Impact of Different Glidepath Techniques on the Overall Performance of WaveOne Gold in an Artificial S-Shape Canal
Abstract
:1. Introduction
2. Materials and Methods
- The simulated canal was explored with a new #10 K-file;
- Working length was established;
- The endodontic motor was set on PG program, using 300 RPM and a torque of 4 Ncm;
- In the presence of saline solution, the PG file was used to passively follow the canal, in one or more passes until working length was reached;
- Working length was reconfirmed.
- The simulated canal was explored with a new #10 K-file;
- Working length was established;
- The endodontic motor was set on the WOGG program that used the manufacturer’s settings for speed, torque, and angles of reciprocation;
- In the presence of saline solution, the WOGG file was used with gentle inward pressure so the file would passively follow the secured canal in one or more passes until working length was reached;
- Working length was reconfirmed.
- The endodontic motor was set on WOG Primary program that used the manufacturer’s settings for speed, torque, and angles of reciprocation;
- In the presence of saline solution, the file was used with gentle inward pressure and was let to passively progress inside the canal with in-and-out pecking motion;
- After progressing 2–3 mm (about three pecks) of the canal the file was removed and cleaned, and the canal was irrigated following the IRI protocol;
- The pecking motions were repeated until the instrument reached full working length, separated, blocked the canal, or created a iatrogenic error (unpassable ledge, via falsa).
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Park, S.; Kim, Y.; Shon, W.; You, S.; Moon, Y.; Kim, H.; Lee, W. Clinical Efficiency and Reusability of the Reciprocating Nickel–Titanium Instruments According to the Root Canal Anatomy. Scanning 2014, 36, 246–251. [Google Scholar] [CrossRef] [PubMed]
- Berutti, E.; Cantatore, G.; Castellucci, A.; Chiandussi, G.; Pera, F.; Migliaretti, G.; Pasqualini, D. Use of Nickel-Titanium Rotary PathFile to Create the Glide Path: Comparison With Manual Preflaring in Simulated Root Canals. J. Endod. 2009, 35, 408–412. [Google Scholar] [CrossRef] [PubMed]
- Blum, J.Y.; Cohen, A.; Machtou, P.; Micallef, J.P. Analysis of Forces Developed during Mechanical Preparation of Extracted Teeth Using Profile NiTi Rotary Instruments. Int. Endod. J. 1999, 32, 24–31. [Google Scholar] [CrossRef]
- Berutti, E.; Negro, A.; Lendini, M.; Pasqualini, D. Influence of Manual Preflaring and Torque on the Failure Rate of ProTaper Rotary Instruments. J. Endod. 2004, 30, 228–230. [Google Scholar] [CrossRef] [PubMed]
- Patino, P.; Biedma, B.; Liebana, C.; Cantatore, G.; Bahillo, J. The Influence of a Manual Glide Path on the Separation Rate of NiTi Rotary Instruments. J. Endod. 2005, 31, 114–116. [Google Scholar] [CrossRef] [PubMed]
- Bürklein, S.; Schäfer, E. Critical Evaluation of Root Canal Transportation by Instrumentation. Endod. Top. 2013, 29, 110–124. [Google Scholar] [CrossRef]
- West, J.D. The Endodontic Glidepath: “Secret to Rotary Safety”. Dent. Today 2010, 29, 86+88+90–93. [Google Scholar] [PubMed]
- van der Vyver, P. Creating a Glide Path for Rotary NiTi Instruments: Part One. Int. Dent. SA 2010, 13, 46–53. [Google Scholar]
- de Oliveira Alves, V.; da Silveira Bueno, C.E.; Cunha, R.S.; Pinheiro, S.L.; Fontana, C.E.; de Martin, A.S. Comparison among Manual Instruments and PathFile and Mtwo Rotary Instruments to Create a Glide Path in the Root Canal Preparation of Curved Canals. J. Endod. 2012, 38, 117–120. [Google Scholar] [CrossRef]
- Paleker, F.; van der Vyver, P.J. Comparison of Canal Transportation and Centering Ability of K-Files, ProGlider File, and G-Files: A Micro-Computed Tomography Study of Curved Root Canals. J. Endod. 2016, 42, 1105–1109. [Google Scholar] [CrossRef]
- Nakagawa, R.K.L.; Alves, J.L.; Buono, V.T.L.; Bahia, M.G.A. Flexibility and Torsional Behaviour of Rotary Nickel-Titanium PathFile, RaCe ISO 10, Scout RaCe and Stainless Steel K-File Hand Instruments. Int. Endod. J. 2014, 47, 290–297. [Google Scholar] [CrossRef]
- Knowles, K.I.; Hammond, N.B.; Biggs, S.G.; Ibarrola, J.L. Incidence of Instrument Separation Using LightSpeed Rotary Instruments. J. Endod. 2006, 32, 14–16. [Google Scholar] [CrossRef]
- Dagna, A.; El Abed, R.; Hussain, S.; Abu-Tahun, I.H.; Visai, L.; Bertoglio, F.; Bosco, F.; Beltrami, R.; Poggio, C.; Kim, H.-C. Comparison of Apical Extrusion of Intracanal Bacteria by Various Glide-Path Establishing Systems: An in Vitro Study. Restor. Dent. Endod. 2017, 42, 316. [Google Scholar] [CrossRef]
- Ruddle, C.J.; Machtou, P.; West, J.D. Endodontic Canal Preparation: Innovations in Glide Path Management and Shaping Canals. Dent. Today 2014, 33, 118–123. [Google Scholar]
- Grande, N.M.; Ahmed, H.M.A.; Cohen, S.; Bukiet, F.; Plotino, G. Current Assessment of Reciprocation in Endodontic Preparation: A Comprehensive Review—Part I: Historic Perspectives and Current Applications. J. Endod. 2015, 41, 1778–1783. [Google Scholar] [CrossRef]
- Ahn, S.-Y.; Kim, H.-C.; Kim, E. Kinematic Effects of Nickel-Titanium Instruments with Reciprocating or Continuous Rotation Motion: A Systematic Review of In Vitro Studies. J. Endod. 2016, 42, 1009–1017. [Google Scholar] [CrossRef]
- Cunha, R.S.; Junaid, A.; Ensinas, P.; Nudera, W.; da Silveira Bueno, C.E. Assessment of the Separation Incidence of Reciprocating WaveOne Files: A Prospective Clinical Study. J. Endod. 2014, 40, 922–924. [Google Scholar] [CrossRef]
- Coelho, M.S.; Fontana, C.E.; Kato, A.S.; de Martin, A.S.; da Silveira Bueno, C.E. Effects of Glide Path on the Centering Ability and Preparation Time of Two Reciprocating Instruments. Iran. Endod. J. 2016, 11, 33–37. [Google Scholar]
- De-Deus, G.; Arruda, T.E.P.; Souza, E.M.; Neves, A.; Magalhães, K.; Thuanne, E.; Fidel, R.A.S. The Ability of the Reciproc R25 Instrument to Reach the Full Root Canal Working Length without a Glide Path. Int. Endod. J. 2013, 46, 993–998. [Google Scholar] [CrossRef]
- Bartols, A.; Robra, B.-P.; Walther, W. The Ability of Reciproc Instruments to Reach Full Working Length without Glide Path Preparation: A Clinical Retrospective Study. PeerJ 2017, 5, e3583. [Google Scholar] [CrossRef]
- De-Deus, G.; Cardoso, M.L.; Belladonna, F.G.; Cavalcante, D.M.; Simões-Carvalho, M.; Souza, E.M.; Lopes, R.T.; Silva, E.J.N.L. Performance of Reciproc Blue R25 Instruments in Shaping the Canal Space without Glide Path. J. Endod. 2019, 45, 194–198. [Google Scholar] [CrossRef]
- Jonker, C.H.; Van der Vyver, P.J.; De Wet, F.A. The Influence of Glide Path Preparation on the Failure Rate of WaveOne Reciprocating Instruments. SADJ 2014, 69, 266–269. [Google Scholar]
- Yoshimine, Y.; Ono, M.; Akamine, A. The Shaping Effects of Three Nickel-Titanium Rotary Instruments in Simulated S-Shaped Canals. J. Endod. 2005, 31, 373–375. [Google Scholar] [CrossRef]
- Keskin, C.; Inan, U.; Demiral, M.; Keleş, A. Cyclic Fatigue Resistance of Reciproc Blue, Reciproc, and WaveOne Gold Reciprocating Instruments. J. Endod. 2017, 43, 1360–1363. [Google Scholar] [CrossRef]
- Adıgüzel, M.; Capar, I.D. Comparison of Cyclic Fatigue Resistance of WaveOne and WaveOne Gold Small, Primary, and Large Instruments. J. Endod. 2017, 43, 623–627. [Google Scholar] [CrossRef]
- Zheng, L.; Ji, X.; Li, C.; Zuo, L.; Wei, X. Comparison of Glide Paths Created with K-Files, PathFiles, and the ProGlider File, and Their Effects on Subsequent WaveOne Preparation in Curved Canals. BMC Oral Health 2018, 18, 152. [Google Scholar] [CrossRef]
- Vorster, M.; van der Vyver, P.J.; Paleker, F. Influence of Glide Path Preparation on the Canal Shaping Times of WaveOne Gold in Curved Mandibular Molar Canals. J. Endod. 2018, 44, 853–855. [Google Scholar] [CrossRef]
- Toshihiro, U.; Tomoya, H.; Masashi, T.; Kazuki, K.; Satoshi, K. A Comparison of Glide Path Preparation Times of Three Different Techniques in Constricted Canals. J. Japan Endod. Assoc. 2023, 44, 22–26. [Google Scholar]
- Han, Y.; Hou, X.-M. Glide Path Enlargement of Curved Molar Canals Using HyFlex EDM Glide Path File versus PathFile: A Comparative Study of Preparation Time and Postoperative Pain. BMC Oral Health 2021, 21, 150. [Google Scholar] [CrossRef] [PubMed]
- Paleker, F.; van der Vyver, P.J. Glide Path Enlargement of Mandibular Molar Canals by Using K-Files, the ProGlider File, and G-Files: A Comparative Study of the Preparation Times. J. Endod. 2017, 43, 609–612. [Google Scholar] [CrossRef] [PubMed]
- Kirchhoff, A.L.; Chu, R.; Mello, I.; Garzon, A.D.P.; dos Santos, M.; Cunha, R.S. Glide Path Management with Single- and Multiple-Instrument Rotary Systems in Curved Canals: A Micro-Computed Tomographic Study. J. Endod. 2015, 41, 1880–1883. [Google Scholar] [CrossRef]
- Kırıcı, D.; Koç, S.; Kuştarcı, A. Effects of Different Glide Path Techniques on the Amount of Extruded Debris and Preparation Times during Root Canal Preparation. J. Dent. Res. Dent. Clin. Dent. Prospect. 2020, 14, 187–190. [Google Scholar] [CrossRef]
- Berutti, E.; Paolino, D.S.; Chiandussi, G.; Alovisi, M.; Cantatore, G.; Castellucci, A.; Pasqualini, D. Root Canal Anatomy Preservation of WaveOne Reciprocating Files with or without Glide Path. J. Endod. 2012, 38, 101–104. [Google Scholar] [CrossRef]
GP_Time (s) | Mean | SD | Min | Max | p-Value |
---|---|---|---|---|---|
PG | 5.2005 | 1.38373 | 2.92 | 8.95 | 0.089 |
WOGG | 4.5555 | 0.8872 | 3.02 | 6.63 |
Variable | NGP | PG | WOGG | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Mean | SD | Min | Max | Mean | SD | Min | Max | Mean | SD | Min | Max | |
Shape_time (s) | 17.1682 | 3.99178 | 12.58 | 29.29 | 14.8668 | 2.64826 | 11.08 | 21.67 | 13.1847 | 1.9147 | 9.04 | 15.85 |
Total_time (s) | 17.1682 | 2.99178 | 12.58 | 29.29 | 20.0805 | 3.24897 | 15.31 | 28.73 | 17.2984 | 2.52387 | 12.65 | 22.37 |
RFWL | 0.85 | 0.36635 | 0 | 1 | 0.95 | 0.22361 | 0 | 1 | 0.95 | 0.22361 | 0 | 1 |
NPM | 9.7059 | 1.15999 | 8 | 12 | 8.2632 | 1.36797 | 6 | 12 | 7.0526 | 1.02598 | 4 | 8 |
Shape_Time (s) | Total_Time (s) | NPM | |||||
---|---|---|---|---|---|---|---|
Group | Mean Difference | p-Value | Mean Difference | p-Value | Mean Difference | p-Value | |
NGP | PG | 2.30139 | 0.058 | −2.91229 | 0.028 | 1.44272 | 0.002 |
WOGG | 3.98350 | 0.000 | −0.53019 | 0.879 | 2.65325 | 0.000 | |
PG | NGP | −2.30139 | 0.058 | 2.91229 | 0.028 | −1.44272 | 0.002 |
WOGG | 1.68211 | 0.190 | 2.38211 | 0.074 | 1.21053 | 0.008 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Lup, V.M.; Marcu, O.A.; Gaeta, C.; Ciavoi, G. Impact of Different Glidepath Techniques on the Overall Performance of WaveOne Gold in an Artificial S-Shape Canal. Dent. J. 2024, 12, 182. https://doi.org/10.3390/dj12060182
Lup VM, Marcu OA, Gaeta C, Ciavoi G. Impact of Different Glidepath Techniques on the Overall Performance of WaveOne Gold in an Artificial S-Shape Canal. Dentistry Journal. 2024; 12(6):182. https://doi.org/10.3390/dj12060182
Chicago/Turabian StyleLup, Vlad Mircea, Olivia Andreea Marcu, Carlo Gaeta, and Gabriela Ciavoi. 2024. "Impact of Different Glidepath Techniques on the Overall Performance of WaveOne Gold in an Artificial S-Shape Canal" Dentistry Journal 12, no. 6: 182. https://doi.org/10.3390/dj12060182
APA StyleLup, V. M., Marcu, O. A., Gaeta, C., & Ciavoi, G. (2024). Impact of Different Glidepath Techniques on the Overall Performance of WaveOne Gold in an Artificial S-Shape Canal. Dentistry Journal, 12(6), 182. https://doi.org/10.3390/dj12060182