Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices
Abstract
:1. Introduction
2. Methods
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Kalin, R.; Stanton, M.S. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin. Electrophysiol. 2005, 28, 326–328. [Google Scholar] [CrossRef]
- Stühlinger, M.; Burri, H.; Vernooy, K.; Garcia, R.; Lenarczyk, R.; Sultan, A.; Brunner, M.; Sabbag, A.; Özcan, E.E.; Ramos, J.T.; et al. EHRA consensus on prevention and management of interference due to medical procedures in patients with cardiac implantable electronic devices. Europace 2022, 24, 1512–1537. [Google Scholar] [CrossRef]
- Gillam, M.H.; Inacio, M.C.; Pratt, N.L.; Shakib, S.; Roughead, E.E. Magnetic Resonance Imaging in People With Cardiac Implantable Electronic Devices: A Population Based Cohort Study. Heart Lung Circ. 2018, 27, 748–751. [Google Scholar] [CrossRef] [PubMed]
- Roguin, A.; Schwitter, J.; Vahlhaus, C.; Lombardi, M.; Brugada, J.; Vardas, P.; Auricchio, A.; Priori, S.; Sommer, T. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace 2008, 10, 336–346. [Google Scholar] [CrossRef] [PubMed]
- Maass, A.H.; Hemels, M.E.W.; Allaart, C.P. Magnetic resonance imaging in patients with cardiac implantable electronic devices. Neth. Heart J. 2018, 26, 584–590. [Google Scholar] [CrossRef] [PubMed]
- Wilkoff, B.L.; Bello, D.; Taborsky, M.; Vymazal, J.; Kanal, E.; Heuer, H.; Hecking, K.; Ben Johnson, W.; Young, W.; Ramza, B.; et al. Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment. Heart Rhythm. 2011, 8, 65–73. [Google Scholar] [CrossRef]
- Poole, J.E.; Gleva, M.J.; Mela, T.; Chung, M.K.; Uslan, D.Z.; Borge, R.; Gottipaty, V.; Shinn, T.; Dan, D.; Feldman, L.A.; et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: Results from the REPLACE registry. Circulation 2010, 122, 1553–1561. [Google Scholar] [CrossRef]
- Shenthar, J.; Milasinovic, G.; Al Fagih, A.; Götte, M.; Engel, G.; Wolff, S.; Tse, H.-F.; Herr, J.; Carrithers, J.; Cerkvenik, J.; et al. MRI scanning in patients with new and existing CapSureFix Novus 5076 pacemaker leads: Randomized trial results. Heart Rhythm. 2015, 12, 759–765. [Google Scholar] [CrossRef]
- Glikson, M.; Nielsen, J.C.; Kronborg, M.B.; Michowitz, Y.; Auricchio, A.; Barbash, I.M.; Barrabés, J.A.; Boriani, G.; Braunschweig, F.; Brignole, M.; et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) with the special contribution of the European Heart Rhythm Association (EHRA). Eur. Heart J. 2021, 42, 3427–3520. [Google Scholar]
- Russo, R.J.; Costa, H.S.; Silva, P.D.; Anderson, J.L.; Arshad, A.; Biederman, R.W.; Boyle, N.G.; Frabizzio, J.V.; Birgersdotter-Green, U.; Higgins, S.L.; et al. Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. N. Engl. J. Med. 2017, 376, 755–764. [Google Scholar] [CrossRef]
- Mason, S.; Osborn, J.S.; Dhar, R.; Tonkin, A.; Ethington, J.; Le, V.; Benuzillo, J.; Lappe, D.L.; Knowlton, K.U.; Bunch, T.J.; et al. Real world MRI experience with nonconditional and conditional cardiac rhythm devices after MagnaSafe. J. Cardiovasc. Electrophysiol. 2017, 28, 1468–1474. [Google Scholar] [CrossRef] [PubMed]
- Naehle, C.P.; Kreuz, J.; Strach, K.; Schwab, J.O.; Pingel, S.; Luechinger, R.; Fimmers, R.; Schild, H.; Thomas, D. Safety, feasibility, and diagnostic value of cardiac magnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 T. Am. Heart J. 2011, 161, 1096–1105. [Google Scholar] [CrossRef] [PubMed]
- Indik, J.H.; Gimbel, J.R.; Abe, H.; Alkmim-Teixeira, R.; Birgersdotter-Green, U.; Clarke, G.D.; Dickfeld, T.-M.L.; Froelich, J.W.; Grant, J.; Hayes, D.L.; et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017, 14, e97–e153. [Google Scholar] [CrossRef] [PubMed]
- Sommer, T.; Naehle, C.P.; Yang, A.; Zeijlemaker, V.; Hackenbroch, M.; Schmiedel, A.; Meyer, C.; Strach, K.; Skowasch, D.; Vahlhaus, C.; et al. Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: A prospective study with 115 examinations. Circulation 2006, 114, 1285–1292. [Google Scholar] [CrossRef] [PubMed]
- Martin, E.T.; Coman, J.A.; Shellock, F.G.; Pulling, C.C.; Fair, R.; Jenkins, K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. J. Am. Coll. Cardiol. 2004, 43, 1315–1324. [Google Scholar] [CrossRef]
- Lupo, P.; Cappato, R.; Di Leo, G.; Secchi, F.; Papini, G.D.E.; Foresti, S.; Ali, H.; De Ambroggi, G.M.G.; Sorgente, A.; Epicoco, G.; et al. An eight-year prospective controlled study about the safety and diagnostic value of cardiac and non-cardiac 1.5-T MRI in patients with a conventional pacemaker or a conventional implantable cardioverter defibrillator. Eur. Radiol. 2018, 28, 2406–2416. [Google Scholar] [CrossRef]
- Bhuva, A.N.; Moralee, R.; Brunker, T.; Lascelles, K.; Cash, L.; Patel, K.P.; Lowe, M.; Sekhri, N.; Alpendurada, F.; Pennell, D.J.; et al. Evidence to support magnetic resonance conditional labelling of all pacemaker and defibrillator leads in patients with cardiac implantable electronic devices. Eur. Heart J. 2022, 43, 2469–2478. [Google Scholar] [CrossRef]
- DEL Ojo, J.L.; Moya, F.; Villalba, J.; Sanz, O.; Pavón, R.; Garcia, D.; Pastor, L. Is Magnetic Resonance Imaging Safe in Cardiac Pacemaker Recipients? Pacing Clin. Electrophysiol. 2005, 28, 274–278. [Google Scholar] [CrossRef]
- Gimbel, J.R. Magnetic resonance imaging of implantable cardiac rhythm devices at 3.0 tesla. Pacing Clin Electrophysiol. 2008, 31, 795–801. [Google Scholar] [CrossRef]
- Hwang, Y.M.; Kim, J.; Lee, J.H.; Kim, M.; Nam, G.-B.; Choi, K.-J.; Kim, Y.-H. Cardiac Implantable Electronic Device Safety during Magnetic Resonance Imaging. Korean Circ. J. 2016, 46, 804–810. [Google Scholar] [CrossRef]
- Naehle, C.P.; Meyer, C.; Thomas, D.; Remerie, S.; Krautmacher, C.; Litt, H.; Luechinger, R.; Fimmers, R.; Schild, H.; Sommer, T. Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: Pilot prospective study with 51 examinations. Radiology 2008, 249, 991–1001. [Google Scholar] [CrossRef] [PubMed]
- Roguin, A.; Zviman, M.M.; Meininger, G.R.; Rodrigues, E.R.; Dickfeld, T.M.; Bluemke, D.A.; Lardo, A.; Berger, R.D.; Calkins, H.; Halperin, H.R. Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: In vitro and in vivo assessment of safety and function at 1.5 T. Circulation 2004, 110, 475–482. [Google Scholar] [CrossRef] [PubMed]
- Levine, G.N.; Gomes, A.S.; Arai, A.E.; Bluemke, D.A.; Flamm, S.D.; Kanal, E.; Manning, W.J.; Martin, E.T.; Smith, J.M.; Wilke, N.; et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: An American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: Endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance. Circulation 2007, 116, 2878–2891. [Google Scholar] [PubMed]
- Nazarian, S.; Hansford, R.; Roguin, A.; Goldsher, D.; Zviman, M.M.; Lardo, A.C.; Caffo, B.S.; Frick, K.D.; Kraut, M.A.; Kamel, I.R.; et al. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann. Intern. Med. 2011, 155, 415–424. [Google Scholar] [CrossRef]
- Pieri, C.; Bhuva, A.; Moralee, R.; Abiodun, A.; Gopalan, D.; Roditi, G.H.; Moon, J.C.; Manisty, C. Access to MRI for patients with cardiac pacemakers and implantable cardioverter defibrillators. Open Heart 2021, 8, e001598. [Google Scholar] [CrossRef]
Characteristics | Overall (n = 1010) | PM (n = 604) | ICD (n = 110) | sICD (n = 23) | CRT-P (n = 30) | CRT-D (n = 37) | ILR (n = 206) | p-Value |
---|---|---|---|---|---|---|---|---|
Demographics | ||||||||
Age [years], median [IQR] | 74.00 [62.00, 80.00] | 77.00 [70.00, 82.00] | 67.00 [58.00, 73.00] | 45.00 [42.50, 47.00] | 77.00 [75.25, 78.75] | 70.00 [58.00, 80.00] | 62.00 [55.00, 74.00] | <0.001 |
Sex at birth [male], n (%) | 672 (66.5) | 401 (66.4) | 88 (80.0) | 17 (73.9) | 24 (80.0) | 28 (75.7) | 114 (55.3) | <0.001 |
Aggregate | ||||||||
Manufacturer | ||||||||
Boston Scientific, n (%) | 78 (7.7) | 33 (5.5) | 11 (10.0) | 23 (100.0) | 3 (10.0) | 9 (24.3) | 0 (0.0) | <0.001 |
Medtronic, n (%) | 558 (55.2) | 302 (50.0) | 44 (40.0) | 0 (0.0) | 13 (43.3) | 12 (32.4) | 187 (90.8) | <0.001 |
Biotronik, n (%) | 239 (23.7) | 186 (30.8) | 22 (20.0) | 0 (0.0) | 4 (13.3) | 11 (29.7) | 15 (7.3) | <0.001 |
St. Jude Medical, n (%) | 85 (8.4) | 35 (5.8) | 31 (28.2) | 0 (0.0) | 10 (33.3) | 5 (13.5) | 4 (1.9) | <0.001 |
MicroPort, n (%) | 50 (5.0) | 48 (7.9) | 2 (1.8) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | <0.001 |
Time since implantation [years], median [IQR] | 1.75 [0.67, 3.58] | 2.25 [0.75, 4.27] | 2.00 [1.08, 4.27] | 1.08 [0.62, 2.50] | 1.62 [0.85, 2.58] | 2.17 [1.08, 3.67] | 1.00 [0.42, 1.92] | <0.001 |
Labeled as MR-conditional by manufacturer, n (%) | 920 (91.1) | 538 (89.1) | 93 (84.5) | 23 (100.0) | 29 (96.7) | 31 (83.8) | 206 (100.0) | <0.001 |
Right atrial lead | ||||||||
Right atrial lead, n (%) | 643 (63.7) | 535 (88.6) | 43 (39.1) | N/A | 29 (96.7) | 36 (97.3) | N/A | <0.001 |
Manufacturer | ||||||||
Boston Scientific RA lead, n (%) | 40 (4.0) | 23 (3.8) | 7 (6.4) | N/A | 3 (10.0) | 7 (18.9) | N/A | <0.001 |
Medtronic RA lead, n (%) | 313 (31.0) | 279 (46.2) | 10 (9.1) | N/A | 13 (43.3) | 11 (29.7) | N/A | <0.001 |
Biotronik RA lead, n (%) | 190 (18.8) | 166 (27.5) | 8 (7.3) | N/A | 3 (10.0) | 13 (35.1) | N/A | 0.025 |
St. Jude Medical RA lead, n (%) | 61 (6.0) | 29 (4.8) | 18 (16.4) | N/A | 9 (30.0) | 5 (13.5) | N/A | <0.001 |
MicroPort RA lead, n (%) | 24 (2.4) | 24 (4.0) | 0 (0.0) | N/A | 0 (0.0) | 0 (0.0) | N/A | 0.309 |
Time since implantation [years], median [IQR] | 2.50 [0.92, 4.77] | 2.58 [0.92, 4.83] | 2.00 [1.00, 3.25] | N/A | 1.75 [0.92, 2.83] | 2.54 [1.08, 4.35] | N/A | 0.175 |
Labeled as MR-conditional by manufacturer, n (%) | 585 (57.9) | 485 (80.3) | 41 (37.3) | N/A | 27 (90.0) | 32 (86.5) | N/A | 0.157 |
Right ventricular lead | ||||||||
Right ventricular lead, n (%) | 630 (62.4) | 599 (99.2) | 1 (0.9) | N/A | 30 (100.0) | 0 (0.0) | N/A | <0.001 |
Manufacturer | ||||||||
Boston Scientific RV lead, n (%) | 38 (3.8) | 35 (5.8) | 0 (0.0) | N/A | 3 (10.0) | 0 (0.0) | N/A | 0.451 |
Medtronic RV lead, n (%) | 328 (32.5) | 305 (50.5) | 1 (0.9) | N/A | 22 (73.3) | 0 (0.0) | N/A | 0.023 |
Biotronik RV lead, n (%) | 188 (18.6) | 184 (30.5) | 0 (0.0) | N/A | 4 (13.3) | 0 (0.0) | N/A | 0.075 |
St. Jude Medical RV lead, n (%) | 33 (3.3) | 33 (5.5) | 0 (0.0) | N/A | 0 (0.0) | 0 (0.0) | N/A | 0.426 |
MicroPort RV lead, n (%) | 34 (3.4) | 34 (5.6) | 0 (0.0) | N/A | 0 (0.0) | 0 (0.0) | N/A | 0.430 |
Time since implantation [years], median [IQR] | 2.42 [0.75, 4.75] | 2.42 [0.75, 4.83] | 21.92 [21.92, 21.92] * | N/A | 1.79 [0.92, 2.77] | 0.58 [0.58, 0.58] * | N/A | 0.110 |
Labeled as MR-conditional by manufacturer, n (%) | 563 (55.7) | 536 (88.7) | 0 (0.0) | N/A | 27 (90.0) | 0 (0.0) | N/A | 0.084 |
Left ventricular lead | ||||||||
Left ventricular lead, n (%) | 68 (6.7) | 2 (0.3) | 0 (0.0) | N/A | 30 (100.0) | 36 (97.3) | N/A | <0.001 |
Manufacturer | ||||||||
Boston Scientific LV lead, n (%) | 12 (1.2) | 0 (0.0) | 0 (0.0) | N/A | 4 (13.3) | 8 (21.6) | N/A | 0.678 |
Medtronic LV lead, n (%) | 24 (2.4) | 2 (0.3) | 0 (0.0) | N/A | 11 (36.7) | 11 (29.7) | N/A | 0.159 |
Biotronik LV lead, n (%) | 15 (1.5) | 0 (0.0) | 0 (0.0) | N/A | 4 (13.3) | 11 (29.7) | N/A | 0.211 |
St. Jude Medical LV lead, n (%) | 16 (1.6) | 0 (0.0) | 0 (0.0) | N/A | 10 (33.3) | 6 (16.2) | N/A | 0.231 |
MicroPort LV lead, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | N/A | 0 (0.0) | 0 (0.0) | N/A | - |
Time since implantation [years], median [IQR] | 2.12 [0.98, 3.52] | 1.71 [1.23, 2.19] | - | N/A | 1.62 [0.85, 2.58] | 2.54 [1.33, 4.02] | N/A | 0.085 |
Labeled as MR-conditional by manufacturer, n (%) | 60 (5.9) | 2 (0.3) | 0 (0.0) | N/A | 27 (90.0) | 31 (83.8) | N/A | 0.464 |
Implantable cardioverter defibrillator lead | ||||||||
Implantable cardioverter defibrillator lead, n (%) | 169 (16.7) | N/A | 109 (99.1) | 23 (100.0) | N/A | 37 (100.0) | N/A | <0.001 |
Manufacturer | ||||||||
Boston Scientific ICD lead, n (%) | 45 (4.5) | N/A | 14 (12.7) | 22 (95.7) | N/A | 9 (24.3) | N/A | <0.001 |
Medtronic ICD lead, n (%) | 49 (4.9) | N/A | 38 (34.5) | 0 (0.0) | N/A | 11 (29.7) | N/A | 0.001 |
Biotronik ICD lead, n (%) | 37 (3.7) | N/A | 24 (21.8) | 1 (4.3) | N/A | 12 (32.4) | N/A | 0.029 |
St. Jude Medical ICD lead, n (%) | 37 (3.7) | N/A | 32 (29.1) | 0 (0.0) | N/A | 5 (13.5) | N/A | 0.001 |
MicroPort ICD lead, n (%) | 1 (0.1) | N/A | 1 (0.9) | 0 (0.0) | N/A | 0 (0.0) | N/A | >0.999 |
Time since implantation [years], median [IQR] | 2.00 [1.00, 4.08] | N/A | 2.00 [1.08, 5.08] | 1.08 [0.62, 2.50] | N/A | 2.50 [1.08, 3.83] | N/A | 0.049 |
Labeled as MR-conditional by manufacturer, n (%) | 167 (16.5) | N/A | 107 (97.3) | 23 (100.0) | N/A | 37 (100.0) | N/A | >0.999 |
Abandoned leads | ||||||||
Abandoned leads, n (%) | 8 (0.8) | 8 (1.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0.541 |
Number of abandoned leads per patient | ||||||||
One, n (%) | 8 (0.8) | 8 (1.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0.541 |
Manufacturer | ||||||||
Medtronic, n (%) | 2 (0.2) | 2 (0.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Biotronik, n (%) | 4 (0.4) | 4 (0.7) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
St. Jude Medical, n (%) | 2 (0.2) | 2 (0.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Location | ||||||||
RA, n (%) | 5 (0.5) | 5 (0.8) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
RV, n (%) | 3 (0.3) | 3 (0.5) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Time since implantation [years], median [IQR] | 9.88 [6.98, 12.50] | 9.88 [6.98, 12.50] | - | - | - | - | - | - |
Epicardial leads | ||||||||
Epicardial leads, n (%) | 1 (0.1) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (2.7) | 0 (0.0) | 0.089 |
Number of epicardial leads per patient | ||||||||
One, n (%) | 1 (0.1) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (2.7) | 0 (0.0) | 0.089 |
Manufacturer | ||||||||
Medtronic, n (%) | 1 (0.1) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Location | ||||||||
Sinus coronarius, n (%) | 1 (0.1) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (2.7) | 0 (0.0) | 0.089 |
Time since implantation [years], median [IQR] | 0.08 [0.08, 0.08] * | - | - | - | - | - | - | - |
Lead fragment | ||||||||
Lead fragment, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Characteristics | Overall (n = 1010) | PM (n = 604) | ICD (n = 110) | sICD (n = 23) | CRT-P (n = 30) | CRT-D (n = 37) | ILR (n = 206) | p-Value |
---|---|---|---|---|---|---|---|---|
MRI attributes | ||||||||
Anatomic region imaged | ||||||||
Head, n (%) | 543 (53.8) | 324 (53.6) | 74 (67.3) | 7 (30.4) | 15 (50.0) | 10 (27.0) | 113 (54.9) | <0.001 |
Spine, n (%) | 115 (11.4) | 79 (13.1) | 9 (8.2) | 2 (8.7) | 3 (10.0) | 8 (21.6) | 14 (6.8) | 0.046 |
Head and spine, n (%) | 24 (2.4) | 17 (2.8) | 2 (1.8) | 0 (0.0) | 0 (0.0) | 3 (8.1) | 2 (1.0) | 0.189 |
Head and neck, n (%) | 3 (0.3) | 3 (0.5) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0.781 |
Neck, n (%) | 3 (0.3) | 2 (0.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.5) | 1.000 |
Neck and thorax, n (%) | 1 (0.1) | 1 (0.2) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Upper extremity, n (%) | 16 (1.6) | 7 (1.2) | 0 (0.0) | 2 (8.7) | 0 (0.0) | 2 (5.4) | 5 (2.4) | 0.022 |
Upper extremity and spine, n (%) | 1 (0.1) | 1 (0.2) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Thorax, n (%) | 4 (0.4) | 3 (0.5) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (2.7) | 0 (0.0) | 0.355 |
Thorax and abdomen, n (%) | 1 (0.1) | 1 (0.2) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Thorax and spine, n (%) | 1 (0.1) | 1 (0.2) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Breast, n (%) | 8 (0.8) | 8 (1.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0.541 |
Heart, n (%) | 82 (8.1) | 23 (3.8) | 11 (10.0) | 7 (30.4) | 3 (10.0) | 6 (16.2) | 32 (15.5) | <0.001 |
Heart and thorax, n (%) | 1 (0.1) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.5) | 0.402 |
Heart and prostate, n (%) | 1 (0.1) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (2.7) | 0 (0.0) | 0.089 |
Abdomen, n (%) | 95 (9.4) | 67 (11.1) | 1 (0.9) | 0 (0.0) | 2 (6.7) | 2 (5.4) | 23 (11.2) | 0.002 |
Abdomen and lower extremity, n (%) | 1 (0.1) | 1 (0.2) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Rectum, n (%) | 1 (0.1) | 0 (0.0) | 1 (0.9) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0.198 |
Prostate, n (%) | 25 (2.5) | 13 (2.2) | 5 (4.5) | 1 (4.3) | 0 (0.0) | 0 (0.0) | 6 (2.9) | 0.483 |
Lower extremity, n (%) | 74 (7.3) | 45 (7.5) | 7 (6.4) | 4 (17.4) | 7 (23.3) | 4 (10.8) | 7 (3.4) | 0.002 |
Lower extremity and spine, n (%) | 3 (0.3) | 2 (0.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (0.5) | >0.999 |
Lower extremity and prostate, n (%) | 2 (0.2) | 2 (0.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Lower extremity, spine and head, n (%) | 1 (0.1) | 1 (0.2) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Whole body, n (%) | 2 (0.2) | 2 (0.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Characteristics | Overall (n = 1010) | PM (n = 604) | ICD (n = 110) | sICD (n = 23) | CRT-P (n = 30) | CRT-D (n = 37) | ILR (n = 206) | p-Value |
---|---|---|---|---|---|---|---|---|
Related adverse events during and 24h post MRI in patients without abandoned and epicardial leads or lead fragments | ||||||||
Discomfort, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Palpitation, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Heating, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Sensation of device migration, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Abnormal vital signs, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Atrial arrhythmia, n (%) | 2 (0.2) | 2 (0.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | >0.999 |
Ventricular arrhythmia, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
Death, n (%) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | - |
PM | Pre MRI | Post MRI | p-Value |
---|---|---|---|
Battery | |||
Battery level okay, n (%) | 604 (100.0) | 604 (100.0) | - |
Right atrial lead | |||
RA lead Pacing (threshold) [V], median [IQR] | 0.75 [0.50, 0.80] | 0.70 [0.50, 0.80] | 0.290 |
RA lead Sensing [mV], median [IQR] | 3.00 [1.90, 4.20] | 3.00 [1.80, 4.57] | 0.817 |
RA lead Impedance [Ω], median [IQR] | 456.00 [399.00, 526.00] | 447.00 [399.00, 507.00] | 0.411 |
Right ventricular lead | |||
RV lead Pacing (threshold) [V], median [IQR] | 0.75 [0.62, 1.00] | 0.75 [0.62, 1.00] | 0.899 |
RV lead Sensing [mV], median [IQR] | 10.80 [8.00, 13.70] | 10.80 [8.10, 13.90] | 0.779 |
RV lead Impedance [Ω], median [IQR] | 525.00 [456.00, 589.00] | 513.00 [456.00, 589.00] | 0.340 |
Atrial pacing | |||
Atrial pacing rate [%], median [IQR] | 26.00 [2.60, 70.00] | 23.95 [1.55, 74.25] | 0.790 |
Ventricular pacing | |||
Ventricular pacing rate [%], median [IQR] | 14.00 [0.26, 93.00] | 16.00 [0.20, 97.00] | 0.569 |
ICD | Pre MRI | Post MRI | p-value |
Battery | |||
Battery level okay, n (%) | 110 (100.0) | 110 (100.0) | - |
Right atrial lead | |||
RA lead Pacing (threshold) [V], median [IQR] | 0.80 [0.62, 1.25] | 0.80 [0.58, 1.25] | 0.842 |
RA lead Sensing [mV], median [IQR] | 2.35 [1.60, 4.02] | 2.15 [1.52, 4.35] | 0.830 |
RA lead Impedance [Ω], median [IQR] | 446.50 [388.00, 516.30] | 453.00 [388.00, 523.10] | 0.667 |
Right ventricular lead | |||
RV lead Pacing (threshold) [V], median [IQR] | 0.62 [0.60, 1.20] | 0.90 [0.60, 1.32] | >0.999 |
RV lead Sensing [mV], median [IQR] | 11.70 [9.40, 17.20] | 11.70 [11.30, 17.20] | >0.999 |
RV lead Impedance [Ω], median [IQR] | 478.00 [475.00, 487.00] | 478.00 [475.00, 488.00] | >0.999 |
ICD lead | |||
ICD lead Pacing (threshold) [V], median [IQR] | 0.75 [0.60, 1.00] | 0.80 [0.60, 1.00] | 0.606 |
ICD lead Sensing [mV], median [IQR] | 11.50 [8.40, 13.97] | 11.50 [8.40, 13.57] | 0.946 |
ICD lead Impedance [Ω], median [IQR] | 456.00 [403.50, 538.00] | 459.00 [403.00, 538.00] | 0.917 |
Implantable cardioverter defibrillator lead | |||
Shock Impedance [Ω], median [IQR] | 67.50 [59.00, 77.75] | 69.00 [59.00, 77.00] | 0.912 |
Atrial pacing | |||
Atrial pacing rate [%], median [IQR] | 2.35 [0.00, 33.75] | 1.00 [0.00, 35.00] | 0.881 |
Ventricular pacing | |||
Ventricular pacing rate [%], median [IQR] | 0.00 [0.00, 0.91] | 0.00 [0.00, 0.30] | 0.585 |
sICD | Pre MRI | Post MRI | p-value |
Battery | |||
Battery level okay, n (%) | 23 (100.0) | 23 (100.0) | - |
Implantable cardioverter defibrillator lead | |||
Shock Impedance okay, n (%) | 5 (21.7) | 5 (21.7) | - |
CRT-P | Pre MRI | Post MRI | p-value |
Battery | |||
Battery level okay, n (%) | 30 (100.0) | 30 (100.0) | - |
Right atrial lead | |||
RA lead Pacing (threshold) [V], median [IQR] | 0.75 [0.50, 1.00] | 0.75 [0.50, 1.00] | 0.850 |
RA lead Sensing [mV], median [IQR] | 1.90 [0.90, 3.10] | 2.30 [1.00, 3.40] | 0.575 |
RA lead Impedance [Ω], median [IQR] | 418.00 [399.00, 450.00] | 418.00 [380.00, 450.00] | 0.639 |
Right ventricular lead | |||
RV lead Pacing (threshold) [V], median [IQR] | 0.70 [0.50, 0.75] | 0.55 [0.50, 0.75] | 0.906 |
RV lead Sensing [mV], median [IQR] | 12.00 [11.70, 13.20] | 12.00 [11.65, 13.05] | 0.931 |
RV lead Impedance [Ω], median [IQR] | 563.00 [498.75, 592.75] | 563.00 [487.00, 589.00] | 0.749 |
Left ventricular lead | |||
LV lead Pacing (threshold) [V], median [IQR] | 1.65 [1.02, 1.75] | 1.50 [1.00, 2.00] | 0.837 |
LV lead Sensing [mV], median [IQR] | 11.45 [9.45, 14.07] | 9.10 [8.85, 13.20] | >0.999 |
LV lead Impedance [Ω], median [IQR] | 813.50 [556.25, 966.00] | 813.50 [538.25, 966.00] | 0.961 |
Atrial pacing | |||
Atrial pacing rate [%], median [IQR] | 9.10 [0.90, 75.00] | 9.10 [0.90, 75.00] | 0.907 |
Ventricular pacing | |||
Ventricular pacing rate [%], median [IQR] | 94.00 [91.00, 97.70] | 95.00 [91.40, 99.00] | 0.421 |
CRT-D | Pre MRI | Post MRI | p-value |
Battery | |||
Battery level okay, n (%) | 37 (100.0) | 37 (100.0) | - |
Right atrial lead | |||
RA lead Pacing (threshold) [V], median [IQR] | 0.75 [0.68, 0.83] | 0.75 [0.70, 0.90] | 0.616 |
RA lead Sensing [mV], median [IQR] | 2.50 [1.50, 5.30] | 2.40 [1.50, 4.92] | 0.758 |
RA lead Impedance [Ω], median [IQR] | 504.00 [394.25, 565.00] | 505.00 [399.00, 565.00] | 0.995 |
Right ventricular lead | |||
RV lead Pacing (threshold) [V], median [IQR] * | 0.60 [0.60, 0.60] | 0.60 [0.60, 0.60] | - |
RV lead Sensing [mV], median [IQR] * | 23.40 [23.40, 23.40] | 23.40 [23.40, 23.40] | - |
RV lead Impedance [Ω], median [IQR] * | 496.00 [496.00, 496.00] | 496.00 [496.00, 496.00] | - |
* only one value available | |||
Left ventricular lead | |||
LV lead Pacing (threshold) [V], median [IQR] | 1.00 [0.74, 1.32] | 1.00 [0.75, 1.50] | 0.950 |
LV lead Sensing [mV], median [IQR] | 17.90 [13.95, 21.20] | 8.80 [8.75, 16.65] | 0.507 |
LV lead Impedance [Ω], median [IQR] | 559.00 [410.00, 670.00] | 560.00 [381.00, 665.00] | 0.841 |
Implantable cardioverter defibrillator lead | |||
Shock Impedance [Ω], median [IQR] | 62.00 [56.00, 81.50] | 62.00 [56.50, 79.50] | 0.888 |
Atrial pacing | |||
Atrial pacing rate [%], median [IQR] | 26.00 [1.00, 72.50] | 26.00 [1.00, 72.50] | 0.883 |
Ventricular pacing | |||
Ventricular pacing rate [%], median [IQR] | 99.00 [95.10, 99.80] | 99.00 [96.10, 99.80] | 0.857 |
ILR | Pre MRI | Post MRI | p-value |
Battery | |||
Battery level okay, n (%) | 206 (100.0) | 206 (100.0) | - |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Lanz, H.; Strauß, K.; Höpler, J.; Kraft, M.; Hoffmann, S.; Binzenhöfer, L.; Gade, N.; Roden, D.; Saleh, I.; Kääb, S.; et al. Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices. J. Cardiovasc. Dev. Dis. 2024, 11, 313. https://doi.org/10.3390/jcdd11100313
Lanz H, Strauß K, Höpler J, Kraft M, Hoffmann S, Binzenhöfer L, Gade N, Roden D, Saleh I, Kääb S, et al. Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices. Journal of Cardiovascular Development and Disease. 2024; 11(10):313. https://doi.org/10.3390/jcdd11100313
Chicago/Turabian StyleLanz, Hugo, Katharina Strauß, Julia Höpler, Marie Kraft, Sabine Hoffmann, Leonhard Binzenhöfer, Nils Gade, Daniel Roden, Inas Saleh, Stefan Kääb, and et al. 2024. "Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices" Journal of Cardiovascular Development and Disease 11, no. 10: 313. https://doi.org/10.3390/jcdd11100313
APA StyleLanz, H., Strauß, K., Höpler, J., Kraft, M., Hoffmann, S., Binzenhöfer, L., Gade, N., Roden, D., Saleh, I., Kääb, S., Lackermair, K., Sadoni, S., Hagl, C., Massberg, S., Estner, H., Fichtner, S., & Lüsebrink, E. (2024). Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices. Journal of Cardiovascular Development and Disease, 11(10), 313. https://doi.org/10.3390/jcdd11100313