cm2) auxiliary electrode. As the working electrode, three types of electrodes were used: (a) the hanging mercury drop electrode (HMDE, Polaro Sensors, Prague, Czech Republic), drop surface area 1.365 mm2; (b) the glassy carbon electrode (GCE, laboratory-made), disc diameter 3.0 mm, surface area 7.1 mm2, polished on alumina suspension between measurements (particle size 1.1 µm, Elektrochemické detektory, Turnov, Czech Republic); and (c) the laboratory-made electrode from Carbon Fiber PLA (CF-PLA, ProtoPlant, WA, USA), diameter 1.75 mm, surface area 2.4 mm2, with sides insulated by shrinking rubber tube, polished on the sanding paper K/400 (Ciret, Havant, United Kingdom) [4].
Electrochemical pretreatment of GCE and CF-PLA [8,25,26,27,28] was carried out in 3 mol L−1 KCl before each voltammetric experiment, and parameters were set individually for each used electrode type. At GCE, 150 cycles were applied, switching every 0.1 s between potentials 100 mV more negative than the onset of hydrogen evolution reaction (EClean,neg) and 100 mV more positive than the potential of oxygen evolution or material dissolution (EClean,pos), always ending at EClean,neg. At CF-PLA, 10 cycles of CV between EClean,neg, and EClean,pos were applied with a scan rate of 50 mV s−1 [30,32,33]. No pretreatment was applied at the HMDE.
Statistical parameters were calculated using QC Expert software 3.1 (Trilobyte, Pardubice, Czech Republic). Critical level (Cc) was calculated according to IUPAC recommendations as Cc = 1.645 σB, where σB is the standard deviation of the blank noise (n = 5) [23].

2.4. Procedures

The 3D printed models were immersed in 20 mL of the particular tested solvent for 24 h under the ambient temperature. After that time, the solvents were used for voltammetric experiments; measurements were performed in the supporting electrolytes without and with the addition of Cd2+ (1 × 10–3 mol dm−3), [Ru(NH3)6]3+ (1 × 10–3 mol dm−3) or [Fe(CN)6]3 (1 × 10–4 mol dm−3). The solutions were bubbled for 10 min with nitrogen before an experiment with a nitrogen atmosphere over the solution during an experiment. Cyclic voltammetry (CV) was performed with a scan rate of 50 mV s−1. Differential pulse voltammetry (DPV) used a scan rate of 20 mV s−1, a pulse width of 100 ms, and a pulse height of 50 mV. Anodic stripping differential pulse voltammetry (DPASV) was used only for Cd2+ determination with a scan rate of 20 mV s−1, pulse width 100 ms, pulse height 50 mV, and accumulation time 15 s; the accumulation potential varied according to the used solvent and electrode, as summarized in Table 1. All current values were recalculated to current densities (J) or peak current densities (Jp) to compare the registered signals.

2.5. Surface Imaging

The tested 3DPMs objects were inspected by optical microscopy before and after the immersion to selected solvents. The square plates were investigated by the optical microscope (DigiMicroLab 5.0, DNT, Leer, Germany) and NMM800TR Transmitting & Reflecting Metallurgical Microscope (Microteb, Teheran, Iran). Images were taken by Dino-Eye AM4023CT USB C-Mount Microscope Camera (Dino-Lite Digital Microscopes, AnMo Electronics Corporation, New Taipei City, Taiwan) with a magnification of 50×.

3. Results and Discussion

3.1. Effect of Organic Solvents and Supporting Electrolytes on Mechanical and Physical Properties

Initially, the experiments were aimed at the effects of various liquids (organic solvents and inorganic or mixed aqueous–organic electrolytes) on the mechanical and physical stability of 3DPMs objects, including their swelling, which can cause serious distortion of printed objects. Some of the chemical incompatibilities were already studied elsewhere, notably PET in acetone and ethanol; and PLA in various solutions [21,31] under applied mechanical pressure. Our results were fully in agreement, even without the application of the external force. Notably, PLA was delaminated in acetone, and PET was swelled and softened. We have included these experiments anyway for clarity and direct comparison with the other results using additional 18 types of solvents and solutions.
Table 2 summarizes the observed changes of the 3DPMs after 24-h exposure to the selected solvents, including swelling-induced expansion, dissolution, or delamination.
We can see that objects printed from Elastic resin show high stability towards the organic solvents. On the other hand, it suffers from excessive swelling; in some cases, it increases its volume by 50%.
Clear resin and PET showed similar stability. However, they are, to a different extent, vulnerable to chloroform, dichlormethane, tetrahydrofuran, N,N-dimethylformamide, and acetone; the observed swelling is generally accompanied by some other damage. In the case of Clear resin, the breaking seems to be caused by the tension of the irregular swelling; PET is more susceptible to dissolution.
The most significant effect was observed in the case of PLA—half of the solvents caused the dissolution or deformation caused by the delamination of the layers. Although the PLA is a widely available and relatively cheap material, its incompatibility with many commonly used solvents is a significant drawback for some applications.
Equally important, chlorinated solvents are most likely to cause damage to the printing materials, followed by other polar aprotic solvents, e.g., tetrahydrofuran, acetone, and acetonitrile.
Optical microscopy was used to observe the 3DPMs after the immersion in the electrolytes. Some level of disruption was found on all materials after exposition to NaOH solution; in the case of PLA, the damage was macroscopically visible. Clear PET and PLA were also damaged by the solution containing methanol, and PLA by the H2SO4 solution. The complete optical microscopy images are available in the Supplementary Materials as Figures S1–S4.

3.2. Effect of Materials Exposure on the Electrochemical Properties of Electrolytes

The effects of 24-h exposure of the tested 3DPMs on the electrochemical properties of selected electrolytes were investigated by CV. We monitored the presence of electrochemically active interferences relased from the printed objects, and the influence of other species on the potential window, e.g., via hydrogen evolution reaction, oxygen evolution reaction, or by dissolving the 3D printed electrode material.
The anodic or cathodic dissolution/solvent decomposition onset potentials at the end of a potential window (EEnd) were set, when the registered current reached a chosen level (IEnd), set according to the electrode area and background current, with a primary objective to compare potential window changes at a particular electrode type. For GCE, the cathodic IEnd was −50 µA and the anodic IEnd was 40 µA; for HMDE, the cathodic IEnd was −2.0 µA and −10 µA and the anodic IEnd was 0.5 and 2.0 µA in aqueous solutions and mixed organic–aqueous solution, respectively; for CF-PLA, the cathodic IEnd was −2.0 and −20 µA and anodic IEnd was 0.5 and 20 µA in aqueous solutions and mixed organic aqueous solution, respectively. Found potential window widths are shown in Figure 2.
Generally, it is possible to conclude that potential windows before and after exposure to 3DPMs were similar. In the case of HMDE, no significant difference was observed. The changes in the available GCE potential window are mostly negligible except for H2SO4, where depolymerization after H+ attack of PLA [34] and PET led to a rise in pH and the consequent shift of the potential window. A slightly narrower potential window is also observed in NaOH solution after exposure to PET and Clear resin.
CF-PLA exhibits the most distinct difference in the potential window width. Furthermore, narrower potential windows and higher charging currents on CF-PLA were observed in mixed organic–aqueous media. This behavior suggests that the CF-PLA electrode surface was affected by the solvents, dissolving the PLA and thus exposing more electrically conductive carbon fiber, leading to the increase of registered currents. A similar effect was described and employed for a 3D printed electrode activation by Browne et al. [28]. Besides that, a significant change in the potential window width was registered in NaOH solution after the exposure to PLA.

3.3. Effect of Exposure of Supporting Electrolytes to 3DPMs on Voltammetric Behavior of Selected Redox Systems

DPV of the selected simple and well-known redoxactive probes was performed using all three tested electrodes to monitor changes in the voltammetric response after exposure of the selected solvents to 3DPMs. As the selected analytes, [Ru(NH3)6]3+ was employed for all the used electrodes; [Fe(CN)6]3 was used for the electrodes with wide anodic potential range (i.e., GCE and CF-PLA); whilst at HMDE, DPV, and DPASV analyses of Cd2+ ions were performed instead. The 0.1 mol dm−3 NaOH solution was omitted due to the low solubility of Cd2+ ions. The results are summarized in Table 3 (HMDE), Table 4 (GCE), and Table 5 (CF-PLA).
Despite the low level of macroscopically visible interaction of the solvents with the 3DPMs, the electrochemical interference was quite serious. The observed interferences occur primarily in the solvents, which caused mechanical damage as well, namely methanol-containing solution and NaOH solution, partly H2SO4 solution. The signals observed in the KCl solution were not influenced by the presence of any 3DPMs.
Contrary to the stability of 3DPMs, the main source of the interferences is Elastic resin; distinct peak distortion or even its disappearance was observable in most of the solvents. Moreover, this material releases electrochemically active compounds. In Clear resin, the presence of undesired compounds was also observed in methanol-containing solution; otherwise, this material interfered with the analyte signals only rarely.
The suppression of the analyte signals is also electrode dependent—it is most pronounced at HMDE, probably because of the specific sensitivity of the metallic surface. Results obtained using the CF-PLA electrode were somewhat scattered considering the peak height and potential, probably due to the interference of the electrolyte with the electrode material.

4. Conclusions

The research focused on the stability of four different 3DPMs (Clear and Elastic resins for stereolitography, and FMD filaments PLA and PET) in various organic solvents and how they affect the voltammetric behavior of selected redox probes at various electrodes if the material is in contact with the electrolyte solution.
Elastic resin did not suffer mechanical damage in contact with any of the tested solvents, although it underwent severe swelling. The reason might also lay in its flexibility, which protected it during the swelling—the mechanical damage caused to Clear resin seemed to be caused by the irregular swelling in some of the solvents. PET and PLA suffered more from delamination and dissolution; PLA to a greater extent than PET.
Potential window width was only slightly influenced by the contact with the 3DPMs. The only exception was the CF-PLA electrode which might be caused by the interaction of the electrolyte with the electrode material. On the other hand, the exposure of the 3DPMs to selected supporting electrolytes can influence the voltammetric behavior of analytes and suppress their faradaic response; this suppression is electrode-dependent. Elastic resin releases the most active interferences, although its outer appearance is not changed. The main finding achieved in this work is that even seemingly stable materials might not be suitable for the electrochemical application. The objects printed from Clear resin were mostly stable with the least amount of electrochemically interfering properties, and their suitability for the fabrication of 3D-printed electrochemical cells is highest, but their stability in strongly basic or mixed aqueous–organic solvents is also limited. The obtained results underline the necessity of a careful approach to the application of these materials in demanding conditions.

Supplementary Materials

The following are available online at https://www.mdpi.com/article/10.3390/bios12050308/s1, Figure S1: Optical microscopic and transmitting & reflecting microscopic images of 3D printing material PET before and after 24-h exposure to electrolytes), Figure S2: Optical microscopic and transmitting & reflecting microscopic images of 3D printing material PLA before and after 24-h exposure to electrolytes, Figure S3: Optical microscopic and transmitting & reflecting microscopic images of 3D printing material Elastic before and after 24-h exposure to electrolytes, Figure S4: Optical microscopic and transmitting & reflecting microscopic images of 3D printing material Clear before and after 24-h exposure to electrolytes.

Author Contributions

Conceptualization, T.N., J.F., H.D. and V.K.; methodology, J.F. and H.D.; formal analysis, L.M. and M.C.; resources, M.C.; data curation, M.C. and E.V.; writing—original draft preparation, M.C. and V.H.; writing—review and editing, T.N., H.D., V.K. and V.H.; supervision, V.H. and H.D.; project administration, T.N. and J.F.; funding acquisition, T.N. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the project of the Czech Science Foundation (GA ČR) No. 20-01589S (M.C., T.N., V.H.), Specific University Research (SVV260560) (M.C.), and Grant Agency of the Charles University No. 373521 (M.C.).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Cardoso, R.M.; Kalinke, C.; Rocha, R.G.; dos Santos, P.L.; Rocha, D.P.; Oliveira, P.R.; Janegitz, B.C.; Bonacin, J.A.; Richter, E.M.; Munoz, R.A.A. Additive-manufactured (3d-printed) electrochemical sensors: A critical review. Anal. Chim. Acta 2020, 1118, 73–91. [Google Scholar] [CrossRef] [PubMed]
  2. Escobar, J.G.; Vaneckova, E.; Lachmanova, S.N.; Vivaldi, F.; Heyda, J.; Kubista, J.; Shestivska, V.; Spanel, P.; Schwarzova-Peckova, K.; Rathousky, J.; et al. The development of a fully integrated 3d printed electrochemical platform and its application to investigate the chemical reaction between carbon dioxide and hydrazine. Electrochim. Acta 2020, 360, 136984. [Google Scholar] [CrossRef] [PubMed]
  3. Da Silveira, G.D.; Quero, R.F.; Bressan, L.P.; Bonacin, J.A.; de Jesus, D.P.; da Silva, A.F. Ready-to-use 3d-printed electrochemical cell for in situ voltammetry of immobilized microparticles and Raman spectroscopy. Anal. Chim. Acta 2021, 1141, 57–62. [Google Scholar] [CrossRef] [PubMed]
  4. Vaneckova, E.; Bousa, M.; Lachmanova, S.N.; Rathousky, J.; Gal, M.; Sebechlebska, T.; Kolivoska, V. 3d printed polylactic acid/carbon black electrodes with nearly ideal electrochemical behaviour. J. Electroanal. Chem. 2020, 857, 113745. [Google Scholar] [CrossRef]
  5. Sans, V. Emerging trends in flow chemistry enabled by 3d printing: Robust reactors, biocatalysis and electrochemistry. Curr. Opin. Green Sustain. Chem. 2020, 25, 100367. [Google Scholar] [CrossRef]
  6. Abdalla, A.; Patel, B.A. 3d-printed electrochemical sensors: A new horizon for measurement of biomolecules. Curr. Opin. Electrochem. 2020, 20, 78–81. [Google Scholar] [CrossRef]
  7. Hamzah, H.H.; Shafiee, S.A.; Abdalla, A.; Patel, B.A. 3d printable conductive materials for the fabrication of electrochemical sensors: A mini review. Electrochem. Commun. 2018, 96, 27–31. [Google Scholar] [CrossRef]
  8. Vaneckova, E.; Bousa, M.; Sokolova, R.; Moreno-Garcia, P.; Broekmann, P.; Shestivska, V.; Rathousky, J.; Gal, M.; Sebechlebska, T.; Kolivoska, V. Copper electroplating of 3d printed composite electrodes. J. Electroanal. Chem. 2020, 858, 113763. [Google Scholar] [CrossRef]
  9. Waseem, M.; Salah, B.; Habib, T.; Saleem, W.; Abas, M.; Khan, R.; Ghani, U.; Siddiqi, M.U.R. Multi-response optimization of tensile creep behavior of pla 3d printed parts using categorical response surface methodology. Polymers 2020, 12, 2962. [Google Scholar] [CrossRef]
  10. Kumar, M.B.; Sathiya, P. Methods and materials for additive manufacturing: A critical review on advancements and challenges. Thin-Walled Struct. 2021, 159, 107228. [Google Scholar] [CrossRef]
  11. Alghamdi, S.S.; John, S.; Choudhury, N.R.; Dutta, N.K. Additive manufacturing of polymer materials: Progress, promise and challenges. Polymers 2021, 13, 753. [Google Scholar] [CrossRef] [PubMed]
  12. Junpha, J.; Wisitsoraat, A.; Prathumwan, R.; Chaengsawang, W.; Khomungkhun, K.; Subannajui, K. Electronic tongue and cyclic voltammetric sensors based on carbon nanotube/polylactic composites fabricated by fused deposition modelling 3d printing. Mater. Sci. Eng. C-Mater. Biol. Appl. 2020, 117, 111319. [Google Scholar] [CrossRef] [PubMed]
  13. Choinska, M.; Hrdlicka, V.; Sestakova, I.; Navratil, T. Voltammetric determination of heavy metals in honey bee venom using hanging mercury drop electrode and pla/carbon conductive filament for 3d printer. Monatsh. Chem. 2021, 152, 35–41. [Google Scholar] [CrossRef] [PubMed]
  14. Dixit, C.K.; Kadimisetty, K.; Rusling, J. 3d-printed miniaturized fluidic tools in chemistry and biology. TrAC-Trends Anal. Chem. 2018, 106, 37–52. [Google Scholar] [CrossRef] [PubMed]
  15. Stansbury, J.W.; Idacavage, M.J. 3D printing with polymers: Challenges among expanding options and opportunities. Dent. Mater. 2016, 32, 54–64. [Google Scholar] [CrossRef]
  16. Ambrosi, A.; Pumera, M. 3d-printing technologies for electrochemical applications. Chem. Soc. Rev. 2016, 45, 2740–2755. [Google Scholar] [CrossRef] [Green Version]
  17. Lee, J.Y.; An, J.; Chua, C.K. Fundamentals and applications of 3d printing for novel materials. Appl. Mater. Today 2017, 7, 120–133. [Google Scholar] [CrossRef]
  18. Parra-Cabrera, C.; Achille, C.; Kuhn, S.; Ameloot, R. 3d printing in chemical engineering and catalytic technology: Structured catalysts, mixers and reactors. Chem. Soc. Rev. 2018, 47, 209–230. [Google Scholar] [CrossRef]
  19. Xing, D.; Chen, L.; Ma, Q.; Hao, B.; Gutnikov, S.I.; Lazoryak, B.I.; Mader, E.; Ma, P.C. What happens to glass fiber under extreme chemical conditions? J. Non-Cryst. Solids 2020, 548, 120331. [Google Scholar] [CrossRef]
  20. Gu, H. Tensile behaviours of some high performance filaments after naoh treatment. Mater. Des. 2008, 29, 1893–1896. [Google Scholar] [CrossRef]
  21. Heikkinen, I.T.S.; Kauppinen, C.; Liu, Z.J.; Asikainen, S.M.; Spoljaric, S.; Seppala, J.V.; Savin, H.; Pearce, J.M. Chemical compatibility of fused filament fabrication-based 3-d printed components with solutions commonly used in semiconductor wet processing. Addit. Manuf. 2018, 23, 99–107. [Google Scholar] [CrossRef] [Green Version]
  22. Salentijn, G.I.J.; Oomen, P.E.; Grajewski, M.; Verpoorte, E. Fused deposition modeling 3d printing for (bio)analytical device fabrication: Procedures, materials, and applications. Anal. Chem. 2017, 89, 7053–7061. [Google Scholar] [CrossRef] [Green Version]
  23. IUPAC. Compendium of Analytical Nomenclature. Available online: https://media.iupac.org/publications/analytical_compendium/ (accessed on 26 November 2021).
  24. Novotny, F.; Urbanova, V.; Plutnar, J.; Pumera, M. Preserving fine structure details and dramatically enhancing electron transfer rates in graphene 3d-printed electrodes via thermal annealing: Toward nitroaromatic explosives sensing. ACS Appl. Mater. Interfaces 2019, 11, 35371–35375. [Google Scholar] [CrossRef] [PubMed]
  25. Kalinke, C.; Neumsteir, N.V.; Aparecido, G.D.; Ferraz, T.V.D.; dos Santos, P.L.; Janegitz, B.C.; Bonacin, J.A. Comparison of activation processes for 3d printed pla-graphene electrodes: Electrochemical properties and application for sensing of dopamine. Analyst 2020, 145, 1207–1218. [Google Scholar] [CrossRef] [PubMed]
  26. Redondo, E.; Munoz, J.; Pumera, M. Green activation using reducing agents of carbon-based 3d printed electrodes: Turning good electrodes to great. Carbon 2021, 175, 413–419. [Google Scholar] [CrossRef]
  27. Wirth, D.M.; Sheaff, M.J.; Waldman, J.V.; Symcox, M.P.; Whitehead, H.D.; Sharp, J.D.; Doerfler, J.R.; Lamar, A.A.; LeBlanc, G. Electrolysis activation of fused-filament-fabrication 3d-printed electrodes for electrochemical and spectroelectrochemical analysis. Anal. Chem. 2019, 91, 5553–5557. [Google Scholar] [CrossRef]
  28. Browne, M.P.; Novotny, F.; Sofer, Z.; Pumera, M. 3d printed graphene electrodes’ electrochemical activation. ACS App. Mat. Interf. 2018, 10, 40294–40301. [Google Scholar] [CrossRef]
  29. Manzanares-Palenzuela, C.L.; Hermanova, S.; Sofer, Z.; Pumera, M. Proteinase- sculptured 3d-printed graphene/polylactic acid electrodes as potential biosensing platforms: Towards enzymatic modeling of 3d-printed structures dagger. Nanoscale 2019, 11, 12124–12131. [Google Scholar] [CrossRef]
  30. Fischer, J.; Barek, J.; Yosypchuk, B.; Navratil, T. Voltammetric determination of trace amounts of 2-methyl-4,6-dinitrophenol at a silver solid amalgam electrode. Electroanalysis 2006, 18, 127–130. [Google Scholar] [CrossRef]
  31. Erokhin, K.S.; Gordeev, E.G.; Ananikov, V.P. Revealing interactions of layered polymeric materials at solid-liquid interface for building solvent compatibility charts for 3d printing applications. Sci. Rep. 2019, 9, 20177. [Google Scholar] [CrossRef]
  32. Barek, J.; Fischer, J.; Navratil, T.; Peckova, K.; Yosypchuk, B.; Zima, J. Nontraditional electrode materials in environmental analysis of biologically active organic compounds. Electroanalysis 2007, 19, 2003–2014. [Google Scholar] [CrossRef]
  33. Navratil, T.; Yosypchuk, B.; Barek, J. A multisensor for electrochemical sequential autonomous automatic measurements. Chem. Anal.-Warsaw 2009, 54, 3–17. [Google Scholar]
  34. Lyu, S.P.; Untereker, D. Degradability of polymers for implantable biomedical devices. Int. J. Mol. Sci. 2009, 10, 4033–4065. [Google Scholar] [CrossRef] [PubMed] [Green Version]
Figure 1. 3D printed models from TrafficWhite PLA.
Figure 1. 3D printed models from TrafficWhite PLA.
Biosensors 12 00308 g001
Figure 2. Available potential windows in various electrolytes before and after their 24-h exposition to 3D-printed objects from Elastic, Clear, PET, and PLA at HMDE, GCE, and CF-PLA.
Figure 2. Available potential windows in various electrolytes before and after their 24-h exposition to 3D-printed objects from Elastic, Clear, PET, and PLA at HMDE, GCE, and CF-PLA.
Biosensors 12 00308 g002
Table 1. Accumulation potentials in various solvents used during DPASV measurement.
Table 1. Accumulation potentials in various solvents used during DPASV measurement.
Electrode:HMDEGCECF-PLA
Solvent Accumulation potential (Eacc) [mV]
1 mol dm−3 KCl, 1 mol dm−3 NaOH−1800−1600−1000
1 mol dm−3 KCl: MeOH (9:1, v/v)−1000−1600−1000
1 mol dm−3 H2SO4−1000−1000−1000
Table 2. Changes of 3D printed square plate objects (10 × 10 × 1 mm) after 24-h immersion to various solvents and solutions. Numbers denote relative elongation (%) of the objects after the immersion.
Table 2. Changes of 3D printed square plate objects (10 × 10 × 1 mm) after 24-h immersion to various solvents and solutions. Numbers denote relative elongation (%) of the objects after the immersion.
SolventMaterial
ElasticClearPETPLA
Acetone257
cracked
×
soft
13
delamination
Acetonitrile142×17 delamination
Chloroform53---
broken
---
dissolved
---
dissolved
Dichloromethane4516
broken
---
dissolved
---
dissolved
Diethyl ether21×××
Dimethyl sulfoxide24××
soft
5
soft
Ethanol332××
Ethyl acetate29××
soft
12
cracked, soft
Formaldehyde 38%7×××
Hexane5×××
Isopropanol29×××
Methanol248
cracked
××
N,N-dimethylformamide383
deformation
3
soft, deformation
23
delamination
n-penthanol32×××
Tetrahydrofuran464
deformation
---
partly dissolved
---
dissolved
Toluene33××8
cracked, soft,
Aqueous 1M KCl××××
Aqueous 1M NaOH×××etched
Aqueous 1M H2SO4××××
1M KCl: MeOH (1:9, v/v)196××
× no observed change; --- impossible to obtain the value.
Table 3. DPV and DPASV signals of Cd2+ and [Ru(NH3)6]3+ recorded using HMDE before and after 24-h electrolyte exposures to 3DPMs.
Table 3. DPV and DPASV signals of Cd2+ and [Ru(NH3)6]3+ recorded using HMDE before and after 24-h electrolyte exposures to 3DPMs.
Printing MaterialBeforeElasticClearPETPLA
Supporting electrolyte1 mol dm−3 KCl
Cd2+; DPVEp [mV]−589−584−589−586−589
Jp [µA mm−2]1.161.091.321.211.16
Cd2+; DPASVEp [mV]−587−580−584−582−592
Jp [µA mm−2]2.781.902.121.992.21
[Ru(NH3)6]3+; DPVEp [mV]−154−145−151−148−149
Jp [µA mm−2]0.180.190.170.180.22
Supporting electrolyte1 mol dm−3 NaOH
[Ru(NH3)6]3+; DPVEp [mV]−238-−229--
Jp [µA mm−2]0.18<Cc0.04<Cc<Cc
Supporting electrolyte1 mol dm−3 H2SO4
Cd2+; DPVEp [mV]−541-−541−546−540
Jp [µA mm−2]0.94<Cc0.430.430.46
Cd2+; DPASVEp [mV]−587−493−589−589−588
Jp [µA mm−2]2.150.282.101.792.85
[Ru(NH3)6]3+; DPVEp [mV]−180−173−177−176−176
Jp [µA mm−2]0.220.220.210.210.23
Supporting electrolyte1 mol dm−3 KCl: MeOH (1:9, v/v)
Cd2+; DPVEp [mV]−714-−704−726−722
Jp [µA mm−2]0.59<Cc0.682.982.59
Cd2+; DPASVEp [mV]−647−639−636−636−659
Jp [µA mm−2]1.570.320.100.141.11
[Ru(NH3)6]3+; DPVEp [mV]−165−171−153−159−159
Jp [µA mm−2]0.200.060.170.230.24
Cc—Critical level.
Table 4. Evaluation of DPV and DPASV signals of Cd2+, [Ru(NH3)6]3+, and [Fe(CN)6]3− recorded using GCE before and after 24-h electrolyte exposures to 3DPMs.
Table 4. Evaluation of DPV and DPASV signals of Cd2+, [Ru(NH3)6]3+, and [Fe(CN)6]3− recorded using GCE before and after 24-h electrolyte exposures to 3DPMs.
Printing MaterialBeforeElasticClearPETPLA
Supporting electrolyte1 mol dm−3 KCl
Cd2+; DPVEp [mV]–589–584–589–586–589
Jp [µA mm−2]1.161.091.321.211.16
Cd2+; DPASVEp [mV]–587–580–584–582–592
Jp [µA mm−2]2.781.902.121.992.21
[Ru(NH3)6]3+; DPVEp [mV]–154–145–151–148–149
Jp [µA mm−2]0.180.190.170.180.22
[Fe(CN)6]3−Ep [mV]233233232230229
Jp [µA mm−2]2.042.102.242.202.26
Supporting electrolyte1 mol dm−3 NaOH
[Ru(NH3)6]3+; DPVEp [mV]–238-–229--
Jp [µA mm−2]0.18<Cc0.04<Cc<Cc
Supporting electrolyte1 mol dm−3 H2SO4
Cd2+; DPVEp [mV]–541-–541–546–540
Jp [µA mm−2]0.94<Cc0.430.430.46
Cd2+; DPASVEp [mV]–587–493–589–589–588
Jp [µA mm−2]2.150.282.101.792.85
[Ru(NH3)6]3+; DPVEp [mV]–180–173–177–176–176
Jp [µA mm−2]0.220.220.210.210.23
[Fe(CN)6]3−Ep [mV]271263264268268
Jp [µA mm−2]2.001.942.041.841.92
Supporting electrolyte1 mol dm−3 KCl: MeOH (1:9, v/v)
Cd2+; DPVEp [mV]–714-–704–726–722
Jp [µA mm−2]0.59<Cc0.682.982.59
Cd2+; DPASVEp [mV]–647–639–636–636–659
Jp [µA mm−2]1.570.320.100.141.11
[Ru(NH3)6]3+; DPVEp [mV]–165–171–153–159–159
Jp [µA mm−2]0.200.060.170.230.24
[Fe(CN)6]3−Ep [mV]154129179116163
Jp [µA mm−2]0.170.040.220.140.12
Cc—Critical level.
Table 5. Evaluation of DPV of [Fe(CN)6]3 and [Ru(NH3)6]3+ recorded using CF-PLA before and after 24-h electrolyte exposures to 3DPMs.
Table 5. Evaluation of DPV of [Fe(CN)6]3 and [Ru(NH3)6]3+ recorded using CF-PLA before and after 24-h electrolyte exposures to 3DPMs.
Printing MaterialBeforeElasticClearPETPLA
Supporting electrolyte1 mol dm−3 KCl
[Ru(NH3)6]3+; DPVEp [mV]–158–163–158–158–158
Jp [µA mm−2]0.1380.0930.0970.1290.130
[Fe(CN)6]3; DPVEp [mV]519439439413435
Jp [µA mm−2]0.0020.0040.0020.0020.002
Supporting electrolyte1 mol dm−3 NaOH
[Ru(NH3)6]3+; DPVEp [mV]–189–192–197–203–180
Jp [µA mm−2]0.0590.0300.0520.0660.051
[Fe(CN)6]3; DPVEp [mV]339240234242229
Jp [µA mm−2]0.0070.1310.1330.0690.111
Supporting electrolyte1 mol dm−3 H2SO4
[Ru(NH3)6]3+; DPVEp [mV]–198–204–201–201–223
Jp [µA mm−2]0.0370.0230.0220.0230.022
[Fe(CN)6]3; DPVEp [mV]237304303256296
Jp [µA mm−2]0.0140.0110.0090.0110.010
Supporting electrolyte1 mol dm−3 KCl: MeOH (1:9, v/v)
[Ru(NH3)6]3+; DPVEp [mV]–141–250–149–170–170
Jp [µA mm−2]0.5170.6310.5020.5770.415
[Fe(CN)6]3; DPVEp [mV]29953107--
Jp [µA mm−2]0.0050.1830.048<Cc<Cc
Cc—Critical level.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Choińska, M.; Hrdlička, V.; Dejmková, H.; Fischer, J.; Míka, L.; Vaněčková, E.; Kolivoška, V.; Navrátil, T. Applicability of Selected 3D Printing Materials in Electrochemistry. Biosensors 2022, 12, 308. https://doi.org/10.3390/bios12050308

AMA Style

Choińska M, Hrdlička V, Dejmková H, Fischer J, Míka L, Vaněčková E, Kolivoška V, Navrátil T. Applicability of Selected 3D Printing Materials in Electrochemistry. Biosensors. 2022; 12(5):308. https://doi.org/10.3390/bios12050308

Chicago/Turabian Style

Choińska, Marta, Vojtěch Hrdlička, Hana Dejmková, Jan Fischer, Luděk Míka, Eva Vaněčková, Viliam Kolivoška, and Tomáš Navrátil. 2022. "Applicability of Selected 3D Printing Materials in Electrochemistry" Biosensors 12, no. 5: 308. https://doi.org/10.3390/bios12050308

APA Style

Choińska, M., Hrdlička, V., Dejmková, H., Fischer, J., Míka, L., Vaněčková, E., Kolivoška, V., & Navrátil, T. (2022). Applicability of Selected 3D Printing Materials in Electrochemistry. Biosensors, 12(5), 308. https://doi.org/10.3390/bios12050308

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop