Next Article in Journal
Saliva, a Magic Biofluid Available for Multilevel Assessment and a Mirror of General Health—A Systematic Review
Next Article in Special Issue
Optimization of the Electrodeposition of Gold Nanoparticles for the Application of Highly Sensitive, Label-Free Biosensor
Previous Article in Journal
The Recent Advances in Raman Microscopy and Imaging Techniques for Biosensors
Previous Article in Special Issue
Ultrathin Functional Polymer Modified Graphene for Enhanced Enzymatic Electrochemical Sensing
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Response Surface Methodology for the Optimisation of Electrochemical Biosensors for Heavy Metals Detection

by
Giuseppe Egidio De Benedetto
1,
Sabrina Di Masi
2,*,
Antonio Pennetta
1 and
Cosimino Malitesta
2
1
Dipartimento di Beni Culturali, Università del Salento, Via D. Birago 64, 73100 Lecce, Italy
2
Dipartimento di Scienze e Tecnologie Biologiche ed Ambientali, Via per Monteroni 1, 73100 Lecce, Italy
*
Author to whom correspondence should be addressed.
Biosensors 2019, 9(1), 26; https://doi.org/10.3390/bios9010026
Submission received: 11 January 2019 / Revised: 9 February 2019 / Accepted: 9 February 2019 / Published: 13 February 2019
(This article belongs to the Special Issue Enzymatic Electrochemical Biosensors)

Abstract

:
Herein, we report the application of a chemometric tool for the optimisation of electrochemical biosensor performances. The experimental design was performed based on the responses of an amperometric biosensor developed for metal ions detection using the flow injection analysis. The electrode preparation and the working conditions were selected as experimental parameters, and thus, were modelled by a response surface methodology (RSM). In particular, enzyme concentration, flow rates, and number of cycles were reported as continuous factors, while the sensitivities of the biosensor (S, µA·mM−1) towards metals, such as Bi3+ and Al3+ were collected as responses and optimised by a central composite design (CCD). Bi3+ and Al3+ inhibition on the Pt/PPD/GOx biosensor response is for the first time reported. The optimal enzyme concentration, scan cycles and flow rate were found to be 50 U·mL−1, 30 and, 0.3 mL·min−1, respectively. Descriptive/predictive performances are discussed: the sensitivities of the optimised biosensor agreed with the experimental design prediction. The responses under the optimised conditions were also tested towards Ni2+ and Ag+ ions. The multivariate approach used in this work allowed us to obtain a wide working range for the biosensor, coupled with a high reproducibility of the response (RSD = 0.72%).

Graphical Abstract

1. Introduction

Heavy metal ions are pollutants that seriously affect the health of the ecosystem due to their non-biodegradability and enrichment in the environment and biological systems [1]. Nowadays, the determination of heavy-metal ions in water samples represents an important issue. Different analytical techniques are employed for qualitative and quantitative detection of heavy metal ions [2], including atomic absorption spectroscopy (AAS), atomic fluorescence spectroscopy (AFS) and inductively coupled plasma mass spectrometry (ICP-MS) [3]. The above techniques have several advantages, such as high sensitivity, selectivity and accuracy. Otherwise, they present some problems, such as the complexity, cost and long steps of pre-concentration and analysis. For these reasons, portable devices, such as biosensors [4,5], are required for on-site monitoring of heavy-metal ions. Enzymes are widely integrated in biosensors for sensitive determination of their substrates, inhibitors and/or competitors [6]. The entrapment of enzymes in electrosynthesised polymeric networks [7] is now commonly used for the development of biosensors [8]. Among several enzymes, the inhibition of glucose oxidase (GOx) has been extensively studied in the field of electrochemical biosensors for the detection of a wide set of inhibitors [9,10], including heavy-metal ions [11,12,13,14,15]. In addition, enzyme inhibition activities represent an attractive approach to producing a simple and rapid detection of the bioavailable fraction of heavy metal ions in water samples [16,17,18,19,20,21]. Owing to the complexity of these systems, the development of biosensors requires the studying of a set of parameters affecting their analytical responses. In the past, the most commonly-used approach to evaluating and optimising the influence of factors on the experimental responses was based on the univariate statistics. This technique precludes inclusion all the interactions among the variables studied in the experimental plan. The design of experiment (DOE) is useful to study all the interactions between the factors and to overcome the limits offered by a simple “one-at-a-time” approach [22,23,24,25]. It also helps the user to perform the optimisation of the parameters with a reduction of reagents consumption and number of experiments. In the literature, DOE has been little applied for the optimisation of biosensor performances [26,27,28,29,30]. In this work, the Pt/PPD/GOx biosensors, described by our previous works [31,32,33], were used in a flow injection setup and relevant responses were optimised. The effects of varying enzyme concentration (U·mL−1), the number of voltammetric cycles during the preparation of Pt/PPD/GOx biosensor and the flow rate (µL·min−1) on sensitivity (S, µA·mM−1) toward the new tested Bi3+ and Al3+ ions were investigated using response surface methodology (RSM) based on central composite design (CCD). We also report the performance of the biosensor for Ni2+ and Ag+ ions—already tested in our previous works—under the best selected conditions. Lineweaver-Burk and Dixon plots were reported to elucidate information about kinetic parameters and degree of inhibition on the studied biosensors toward previous studied ions.

2. Materials and Methods

2.1. Apparatus

Experiments were carried out with a computer controlled PalmSens potentiostat. All electrochemical data were collected and analysed by the PalmSens PSTrace software. The electrochemical cell was purchased from Dropsens and consisted of a plexiglass flow cell and a disposable screen ink-printed platinum electrode (SPPtE, DRP-150, Dropsens, Italy). They consisted in a platinum disk-shaped (12.6 mm2) working electrode, a paste of silver/silver chloride pseudo-reference and a platinum strip counter electrode, on a ceramic substrate (3.3 cm × 1.0 cm). All the electrochemical measurements were referred to the screen-printed silver pseudo-reference electrode. The screen printed electrodes were interfaced to the potentiostat by a cable connector for SPEs (DRP-CAC, Dropsens, Italy). A Gilson MiniPuls 3 peristaltic pump and a Rheodyne low pressure injector with a 200-µL sample loop completed the apparatus.

2.2. Chemicals

Glucose oxidase (GOx) from Aspergillus niger (Type VII, 248073 U/g) and D (+) glucose were purchased from Sigma-Aldrich. Bismuth standard solution (1000 ppm), Al(NO3)3, Hg(NO3)2, AgNO3, Ni(NO3)2 were purchased from Fluka. Acetate buffer (50 mM, pH = 5.2) was prepared from 50 mM acetic acid brought to pH 5.2 with NaOH. Glucose stock solutions were prepared in acetate buffer and were allowed to mutarotate overnight before use. 1 mM stock solution of metal ions were prepared in acetate buffer (50 mM, pH = 5.2) and diluted with the same buffer to give the required concentrations.

2.3. Preparation of the Pt/PPD/GOx biosensor

In order to prepare the biosensor, platinum screen-printed electrodes were chosen as the transducer element, as reported in our previous works [30,31,32,33]. Therefore, biosensors were grown on the surface working electrodes using a CV technique. Briefly, the surface of a platinum screen-printed electrode was washed with Milli-Q water. The electrodes were conditioned by cyclic voltammetry (CV) in 10 mM K3Fe(CN)6 solution between −0.3 V and +0.5 V until a steady state was reached. Thus, a drop of 50 µL of a solution containing variable concentration of GOx and 5 mmol/L of o-phenylenediamine were casted on the electrode surface, then a cyclic voltammetry between −0.07 V and +0.77 V was performed for the electrochemical grown of the polymer. The electrode was then rinsed with acetate buffer and mounted in the flow cell for the measurements. The Pt/PPD/GOx biosensors were stored in acetate buffer at 4 °C when not in use.

2.4. Electrochemical Estimation of Heavy Metal Ions

All the electrochemical measurements were carried out at room temperature in acetate buffer (50 mM, pH = 5.2, freshly prepared every week) at an applied potential of +0.47 V vs Ag/AgCl [31]. Next, 200 µL of glucose solution containing different amount of metal ions were injected in the flow injection apparatus and the inhibition of metal ions on glucose response was calculated with the following equation (Equation (1)):
Inhibition % = (I0 − I)/I0 × 100
where I and I0 represent the response of glucose on Pt/PPD/GOx biosensor with and without metal ions. All measurements were recorded in triplicate. The calibration curve was 1/i vs [I] (Dixon plot). The slopes were used to collect the sensitivity of the biosensor (S, µA·mM−1) towards metal ions in the optimisation procedure.

2.5. Experimental Design

Minitab 17 software (Minitab Inc., USA) was used for design, mathematical modelling, and optimisation. The independent factors (variables) considered in this work were flow rate, 0.3–1 mL min−1; number of cycles, 10–30; enzyme concentration, 50–800 U·mL−1 (X1, X2, and X3, respectively) (Table S1), whereas the sensitivities of the biosensor towards metal ions were the responses (dependent factors). oPD concentration (5 mmol·L−1 of o-phenylenediamine) during biosensor preparation and applied potential (0.47 V vs. Ag/AgCl) during response measurements were kept constant to avoid the influence of additional factors (more than three) and to have a simpler experimental design. Based on the CCD principle, the design consisted of ‘2k’ fractional factorial points plus ‘2k’ axial points and ‘1′ center point, where ‘k’ is the number of variables (in our case, k = 3). Thus, 20 experiments were conducted with 8 (23) fractional factorial points, 8 (2 × 4) axial points, and 6 replications of the central point to have a final estimation of the experimental error. A circumscribed design with the star and factorial points lying equidistant from the center was used. The range of the experimental domain was fixed onto the axial points. A second order polynomial (Equation (2)), consisting of linear, quadratic and first order interaction terms, was fitted to the measured individual response variables.
y = β 0 + i = 1 k β i x i + i = 1 k β i i x i 2 +   i = 1 k j ( i ) = 1 k β i j x i x j + ε
where y is one of the response variables (i.e., sensitivities), xi represent the dependent variables, β0; βi; βii; βij are the regression coefficients for intercept, linear, quadratic and interaction terms, respectively, k denotes the number of variables and ε represents the unexplained error. The regressions coefficients were estimated by the method of multiple-least square regression that finds the regression coefficients by minimising the sum of squares of the errors. The significance of the overall model, and of each regression coefficient was assessed by analysis of variance (ANOVA).

3. Results and Discussion

3.1. Glucose Responses and Inhibitive Detection of Heavy Metal Ions in A Fia Apparatus

The amperometric biosensors were prepared as reported elsewhere [33] by using different numbers of cycles during the electrosynthesis of the film and different enzyme concentrations. The FIA measurements were recorded in 50 mM acetate buffer (pH = 5.2) at the applied potential of 0.47 V and at different flow rates. The calibration curve to glucose at optimised conditions in the concentration range from 0.01 mM to 50 mM is reported in Figure 1A, whereas the FIA peaks recorded in the same concentration range were presented in Figure 1C. The linear range was from 10 µM to 10 mM, showing a sensitivity to glucose of 0.734 ± 0.010 mM·µA−1 (R2 = 0,997). Lineweaver-Burk plot (1/i vs 1/C) was used to determinate the apparent Michaelis-Menten constant, Km, as the glucose concentration at which the reaction rate is at half-maximum, and the maximum reaction rate achieved by the system in terms of current, imax (Figure 1B). The apparent Km and the maximum rate imax were found to be 33.02 ± 0.08 mM (R2 = 0.999) glucose and 26.08 µA, respectively. The relationship Δi vs. [glucose] curves after 10 mM and saturates at about 25 mM. The response of the biosensor is reproducible in the entire investigated range (RSD% =25 at 10 µM and RSD% = 0.21 at 50 mM), so that the sensor can be beneficial also at high glucose concentrations, which opens up opportunities for applications in food analysis.
In order to show the degree of inhibition of the enzyme to heavy metal ions, we report a typical response of the biosensor to 30 µM of Al3+ ions (Figure 2).

3.2. Optimisation of the Performance of Biosensor Using DOE

Basically, the optimisation process involves three major steps: (1) performing the statistically designed experiments, (2) estimating the coefficients in a mathematical model, and (3) predicting the response and checking the appropriateness of the model. The electrochemical responses of a biosensor can be influenced by many experimental parameters that should be optimised in order to obtain better performances. The CCD was selected because it is a design that includes linear, quadratic and interaction terms and allows greater numbers of levels without performing experiments at every combination of factor levels [23].
Among the electrosynthesis parameters, the enzyme concentration and number of cycles were optimised. The amount of the enzyme and the number of cycles during the electrosynthesis were taken into account in order to understand if the (small) change in the film thickness can affect the polymer permselectivity and/or the amount of immobilised enzyme. The levels of these independent variables were selected starting from those used in previous works and extending their range in both sides (see Table S1). Flow rate also affects responses: on considering that low flow rate gave better responses [33], 1 mL·min−1, the flow rate used in the previous works, was retained as upper level whereas 0.3 mL·min−1 was selected as lower level. We chose to model the new tested metal ions (Bi3+ and Al3+) due to their potential to cause environmental damage. Table 1 shows the experimental design matrix along with measured sensitivities for Al3+ and Bi3+ for each set of independent variables.
The coefficients in a mathematical model (Equation (1)) were estimated by regression analyses and complete software output is reported in Supplementary Materials. The regression equations in uncoded units were (Equations (3) and (4)):
Y1 = −0.171 + 0.000214 X1 − 0.379 X3 + 0.0348 X2 + 0.000000 X1*X1 − 0.013 X3*X3 − 0.000191 X2*X2 + 0.000617 X1*X3 − 0.000057 X1*X2 + 0.0060 X3*X2
Y2 = 2.44 − 0.00464 X1 − 2.44 X3 + 0.0191 X2 + 0.000006 X1*X1 + 1.49 X3*X3 + 0.00117 X2*X2 + 0.00079 X1*X3 − 0.000131 X1*X2 − 0.0021 X3*X2
where Y1 is the sensitivity toward Al3+ and Y2 is the sensitivity toward Bi3+. ANOVA was also performed for each response, and the complete outputs are presented in Supplementary Materials. In a situation where a model fits the data, the mean square of the lack-of-fit error will be close to that of the pure error and the resulting ratio, the F-statistic, will be small. If the relevant P-value, i.e., the error probability, is large compared to the significance level (alpha value = 0.05), then the lack-of-fit is not significant, and the model adequately explains data in the region of experimentation. In the present case, the non-significant lack-of-fit for Al3+ (F = 2.39; P = 0.180 > 0.05) and Bi3+ (F = 3.38; P = 0.104 > 0.05) sensitivities suggest the adequacy of the model for both metal ions to explain data in the region of experimentation.
The Analysis of Variance tables (see Supplementary Materials) summarises the linear terms, the squared terms, and the interactions. The small p-values for the square or interaction terms suggest there is curvature in the response surface. According to the calculated p-values, the enzyme in the linear or square term of the model shows that this variable is statistically significant. Figure S1 shows the surface plots for metal ion sensitivities as function of enzyme concentration and number of cycles, where a curvature of responses is evident. In numerical optimisation, it is possible to choose the desired goal for each response. The overall desirability is an objective function ranging from 0 (if the optimal values are outside the selected ranges) to 1 (if the goal for each response is reached). All the sensitivities were maximised with the same weight and the composite desirability was equal to 1 at 50 U·mL−1 for enzyme concentration and 30 cycles during the synthesis of biosensor and 0.3 mL·min−1 for the flow rate during inhibition experiments.
To confirm the reliability of the model, additional laboratory experiments were conducted using the optimum conditions. The experimental results were in agreement with the predicted results (Table 2): it can be seen that the predicted confidence interval (PI) covered the values of sensitivity obtained from the experiments, confirming the validity of the model.

3.3. Analytical Performances of the Optimised Biosensor

The inhibitive experiments were carried out at varying concentrations of Bi3+ (3.9 to 250 µM) and Al3+ (15.62 µM to 4 mM) in the presence of 20 mM of glucose dissolved in 50 mM acetate buffer (pH = 5.2). As a comparison with previous results, optimisation was also performed for Ag+ and Ni2+, obtaining the same optimal conditions (see below). Figure 3 shows the calibration graph constructed by plotting the inhibition percentage of the enzyme activity against the concentration of the new tested Bi3+ and Al3+ ions.
Table 3 collects the detection limit (LOD, µM) which were evaluated according to recent advances considering the concentration of inhibitor which causes 10% of inhibition [16], instead of the frequently used 3x standard deviation (SD) of the blank. In the same table, the sensitivities of the biosensor obtained toward Bi3+ and Al3+ ions are reported.
The performances of the optimal biosensor were also tested for Ni2+ and Ag+ ions, already investigated in our previous work [33]. While sensitivities were in well agreement with previous results, significant improvements were obtained for Ni2+ ions. In particular, we observed a single linear range for Ni2+ (Figure 4) in place of the two observed in the past (see reference [33]): possibly the lower amount of enzyme entrapped onto the electrode simplify the complex interaction mechanism between metal and enzyme. Also, the reusability of the biosensor was improved with respect to previous results (see below).
Using the same system, the working and the storage stability were investigated; these were not evaluated in previous work. Figure 5A shows the working stability of the optimised PPD/GOx biosensor. Nine measurements over about 2 h were recorded with a low relative standard deviation (RSD) of 0.96%. The storage stability of the biosensor was studied by measuring current response at different times after conservation in acetate buffer at 4 °C in a refrigerator. The responses remained acceptable over more than 7 months, as the biosensor could keep 60% of its original current response with RSD of 0.72% for ten consecutive measurements. These results indicated that the Pt/PPD/GOx biosensor had good working and storage stability.
The reversibility of enzyme activity is also important for the reusability of an inhibitive biosensor. The recovering of enzyme activity was tested by detecting the current response of the PPD/GOx biosensor towards glucose, before and after Ag+ detection [33]. Figure 5B shows the reusability of biosensors after the injection of different concentration of Ag+ ions (up to 0.5 mM) in the presence of 10 mM of glucose. It was possible to achieve 92% and 98% of recovery (slightly higher than the previous one [33]) of its initial current response after simply flushing the acetate buffer for 6 min and 11 min, respectively, without using any chelating agent, such as EDTA. Figure 5C shows the reusability of the biosensor after the injection of different concentrations of Bi3+ and Al3+ after the injection of glucose 20 mM. Also, in this case, we observed the recovery of the responses of the biosensor, confirming the possibility of reusing the sensor after a simple washing with the buffer media.
Even if the biosensors based on enzymes as biological receptors were extensively studied in the literature, these devices suffer from poor selectivity towards the inhibitors of the enzymatic activity. For instance, the optimised biosensor responds towards Zn2+, Cd2+, Cr3+ and Co2+ ions. So, these devices are a portable alternative for the in-situ monitoring of the contaminants, without requiring any sample preparation and transportation; furthermore, the biosensors could be successfully used in screening analysis for the total contamination of metal ions in water samples. The analytical application of the biosensors based on enzyme inhibition is still limited, since these devices were affected by various toxic compounds in the same sample. As an example, the inhibitive activity on different enzymes against herbicides is reported in the literature [34]. In our case, the selectivity of the biosensor was evaluated in the presence of atrazine and 4-(2,4-dichlorophenoxy) butyric acid (2,4-DB). Any inhibition effect was visible after the injection of 0.8 µM of atrazine and 39.6 µM of 2,4-DB (Figure S3), confirming the application of the biosensor for the selective detection of metal ions.

4. Conclusions

In this work, the performance of an amperometric biosensor was successfully maximised using an experimental design. The inhibitive behaviour of our previously studied biosensor was modelled by response surface methodology, and selected experimental parameters were optimised by a CCD design. The optimal setup involved a GOx concentration of 50 U·mL−1 and 30 cycles during the electrosynthesis of the biosensor, whereas 0.3 mL·min−1 was the optimal flow rate of measurements. The model was validated and employed successfully to also detect Al3+ and Bi3+, two metal ions whose inhibition capability is here presented for the first time. The inhibitive biosensor was demonstrated to have similar sensitivities compared to the previous biosensor, but improved reproducibility, stability and reversibility. In this configuration, it can be effectively used as gross biosensor to detect heavy metal pollution. Furthermore, thanks to its stability and reversibility, the sensor can be used as amperometric detector in ion chromatographic system to overcome the lack of selectivity.

Supplementary Materials

The following are available online at https://www.mdpi.com/2079-6374/9/1/26/s1, Table S1: Variables and levels considered for the design of experiment (DOE), Figure S1: Response Surface for Al3+ (A), and Bi3+ (B): the sensitivities were improved when low concentrations of enzyme and a high number of cycles were employed during the synthesis of biosensor, Figure S2: Numerical optimisation performed by the software, Figure S3: Electrochemical responses of PPD/GOx biosensor after the injection of 0.8 µM of atrazine and 36.8 µM of 2,4-DB. Baseline: Glucose 1 mM.

Author Contributions

C.M. contributed in the methodology aspects, curated the funding acquisition, the review editing processes, the project administration and he supervised the work. G.E.D.B. critically supervised the manuscript and helped to write the manuscript. A.P. participates for the data curation in the experimental design. S.D.M. contributes to the investigation, the data curation and to write the manuscript.

Funding

This research was funded by Ministero dell’Istruzione, Università e Ricerca (MIUR) through the project PON Dottorati innovativi con caratterizzazione industriale, grant number DOT1412034.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Nriagu, J.O.; Pacyna, J.M. Quantitative assessment of worldwide contamination of air, water and soils by trace metals. Nature 1988, 333, 134–139. [Google Scholar] [CrossRef] [PubMed]
  2. Ouyang, W.; Wang, Y.; Lin, C.; He, M.; Hao, F.; Liu, H.; Zhu, W. Heavy metal loss from agricultural watershed to aquatic system: A scientometrics review. Sci. Total Environ. 2018, 637–638, 208–220. [Google Scholar] [CrossRef] [PubMed]
  3. Linscheid, M.W. Molecules and elements for quantitative bioanalysis: The allure of using electrospray, MALDI, and ICP mass spectrometry side-by-side. Mass Spectrom. Rev. 2018, 38(2), 1–18. [Google Scholar] [CrossRef] [PubMed]
  4. Rodriguez-Mozaz, S.; De Alda, M.J.L.; Marco, M.P.; Barceló, D. Biosensors for environmental monitoring: A global perspective. Talanta 2005, 65, 291–297. [Google Scholar] [CrossRef]
  5. Wang, X.; Lu, X.; Chen, J. Development of biosensor technologies for analysis of environmental contaminants. Trends Environ. Anal. Chem. 2014, 2, 25–32. [Google Scholar] [CrossRef]
  6. Bachan Upadhyay, L.S.; Verma, N. Enzyme Inhibition Based Biosensors: A Review. Anal. Lett. 2012, 46, 225–241. [Google Scholar] [CrossRef]
  7. Malitesta, C.; Palmisano, F.; Torsi, L.; Zambonin, P.G. Glucose Fast-Response Amperometric Sensor Based on Glucose Oxidase Immobilized in an Electropolymerized Poly(o-phenylenediamine) Film. Anal. Chem. 1990, 62, 2735–2740. [Google Scholar] [CrossRef]
  8. Ramanavičius, A.; Mikoliunaite, L.; Gicevičius, M.; Popov, A.; Ramanavičiene, A. Conducting polymers in the design of enzymatic sensors. In Proceedings of the IEEE 7th International Conference on Nanomaterials: Applications and Properties (NAP), Odessa, Ukraine, 10–15 September 2017. [Google Scholar] [CrossRef]
  9. Reyes-De-Corcuera, J.I.; Olstad, H.E.; García-Torres, R. Stability and Stabilization of Enzyme Biosensors: The Key to Successful Application and Commercialization. Annu. Rev. Food Sci. Technol. 2018, 9, 293–322. [Google Scholar] [CrossRef]
  10. Bolella, P.; Gorton, L. Enzyme based amperometric biosensors. Curr. Opin. Electroc. 2018, 10, 157–173. [Google Scholar] [CrossRef]
  11. Mugheri, A.Q.; Tahira, A.; Sirajuddin; Sherazi, S.T.H.; Ishaque Abro, M.; Willander, M.; Ibupoto, Z.H. An amperometric indirect determination of heavy metal ions through inhibition of glucose oxidase immobilized on cobalt oxide nanostructures. Sens. Lett. 2016, 14. [Google Scholar] [CrossRef]
  12. Rust, I.M.; Goran, J.M.; Stevenson, K.J. Amperometric Detection of Aqueous Silver Ions by Inhibition of Glucose Oxidase Immobilized on Nitrogen-Doped Carbon Nanotube Electrodes. Anal. Chem. 2015, 87, 7250–7257. [Google Scholar] [CrossRef] [PubMed]
  13. Ayenimo, J.G.; Adeloju, S.B. Inhibitive potentiometric detection of trace metals with ultrathin polypyrrole glucose oxidase biosensor. Talanta 2015, 137, 62–70. [Google Scholar] [CrossRef] [PubMed]
  14. Ghica, M.E.; Carvalho, R.C.; Amine, A.; Brett, C.M.A. Glucose oxidase enzyme inhibition sensors for heavy metals at carbon film electrodes modified with cobalt or copper hexacyanoferrate. Sens. Actuators B Chem. 2013, 178, 270–278. [Google Scholar] [CrossRef]
  15. Ghica, M.E.; Brett, C.M.A. Glucose oxidase inhibition in poly(neutral red) mediated enzyme biosensors for heavy metal determination. Microchim. Acta. 2008, 163, 185–193. [Google Scholar] [CrossRef] [Green Version]
  16. Amine, A.; Mohammadi, H.; Bourais, I.; Palleschi, G. Enzyme inhibition-based biosensors for food safety and environmental monitoring. Biosens. Bioelectron. 2006, 21, 1405–1423. [Google Scholar] [CrossRef] [PubMed]
  17. Amine, A.; Arduini, F.; Moscone, D.; Palleschi, G. Recent advances in biosensors based on enzyme inhibition. Biosens. Bioelectron. 2016, 76, 180–194. [Google Scholar] [CrossRef] [PubMed]
  18. Bansod, B.; Kumar, T.; Thakur, R.; Rana, S.; Singh, I. Biosensors and Bioelectronics A review on various electrochemical techniques for heavy metal ions detection with different sensing platforms. Biosens. Bioelectron. 2017, 94, 443–455. [Google Scholar] [CrossRef] [PubMed]
  19. Luque de Castro, M.D.; Herrera, M.C. Enzyme inhibition-based biosensors and biosensing systems: Questionable analytical devices. Biosens. Bioelectron. 2003, 18, 279–294. [Google Scholar] [CrossRef]
  20. Mehta, J.; Bhardwaj, S.K.; Bhardwaj, N.; Paul, A.K.; Kumar, P.; Kim, K.H.; Deep, A. Progress in the biosensing techniques for trace-level heavy metals. Biotechnol. Adv. 2016, 34, 47–60. [Google Scholar] [CrossRef] [PubMed]
  21. Nurani, S.; Abdul, H.; Lau, W. Chemical Detection of contaminants in water supply: A review on state-of-the-art monitoring technologies and their applications. Sens. Actuators B. Chem. 2018, 255, 2657–2689. [Google Scholar] [CrossRef]
  22. Bezerra, M.A.; Santelli, R.E.; Oliveira, E.P.; Villar, L.S.; Escaleira, L.A. Response surface methodology (RSM) as a tool for optimization in analytical chemistry. Talanta 2008, 76, 965–977. [Google Scholar] [CrossRef] [PubMed]
  23. Tarley, C.R.T.; Silveira, G.; dos Santos, W.N.L.; Matos, G.D.; da Silva, E.G.P.; Bezerra, M.A.; Miró, M.; Ferreira, S.L.C. Chemometric tools in electroanalytical chemistry: Methods for optimization based on factorial design and response surface methodology. Microchem. J. 2009, 92, 58–67. [Google Scholar] [CrossRef]
  24. Vera, L.; De Zan, M.M.; Cámara, M.S.; Goicoechea, C. Talanta Experimental design and multiple response optimization. Using the desirability function in analytical methods development. Talanta 2014, 124, 123–138. [Google Scholar] [CrossRef] [PubMed]
  25. Monaci, L.; Vatinno, R.; De Benedetto, G.E. Fast detection of cyclopiazonic acid in cheese using Fourier transform mid-infrared ATR spectroscopy. Eur. Food. Res. Technol. 2007, 225, 585–588. [Google Scholar] [CrossRef]
  26. Mignani, A.; Luciano, G.; Lanteri, S.; Leardi, R.; Scavetta, E.; Tonelli, D. Optimization of a glucose biosensor setup based on a Ni/Al HT matrix. Anal. Chim. Acta. 2007, 599, 36–40. [Google Scholar] [CrossRef] [PubMed]
  27. Prakash, O.; Talat, M.; Hasan, S.H.; Pandey, R.K. Factorial design for the optimization of enzymatic detection of cadmium in aqueous solution using immobilized urease from vegetable waste. Bioresour. Technol. 2008, 99, 7565–7572. [Google Scholar] [CrossRef]
  28. Uliana, C.V.; Tognolli, J.O.; Yamanaka, H. Application of factorial design experiments to the development of a disposable amperometric DNA biosensor. Electroanalysis 2011, 23, 2607–2615. [Google Scholar] [CrossRef]
  29. Horry, H.; Maul, A. Optimization of a bacterial bioluminescent biosensor through experimental design. Sens. Actuators B. Chem. 2007, 127, 649–657. [Google Scholar] [CrossRef]
  30. Ebrahimi, B.; Shojaosadati, S.A.; Ranaie, S.O.; Mousavi, S.M. Optimization and evaluation of acetylcholine esterase immobilization on ceramic packing using response surface methodology. Process. Biochem. 2010, 45, 81–87. [Google Scholar] [CrossRef]
  31. Malitesta, C.; Guascito, M.R. Heavy metal determination by biosensors based on enzyme immobilised by electropolymerisation. Biosens. Bioelectron. 2005, 20, 1643–1647. [Google Scholar] [CrossRef]
  32. Guascito, M.R.; Malitesta, C.; Mazzotta, E.; Turco, A. Inhibitive determination of metal ions by an amperometric glucose oxidase biosensor. Study of the effect of hydrogen peroxide decomposition. Sens. Actuators B Chem. 2008, 131, 394–402. [Google Scholar] [CrossRef]
  33. Guascito, M.R.; Malitesta, C.; Mazzotta, E.; Turco, A. Screen-Printed Glucose Oxidase-Based Biosensor for Inhibitive Detection of Heavy Metal Ions in a Flow Injection System. Sens. Lett. 2009, 7, 153–159. [Google Scholar] [CrossRef]
  34. Yang, Q.; Qu, Y.; Bo, Y.; Wen, Y.; Huang, S. Biosensor for atrazin based on aligned carbon nanotubes modified with glucose oxidase. Microchim. Acta. 2010, 168, 197–203. [Google Scholar] [CrossRef]
Figure 1. (A) Amperometric response of optimised Pt/PPD/GOx biosensor (50 U·mL−1, 30 cycles of CV) to glucose standard solution prepared in acetate buffer (0.05 M, pH = 5.2) and linear fit to the calibration curve (0.01–10 mM); (B) Lineaweaver-Burk plot; (C) FIA peaks recorded for triplicate injections of different concentrations of glucose (0.01–50 mM) at a flow rate of 0.3 mL·min−1.
Figure 1. (A) Amperometric response of optimised Pt/PPD/GOx biosensor (50 U·mL−1, 30 cycles of CV) to glucose standard solution prepared in acetate buffer (0.05 M, pH = 5.2) and linear fit to the calibration curve (0.01–10 mM); (B) Lineaweaver-Burk plot; (C) FIA peaks recorded for triplicate injections of different concentrations of glucose (0.01–50 mM) at a flow rate of 0.3 mL·min−1.
Biosensors 09 00026 g001
Figure 2. FIA peaks recorded for glucose (20 mM) and in presence of 30 µM of Al3+ ions prepared in acetate buffer (0.05 M, pH = 5.2). Experimental conditions as in Figure 1.
Figure 2. FIA peaks recorded for glucose (20 mM) and in presence of 30 µM of Al3+ ions prepared in acetate buffer (0.05 M, pH = 5.2). Experimental conditions as in Figure 1.
Biosensors 09 00026 g002
Figure 3. Inhibitive effects and Dixon plot for Pt/PPD/GOx biosensor after the injection of different concentration of Bi3+ and Al3+ ions in presence of 20 mM of glucose.
Figure 3. Inhibitive effects and Dixon plot for Pt/PPD/GOx biosensor after the injection of different concentration of Bi3+ and Al3+ ions in presence of 20 mM of glucose.
Biosensors 09 00026 g003
Figure 4. Inhibitive effects and Dixon plots for Pt/PPD/GOx biosensor after the injection of different concentration of Ni2+ (A) and Ag+ (B) in presence of 20 mM of glucose.
Figure 4. Inhibitive effects and Dixon plots for Pt/PPD/GOx biosensor after the injection of different concentration of Ni2+ (A) and Ag+ (B) in presence of 20 mM of glucose.
Biosensors 09 00026 g004
Figure 5. (A) Evaluation of working life of biosensor after subsequently injection of 31.25 µM of Ni2+. (B) Inhibition effect of Ag+ ions and recovery after washing with acetate buffer (50 mM, pH = 5.2). (C) Reusability of the biosensor after the injection of 15 µM and 31.25 µM of Bi3+ and Al3+ ions in presence of 20 mM of glucose and acetate buffer media.
Figure 5. (A) Evaluation of working life of biosensor after subsequently injection of 31.25 µM of Ni2+. (B) Inhibition effect of Ag+ ions and recovery after washing with acetate buffer (50 mM, pH = 5.2). (C) Reusability of the biosensor after the injection of 15 µM and 31.25 µM of Bi3+ and Al3+ ions in presence of 20 mM of glucose and acetate buffer media.
Biosensors 09 00026 g005aBiosensors 09 00026 g005b
Table 1. Experimental plan and sensitivity in the range of concentration between 15.62 µM and 31.25 µM reported for Al3+ and Bi3+.
Table 1. Experimental plan and sensitivity in the range of concentration between 15.62 µM and 31.25 µM reported for Al3+ and Bi3+.
ExperimentFactors Responses (Sensitivity, µA·mM−1)
RdesStdOrderRunOrderPtTypeEnzyme Concentration (U/mL)Flow Rate (mL/min)Number of Cycles during CVAl3+Bi3+
3112020.86140.0681.059
122−14251200.0350.657
143−14250.65300.0260.509
20404250.65200.0600.438
19504250.65200.0250.454
116−14250.3200.0770.599
5712020.44260.4171.675
1812020.44140.0741.222
139−14250.65100.0510.616
161004250.65200.2250.443
71112020.86260.3231.553
151204250.65200.1240.541
1013−18000.65200.0330.838
21416480.44140.0780.697
171504250.65200.0940.738
916−1500.65200.1671.642
61716480.44260.0010.506
41816480.86140.0690.532
81916480.86260.1390.479
182004250.65200.1180.579
Table 2. Optimal and laboratory experiment validation.
Table 2. Optimal and laboratory experiment validation.
Experimental Value (µA/mM)Fit (µA/mM)95% PI (From Design)
S Bi3+1.8042.684(1.759; 3.608)
S Al3+0.5240.574(0.095; 1.054)
Table 3. Analytical data collected from Dixon plots obtained for Bi3+, Al3+ ion in presence of 20 mM of glucose.
Table 3. Analytical data collected from Dixon plots obtained for Bi3+, Al3+ ion in presence of 20 mM of glucose.
Metal IonLOD (µM)Upper limit of Linearity (µM)Sensitivity (µA mM−1)
Bi3+3.9 1251.80 ± 0.12
Al3+16 5000.52 ± 0.02

Share and Cite

MDPI and ACS Style

De Benedetto, G.E.; Di Masi, S.; Pennetta, A.; Malitesta, C. Response Surface Methodology for the Optimisation of Electrochemical Biosensors for Heavy Metals Detection. Biosensors 2019, 9, 26. https://doi.org/10.3390/bios9010026

AMA Style

De Benedetto GE, Di Masi S, Pennetta A, Malitesta C. Response Surface Methodology for the Optimisation of Electrochemical Biosensors for Heavy Metals Detection. Biosensors. 2019; 9(1):26. https://doi.org/10.3390/bios9010026

Chicago/Turabian Style

De Benedetto, Giuseppe Egidio, Sabrina Di Masi, Antonio Pennetta, and Cosimino Malitesta. 2019. "Response Surface Methodology for the Optimisation of Electrochemical Biosensors for Heavy Metals Detection" Biosensors 9, no. 1: 26. https://doi.org/10.3390/bios9010026

APA Style

De Benedetto, G. E., Di Masi, S., Pennetta, A., & Malitesta, C. (2019). Response Surface Methodology for the Optimisation of Electrochemical Biosensors for Heavy Metals Detection. Biosensors, 9(1), 26. https://doi.org/10.3390/bios9010026

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop